A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation

被引:232
作者
Sommers, MB
Roth, C
Hall, H
Kam, BCC
Ehmke, LW
Krieg, JC
Madey, SM
Bottlang, M
机构
[1] Legacy Clin Res & Technol Ctr, Biomech Lab, Portland, OR 97232 USA
[2] Synthes USA, W Chester, PA USA
[3] Oregon Hlth & Sci Univ, Dept Orthopaed & Rehabil, Portland, OR USA
关键词
hip; cutout model; pertrochanteric fracture; lag screw; fixation;
D O I
10.1097/00005131-200407000-00006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To establish a laboratory model of implant cutout, which can evaluate the effect of implant design on cutout resistance in a clinically realistic "worst case" scenario. Setting: Orthopaedic biomechanics laboratory. Design: Implant cutout was simulated in an unstable pertrochanteric fracture model, which accounted for dynamic loading, osteoporotic bone, and a defined implant offset. For model characterization, lag screw cutout was simulated in human cadaveric specimens and in polyurethane foam surrogates. Subsequently, foam surrogates were used to determine differences in cutout resistance between 2 common lag screws (dynamic hip screw, Gamma) and 2 novel blade-type implant designs (dynamic helical hip system, trochanteric fixation nail). Main Outcome Measures: Implant migration was continuously recorded with a spatial motion tracking system as a function of the applied loading cycles. In addition, the total number of loading cycles to cutout failure was determined for specific load amplitudes. Results: Implant migration in polyurethane surrogates closely correlated with that in cadaveric specimens, but yielded higher reproducibility and consistent cutout failure. The cutout model was able to delineate significant differences in cutout resistance between specific implant designs. At any of 4 load amplitudes (0.8 kN, 1.0 kN, 1.2 kN, 1.4 kN) dynamic hip screw lag screws failed earliest. The gamma nail lag screw could sustain significantly more loading cycles than the dynamic hip screw. Of all implants, trochanteric fixation nail implants demonstrated the highest cutout resistance. Conclusions: Implant design can significantly affect the fixation strength and cutout resistance of implants for pertrochanteric fracture fixation. The novel cutout model can predict differences in cutout resistance between distinct implant designs.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 29 条
[1]   THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[2]  
CLAWSON D K, 1964, J Trauma, V4, P737, DOI 10.1097/00005373-196411000-00001
[3]   INTERTROCHANTERIC FEMORAL FRACTURES - MECHANICAL FAILURE AFTER INTERNAL-FIXATION [J].
DAVIS, TRC ;
SHER, JL ;
HORSMAN, A ;
SIMPSON, M ;
PORTER, BB ;
CHECKETTS, RG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (01) :26-31
[4]   TREATMENT OF THE UNSTABLE INTERTROCHANTERIC FRACTURE - EFFECT OF THE PLACEMENT OF THE SCREW, ITS ANGLE OF INSERTION, AND OSTEOTOMY [J].
DENHARTOG, BD ;
BARTAL, E ;
COOKE, F .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1991, 73A (05) :726-733
[5]   A PROSPECTIVE COMPARATIVE-STUDY OF THE COMPRESSION HIP SCREW AND THE GAMMA-NAIL [J].
GOLDHAGEN, PR ;
OCONNOR, DR ;
SCHWARZE, D ;
SCHWARTZ, E .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1994, 8 (05) :367-372
[6]  
Goodman S B, 1992, J Orthop Trauma, V6, P66
[7]   An experimental study of the failure modes of the gamma locking nail and AO dynamic hip screw under static loading: a cadaveric study [J].
Haynes, RC ;
Poll, RG ;
Miles, AW ;
Weston, RB .
MEDICAL ENGINEERING & PHYSICS, 1997, 19 (05) :446-453
[8]   Failure of femoral head fixation: a cadaveric analysis of lag screw cut-out with the gamma locking nail and AO dynamic hip screw [J].
Haynes, RC ;
Poll, RG ;
Miles, AW ;
Weston, RB .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1997, 28 (5-6) :337-341
[9]  
Jacobs R.R., 1980, CLIN ORTHOP RELAT R, V146, P62
[10]   Type of screw does not influence holding power in the femoral head - A cadaver study with shearing test [J].
Jenny, JY ;
Rapp, E ;
Cordey, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1999, 70 (05) :435-438