Safety and efficacy of transvenous mechanical lead extraction in patients with abandoned leads

被引:16
作者
Segreti, Luca [1 ]
Santoro, Mario Giannotti [1 ]
Di Cori, Andrea [1 ]
Fiorentini, Federico [1 ]
Zucchelli, Giulio [1 ]
Bernini, Giulia [1 ]
De Lucia, Raffaele [1 ]
Viani, Stefano [1 ]
Paperini, Luca [1 ]
Barletta, Valentina [1 ]
Soldati, Ezio [1 ]
Bongiorni, Maria Grazia [1 ]
机构
[1] Univ Hosp Pisa, Cardiol Dept, Via Paradisa 2, I-56124 Pisa, Italy
来源
EUROPACE | 2020年 / 22卷 / 09期
关键词
Lead extraction; Abandoned leads; Implantable cardioverter-defibrillator; Pacemaker; Safety; PROCEDURAL OUTCOMES; SINGLE-CENTER; PREDICTORS; EXPERIENCE; INFECTION; REGISTRY;
D O I
10.1093/europace/euaa134
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims Optimal management of redundant or malfunctioning leads is controversial. We aimed to assess safety and efficacy of mechanical transvenous lead extraction (TLE) in patients with abandoned leads. Methods and results Consecutive TLE procedures performed in our centre from January 2009 to December 2017 were considered. We evaluated the safety and efficacy of mechanical TLE in patients with abandoned (Group 1) compared to non-abandoned (Group 2) leads. We analysed 1210 consecutive patients that required transvenous removal of 2343 leads. Group 1 accounted for 250 patients (21%) with a total of 617 abandoned leads (26%). Group 2 comprised 960 patients (79%) with 1726 leads (74%). The total number of leads (3.0 vs. 2.0), dwelling time of the oldest lead (108.00months vs. 60.00months) and infectious indications for TLE were higher in Group 1. Clinical success was achieved in 1168 patients (96.5%) with a lower rate in Group 1 (90.4% vs. 98.1%; P<0.001). Major complications occurred in only 9 patients (0.7%), without significant differences among the two groups. The presence of one or more abandoned leads [odds ratio (OR) 3.47; 95% confidence interval (CI) 1.07-11.19; P=0.037] and dwelling time of the oldest lead (OR 1.01 for a month; 95% CI 1.01-1.02; P<0.001) were associated with a higher risk of clinical failure. Conclusion Transvenous mechanical lead extraction is a safe procedure also in high-risk settings, as patients with abandoned leads. Success rate resulted a bit lower, especially in the presence of abandoned leads with long implantation time.
引用
收藏
页码:1401 / 1408
页数:8
相关论文
共 20 条
  • [1] Predictors of Complications of Endovascular Chronic Lead Extractions from Pacemakers and Defibrillators: A Single-Operator Experience
    Agarwal, Sunil K.
    Kamireddy, Swapna
    Nemec, Jan
    Voigt, Andrew
    Saba, Samir
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (02) : 171 - 175
  • [2] 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS
    Bongiorni, Maria G.
    Burri, Haran
    Deharo, Jean C.
    Starck, Christoph
    Kennergren, Charles
    Saghy, Laszlo
    Rao, Archana
    Tascini, Carlo
    Lever, Nigel
    Kutarski, Andrzej
    Lozano, Ignacio Fernandez
    Strathmore, Neil
    Costa, Roberto
    Epstein, Laurence
    Love, Charles
    Blomstrom-Lundqvist, Carina
    Fauchier, Laurent
    Defaye, Pascal
    Arnar, David O.
    Klug, Didier
    Boveda, Serge
    Nielsen, Jens Cosedis
    Boriani, Giuseppe
    Zhang, Shu
    Martin, Andrew Paul
    Prutkin, Jordan M.
    de Zuloaga, Claudio
    [J]. EUROPACE, 2018, 20 (07): : 1217 - +
  • [3] The European Lead Extraction ConTRolled (ELECTRa) study: a European Heart Rhythm Association (EHRA) Registry of Transvenous Lead Extraction Outcomes
    Bongiorni, Maria Grazia
    Kennergren, Charles
    Butter, Christian
    Deharo, Jean Claude
    Kutarski, Andrzej
    Rinaldi, Christopher A.
    Romano, Simone L.
    Maggioni, Aldo P.
    Andarala, Maryna
    Auricchio, Angelo
    Kuck, Karl-Heinz
    Blomstrom-Lundqvist, Carina
    [J]. EUROPEAN HEART JOURNAL, 2017, 38 (40) : 2995 - 3005
  • [4] Transvenous extraction profile of Riata leads: Procedural outcomes and technical complexity of mechanical removal
    Bongiorni, Maria Grazia
    Di Cori, Andrea
    Segreti, Luca
    Zucchelli, Giulio
    Viani, Stefano
    Paperini, Luca
    De Lucia, Raffaele
    Levorato, Dianora
    Boem, Adriano
    Soldati, Ezio
    [J]. HEART RHYTHM, 2015, 12 (03) : 580 - 587
  • [5] Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads
    Bongiorni, Maria Grazia
    Soldati, Ezio
    Zucchelli, Giulio
    Di Cori, Andrea
    Segreti, Luca
    De Lucia, Raffaele
    Solarino, Gianluca
    Balbarini, Alberto
    Marzilli, Mario
    Mariani, Mario
    [J]. EUROPEAN HEART JOURNAL, 2008, 29 (23) : 2886 - 2893
  • [6] Lead extraction for cardiac implantable electronic device infection: comparable complication rates with or without abandoned leads
    Bracke, Frank
    Verberkmoes, Niels
    van't Veer, Marcel
    van Gelder, Berry
    [J]. EUROPACE, 2019, 21 (09): : 1378 - 1384
  • [7] Cock CCd, 2000, PACING CLIN ELECTROP, V23, P423
  • [8] Large, Single-Center Experience in Transvenous Coronary Sinus Lead Extraction: Procedural Outcomes and Predictors for Mechanical Dilatation
    Di Cori, Andrea
    Bongiorni, Maria Grazia
    Zucchelli, Giulio
    Segreti, Luca
    Viani, Stefano
    De Lucia, Raffaele
    Paperini, Luca
    Soldati, Ezio
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (02): : 215 - 222
  • [9] Transvenous Extraction Performance of Expanded Polytetrafluoroethylene Covered ICD Leads in Comparison to Traditional ICD Leads in Humans
    Di Cori, Andrea
    Bongiorni, Maria Grazia
    Zucchelli, Giulio
    Segreti, Luca
    Viani, Stefano
    Paperini, Luca
    Soldati, Ezio
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2010, 33 (11): : 1376 - 1381
  • [10] Risk factors for infection of implantable cardiac devices: data from a registry of 2496 patients
    Herce, Benoit
    Nazeyrollas, Pierre
    Lesaffre, Francois
    Sandras, Raphael
    Chabert, Jean-Pierre
    Martin, Angeline
    Tassan-Mangina, Sophie
    Huu Tri Bui
    Metz, Damien
    [J]. EUROPACE, 2013, 15 (01): : 66 - 70