Evaluation of a new troponin I method on the Bayer Immuno 1™ immunoassay analyser

被引:3
作者
Clark, GH [1 ]
Kennon, SRO
Price, CP
机构
[1] Barts & London NHS Trust, Dept Clin Biochem, London, England
[2] Barts & London NHS Trust, Dept Cardiol, London, England
[3] St Bartholomews & Royal London Sch Med & Dent, Dept Clin Biochem, London, England
来源
JOURNAL OF IMMUNOASSAY | 1999年 / 20卷 / 04期
关键词
troponins; immunoassay; myocardial infarction; clinical cut off values;
D O I
10.1080/01971529909349354
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
We have evaluated the analytical and clinical performance of an automated immunoassay for serum cardiac troponin I (Bayer Immune 1(TM), Bayer Diagnostics, Tarrytown, NY). The between batch imprecision was found to be between 1.2 and 3.2% over the concentration range 2.5 - 34.0 mu g/L. The analytical range obtained from duplicate analysis of patient samples and defined as a coefficient of variation of 10% or less was 0.3 - 200 mu g/L. The detection limit was found to be less than 0.1 mu g/L. A method comparison with the Dade Stratus method (Dade Behring, Wilmington, DE) yielded regression statistics with a slope of 0.705 and an intercept of -0.260. An analysis of samples from 40 patients with renal failure demonstrated six with detectable levels of troponin 1 (0.2 - 1.9 mu g/L). Samples from patients with paraproteinaemia did not demonstrate detectable troponin I (from n = 30); however, two patients with elevated rheumatoid factor titers (from n = 20) demonstrated a detectable amount of troponin I (0.1 and 0.2 mu g/L). In a study of 100 patients admitted with acute chest pain and a diagnosis of unstable angina, 6 were subsequently diagnosed as having suffered a myocardial infarction. On admission the sensitivity and specificity of the troponin I results were 26.7% and 94.7%, respectively, moving to 100% and 83% 12 hours after admission.
引用
收藏
页码:253 / 273
页数:21
相关论文
共 34 条
  • [1] ADAMS JE, 1994, CLIN CHEM, V40, P1291
  • [2] DIAGNOSIS OF PERIOPERATIVE MYOCARDIAL-INFARCTION WITH MEASUREMENT OF CARDIAC TROPONIN-I
    ADAMS, JE
    SICARD, GA
    ALLEN, BT
    BRIDWELL, KH
    LENKE, LG
    DAVILAROMAN, VG
    BODOR, GS
    LADENSON, JH
    JAFFE, AS
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1994, 330 (10) : 670 - 674
  • [3] CARDIAC TROPONIN-I - A MARKER WITH HIGH SPECIFICITY FOR CARDIAC INJURY
    ADAMS, JE
    BODOR, GS
    DAVILAROMAN, VG
    DELMEZ, JA
    APPLE, FS
    LADENSON, JH
    JAFFE, AS
    [J]. CIRCULATION, 1993, 88 (01) : 101 - 106
  • [4] Cardiac-specific troponin I levels to predict the risk of mortality in patients with acute coronary syndromes
    Antman, EM
    Tanasijevic, MJ
    Thompson, B
    Schactman, M
    McCabe, CH
    Cannon, CP
    Fischer, GA
    Fung, AY
    Thompson, C
    Wybenga, D
    Braunwald, E
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (18) : 1342 - 1349
  • [5] Apple FS, 1999, CLIN CHEM, V45, P206
  • [6] Apple FS, 1998, PATHOL LAB MED, V3, P229
  • [7] Apple FS, 1999, CLIN CHEM, V45, P18
  • [8] Cardiac troponin I should replace CKMB for the diagnosis of acute myocardial infarction
    Bhagat, CI
    Langton, P
    Lewer, M
    Ching, S
    Beilby, JP
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1997, 34 : 511 - 516
  • [9] BHAYANA V, 1995, CLIN CHEM, V41, P312