Investigations of soft and hard tissues in oral cavity by Spectral Domain Optical Coherence Tomography

被引:3
作者
Madjarova, Violeta Dimitrova [1 ]
Yasuno, Yoshiaki [1 ]
Makita, Shuichi [1 ]
Hori, Yasuaki [1 ]
Voeffray, Jean-Baptiste [1 ]
Itoh, Masahide [1 ]
Yatagai, Toyohiko [1 ]
Tamura, Masami [2 ]
Nanbu, Toshiyuki [2 ]
机构
[1] Univ Tsukuba, Inst Appl Phys, Tennodai 1-1-1, Tsukuba, Ibaraki 3058573, Japan
[2] Shofu Inc, Higashiyama Ku, Kyoto 6050983, Japan
来源
COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE X | 2006年 / 6079卷
基金
日本科学技术振兴机构; 日本学术振兴会;
关键词
Optical Coherence Tomography; SD-OCT; SS-OCT; dental measurements;
D O I
10.1117/12.648359
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Fourier Domain Optical Coherence Tomography (SD-OCT) systems for dental measurements are demonstrated. Two systems have been developed. The first system is fiber based Michelson interferometer with super luminescent diodes at 1310 nm and 100 nm FWHM as a light source. The sensitivity of the system was 106 dB with depth measurement range in air of 2.5 mm. The second systems is a fiber based Mach-Zehnder interferometer with wavelength scanning laser as light source at center wavelength of 1310 nm, wavelength range of 110 nm and scanning rate of 20 KHz. The sensitivity of the system is 112 dB and depth measurement range in air is 6 mm. Both systems can acquire real-time three dimensional (3-D) images in the range of several second. The systems were applied for early caries detection in tooth, for diagnostics of tooth condition after operational tooth treatment, and for diagnostics of the alveolar bone structure. In-vivo measurements were performed on two volunteers. The systems were able to detect discontinuities in tooth and resin filling after tooth treatment. In addition early carries lesion was detected in one of the volunteers. The 3-D profile of the alveolar bone was acquired for first time with non-contact method.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] Bouma B.E., 2002, Handbook of Optical Coherence Tomography
  • [2] Real-time in vivo imaging of dental tissue by means of optical coherence tomography (OCT)
    Brandenburg, R
    Haller, B
    Hauger, C
    [J]. OPTICS COMMUNICATIONS, 2003, 227 (4-6) : 203 - 211
  • [3] Optical coherence tomography using a frequency-tunable optical source
    Chinn, SR
    Swanson, EA
    Fujimoto, JG
    [J]. OPTICS LETTERS, 1997, 22 (05) : 340 - 342
  • [4] Sensitivity advantage of swept source and Fourier domain optical coherence tomography
    Choma, MA
    Sarunic, MV
    Yang, CH
    Izatt, JA
    [J]. OPTICS EXPRESS, 2003, 11 (18): : 2183 - 2189
  • [5] Dental OCT
    Colston, BW
    Sathyam, US
    DaSilva, LB
    Everett, MJ
    Stroeve, P
    Otis, LL
    [J]. OPTICS EXPRESS, 1998, 3 (06): : 230 - 238
  • [6] In vivo OCT imaging of hard and soft tissue of the oral cavity
    Feldchtein, FI
    Gelikonov, GV
    Gelikonov, VM
    Iksanov, RR
    Kuranov, RV
    Sergeev, AM
    Gladkova, ND
    Ourutina, MN
    Warren, JA
    Reitze, DH
    [J]. OPTICS EXPRESS, 1998, 3 (06): : 239 - 250
  • [7] Fercher AF, 2003, REP PROG PHYS, V66, P239, DOI [10.1088/0034-4885/66/2/204, 10.2184/lsj.31.635]
  • [8] NATURE OF LIGHT-SCATTERING IN DENTAL ENAMEL AND DENTIN AT VISIBLE AND NEAR-INFRARED WAVELENGTHS
    FRIED, D
    GLENA, RE
    FEATHERSTONE, JDB
    SEKA, W
    [J]. APPLIED OPTICS, 1995, 34 (07) : 1278 - 1285
  • [9] Imaging caries lesions and lesion progression with polarization sensitive optical coherence tomography
    Fried, D
    Xie, J
    Shafi, S
    Featherstone, JDB
    Breunig, TM
    Le, C
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2002, 7 (04) : 618 - 627
  • [10] Häusler G, 1998, J BIOMED OPT, V3, P21, DOI 10.1117/1.429899