摘要
A new microflow injection analysis (mu FIA) system on a chip for the determination of nitrite is described. The chip is produced by using two transparent poly(methylmethacrylate) (PMMA) slices measured 50 x 40 x 5 mm, and the microchannels etched by CO2 laser are 200 mu m wide and 100 mu m deep with the volume of reaction area about 1.8 mu L. Nitrite is sensed by the chemiluminescence (CL) reaction of luminol with ferricyanide that is the product of the reaction of ferrocyanide with nitrite in acidic medium. The syringe pump with an accurate timer controls all reagents, including the sample. The linear range of the nitrite concentration is 8-100 mu g L-1 and the detection limit is 4 mu g L-1 (S/N = 3). The proposed method has good reproducibility with the relative standard deviation 4.1% for 50 mu g L-1 of nitrite (n = 9) and is very sensitive and simple. It has been successfully applied to. the determination of nitrite in food. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:667 / 672
页数:6
相关论文
共 23 条
  • [1] AOAC, 1997, OFF METH AN
  • [2] Determination of nitrite in waters by microplate fluorescence spectroscopy and HPLC with fluorescence detection
    Büldt, A
    Karst, U
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (15) : 3003 - 3007
  • [4] Sequential injection sample introduction microfluidic-chip based capillary electrophoresis system
    Fang, Q
    Wang, FR
    Wang, SL
    Liu, SS
    Xu, SK
    Fang, ZL
    [J]. ANALYTICA CHIMICA ACTA, 1999, 390 (1-3) : 27 - 37
  • [5] A nanoinjector for microanalysis
    Gorbounov, V
    Kuban, P
    Dasgupta, PK
    Temkin, H
    [J]. ANALYTICAL CHEMISTRY, 2003, 75 (15) : 3919 - 3923
  • [6] Tris(2,2-bipyridyl)ruthenium (II) chemiluminescence in a microflow injection system for codeine determination
    Greenway, GM
    Nelstrop, LJ
    Port, SN
    [J]. ANALYTICA CHIMICA ACTA, 2000, 405 (1-2) : 43 - 50
  • [7] Chemiluminescence μTAS for the determination of atropine and pethidine
    Greenwood, PA
    Merrin, C
    McCreedy, T
    Greenway, GM
    [J]. TALANTA, 2002, 56 (03) : 539 - 545
  • [8] Griess P., 1879, Berichte der deutschen chemischen Gesellschaft, V12, P426, DOI DOI 10.1002/CBER.187901201117
  • [9] HARRISON DJ, 1993, ANAL CHIM ACTA, V283, P361
  • [10] FUSED QUARTZ SUBSTRATES FOR MICROCHIP ELECTROPHORESIS
    JACOBSON, SC
    MOORE, AW
    RAMSEY, JM
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (13) : 2059 - 2063