Evaluation of the Sebia CAPILLARYS 2 Flex Piercing for the Measurement of HbA1c on Venous and Capillary Blood Samples

被引:20
|
作者
Heylen, Olivier [1 ]
Van Neyghem, Stefaan [1 ]
Exterbille, Sandra [1 ]
Wehlou, Charline [1 ]
Gorus, Frans [1 ]
Weets, Ilse [1 ]
机构
[1] Univ Ziekenhuis Brussel, Dept Clin Chem & Radioimmunol, Brussels, Belgium
关键词
Analytical performance; Capillary zone electrophoresis; Hemoglobin A(1c); HPLC; HEMOGLOBIN A(1C) MEASUREMENT; ANALYTICAL PERFORMANCE; MULTICENTER EVALUATION; ASSAY; ANALYZER; GOALS;
D O I
10.1309/AJCPRU5QC2JBANSV
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Objectives: We evaluated the Sebia CAPILLARYS 2 Flex Piercing (Cap 2FP; Sebia, Lisses, France) for measurement of hemoglobin A(1c) (HbA(1c)) on venous and capillary blood samples. Methods: We analyzed whole-blood samples and control materials with the Cap 2FP and Tosoh G8 (Tosoh Corporation, Tokyo, Japan). Capillary blood samples were analyzed on the Cap 2FP on different storage conditions and were compared with venous samples. Results: Both instruments achieved total imprecision of less than 2.5% (International Federation of Clinical Chemistry units). Bias was 1 mmol/mol or less and 4 mmol/mol or less for the Cap 2FP and Tosoh G8, respectively. The Cap 2FP was not prone to common interferences. The Tosoh G8 showed significant bias only for carbamylated hemoglobin and did not completely separate hemoglobin D and hemoglobin E. On the Cap 2FP, storage of capillary blood at room temperature showed no significant bias. There was good agreement with venous blood Conclusions: The Cap 2FP and Tosoh G8 perform excellently for HbA(1c) determination. Capillary blood can be analyzed on the Cap 2FP as an acceptable alternative to venous blood and point-of-care testing. Home collection and central analysis of capillary blood could contribute to a reduction of health care costs without reducing quality of HbA(1c) determination.
引用
收藏
页码:867 / 877
页数:11
相关论文
共 50 条
  • [1] Evaluation of HbA1c on capillary blood samples with Sebia CAPILLARYS 2 Flex Piercing
    Heylen, O.
    Van Neyghem, S.
    Exterbille, S.
    Wehlou, C.
    Gorus, F.
    Weets, I.
    ACTA CLINICA BELGICA, 2014, 69 : 7 - 8
  • [2] HbA1c: performance of the Sebia Capillarys 2 Flex Piercing
    Weykamp, Cas
    Waenink-Wieggers, Helene
    Kemna, Erwin
    Siebelder, Carla
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (06) : E129 - E131
  • [3] EVALUATION OF CAPILLARYS 2 FLEX PIERCING® (SEBIA) AS A NEW ANALYZER FOR HBA1C ASSAY
    Jaisson, S.
    Meurice, J.
    Leroy, N.
    Gillery, P.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S383 - S383
  • [4] First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis
    Jaisson, Stephane
    Leroy, Nathalie
    Meurice, Julie
    Guillard, Emmanuelle
    Gillery, Philippe
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (10) : 1769 - 1775
  • [5] Evaluation of HbA1c measurement on Capillarys 2 Flex Piercing and Menarini HA8160
    Van den Poel, B.
    Driesen, M.
    Vastmans, L.
    Reynders, C.
    Penders, J.
    ACTA CLINICA BELGICA, 2014, 69 : 21 - 21
  • [6] Measurement of HbA1C and HbA2 by Capillarys 2 Flex Piercing HbA1c programme for simultaneous management of diabetes and screening for thalassemia
    Ke, Peifeng
    Liu, Jiawei
    Chao, Yan
    Wu, Xiaobin
    Xiong, Yujuan
    Lin, Li
    Wan, Zemin
    Wu, Xinzhong
    Xu, Jianhua
    Zhuang, Junhua
    Huang, Xianzhang
    BIOCHEMIA MEDICA, 2017, 27 (03) : 515 - 521
  • [7] Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser)
    Kuna, Andrea Tesija
    Dukic, Kristina
    Gabaj, Nora Nikolac
    Miler, Marijana
    Vukasovic, Ines
    Langer, Sanja
    Simundic, Ana-Maria
    Vrkic, Nada
    LABORATORY MEDICINE, 2018, 49 (03) : 231 - 238
  • [8] Capillarys 2 Flex Piercing: Analytical performance assessment according to CLSI protocols for HbA1c quantification
    Doggui, Radhouene
    Sahli, Chaima Abdelhafidh
    Aissa, Wassef Lotfi
    Hammami, Maroua
    Ben Sedrine, Maha
    Mahjoub, Rahma
    Zouaoui, Khemais
    Daboubi, Rim
    Siala, Hajer
    Messaoud, Taieb
    Bibi, Amina
    ELECTROPHORESIS, 2017, 38 (17) : 2210 - 2218
  • [9] Evaluation of the Capillarys 2 Flex Piercing system for the determination of hemoglobin A(1c) (HbA(1c))
    Paleari, Renata
    Mosca, Andrea
    BIOCHIMICA CLINICA, 2014, 38 (02) : 110 - 114
  • [10] Hemoglobin fractionation by Sebia Capillarys 2 Flex Piercing System as primary method for evaluation of hemoglobinopathies
    Poventud-Fuentes, Izmarie
    Garnett, Emily
    Vispo, Beverly
    Elghetany, M. Tarek
    Devaraj, Sridevi
    CLINICA CHIMICA ACTA, 2021, 519 : 193 - 197