Internal quality control system for non-stationary, non-ergodic analytical processes based upon exponentially weighted estimation of process means and process standard deviation

被引:12
作者
Jansen, RTP [1 ]
Laeven, M [1 ]
Kardol, W [1 ]
机构
[1] St Anna Hosp, Dept Clin Chem, NL-5660 AB Geldrop, Netherlands
关键词
quality control; statistics; exponentially weighted estimation; moving standard deviation; Westgard;
D O I
10.1515/CCLM.2002.107
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The analytical processes in clinical laboratories should be considered to be non-stationary, non-ergodic and probably non-stochastic processes. Both the process mean and the process standard deviation vary. The variation can be different at different levels of concentration. This behavior is shown in five examples of different analytical systems: alkaline phosphatase on the Hitachi 911 analyzer (Roche), vitamin B-12 on the Access analyzer (Beckman), prothrombin time and activated partial thromboplastin time on the STA Compact analyzer (Roche) and pO(2) on the ABL 520 analyzer (Radiometer). A model is proposed to assess the status of a process. An exponentially weighted moving average and standard deviation was used to estimate process mean and standard deviation. Process means were estimated overall and for each control level. The process standard deviation was estimated in terms of withinrun standard deviation. Limits were defined in accordance with state of the art- or biological variancederived cut-offs. The examples given are real, not simulated, data. Individual control sample results were normalized to a target value and target standard deviation. The normalized values were used in the exponentially weighted algorithm. The weighting factor was based on a process time constant, which was estimated from the period between two calibration or maintenance procedures. The proposed system was compared with Westgard rules. The Westgard rules perform well, despite the underlying presumption of ergodicity. This is mainly caused by the introduction of the starting rule of 1(2s), which proves essential to prevent a large number of rule violations. The probability of reporting a test result with an analytical error that exceeds the total allowable error was calculated for the proposed system as well as for the Westgard rules. The proposed method performed better. The proposed algorithm was implemented in a computer program running on computers to which the analyzers were linked on-line. Each result was evaluated on-line, and a limit violation was immediately reported. The system has performed satisfactorily in our laboratory for ten analyzers for over 1 year.
引用
收藏
页码:616 / 624
页数:9
相关论文
共 10 条
[1]  
Dechert J, 1998, CLIN CHEM, V44, P1959
[2]  
HENRY RJ, 1959, CLIN CHEM, V5, P309
[3]   VARIANCE REDUCTION IN ANALYTICAL PROCESSES IN THE CLINICAL LABORATORY BY DIGITAL FILTERING [J].
JANSEN, RTP ;
BONANTS, PJM .
ANNALS OF CLINICAL BIOCHEMISTRY, 1983, 20 (MAY) :174-181
[4]   REDUCTION OF ANALYTICAL VARIANCE BY USING A DISCRETE-TIME DATA-WEIGHTING FILTER TO ESTIMATE ABRUPT CHANGES IN BATCH-TYPE PROCESSES .2. APPLICATIONS [J].
JANSEN, RTP ;
POULISSE, HNJ .
ANALYTICA CHIMICA ACTA, 1983, 151 (02) :441-446
[5]   THE USE OF CONTROL CHARTS IN THE CLINICAL LABORATORY [J].
LEVEY, S ;
JENNINGS, ER .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1950, 20 (11) :1059-1066
[6]  
Neubauer AS, 1997, CLIN CHEM, V43, P594
[7]  
Parvin CA, 1997, CLIN CHEM, V43, P602
[8]   REDUCTION OF ANALYTICAL VARIANCE BY USING A DISCRETE-TIME DATA-WEIGHTING FILTER TO ESTIMATE ABRUPT CHANGES IN BATCH-TYPE PROCESSES .1. THEORY [J].
POULISSE, HNJ ;
JANSEN, RTP .
ANALYTICA CHIMICA ACTA, 1983, 151 (02) :433-439
[9]  
STEIGSTRA H, 1991, CLIN CHEM, V37, P1196
[10]  
WESTGARD JO, 1981, CLIN CHEM, V27, P493