Analysis of subtraction methods in three-dimensional contrast-enhanced peripheral MR angiography

被引:6
作者
Huang, YX
Webster, CA
Wright, GA
机构
[1] Sunnybrook & Womens Coll, Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
关键词
magnetic resonance (MR) digital subtraction; angiography; complex subtraction; maximum intensity projection; peripheral vascular disease; contrast-enhanced;
D O I
10.1002/jmri.10100
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the effectiveness of three image subtraction algorithms designed to improve arterial conspicuity in first-pass contrast-enhanced magnetic resonance (MR) angiography. Materials and Methods: Three subtraction methods were analyzed through computer simulations, phantom studies, and clinical studies. These algorithms were: complex subtraction, magnitude subtraction, and maximum intensity projection subtraction. Results: In high resolution three-dimensional imaging, maximum intensity projection subtraction generally yields the best background suppression. Complex subtraction is effective in reducing partial volume effects in low resolution imaging, Magnitude subtraction works better in high resolution, low contrast concentration protocols. Conclusion: Choosing the appropriate subtraction method according to the protocol is helpful in optimizing image quality. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:541 / 550
页数:10
相关论文
共 16 条
[1]   CONTRAST-TO-NOISE RATIOS IN MAXIMUM INTENSITY PROJECTION IMAGES [J].
BROWN, DG ;
RIEDERER, SJ .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) :130-137
[2]   A SIGNAL-TO-NOISE CALIBRATION PROCEDURE FOR NMR IMAGING-SYSTEMS [J].
EDELSTEIN, WA ;
BOTTOMLEY, PA ;
PFEIFER, LM .
MEDICAL PHYSICS, 1984, 11 (02) :180-185
[3]  
GAUCHER H, 2000, P 8 ANN M ISMRM DENV, P184
[4]   THE RICIAN DISTRIBUTION OF NOISY MRI DATA [J].
GUDBJARTSSON, H ;
PATZ, S .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (06) :910-914
[5]   CORRECTION [J].
HENKELMAN, RM .
MEDICAL PHYSICS, 1986, 13 (04) :544-544
[6]   MEASUREMENT OF SIGNAL INTENSITIES IN THE PRESENCE OF NOISE IN MR IMAGES [J].
HENKELMAN, RM .
MEDICAL PHYSICS, 1985, 12 (02) :232-233
[7]   Peripheral vascular tree stenoses: Detection with subtracted and nonsubtracted MR angiography [J].
Ho, KYJAM ;
de Haan, MW ;
Kessels, AGH ;
Kitslaar, PJEHM ;
van Engelshoven, JMA .
RADIOLOGY, 1998, 206 (03) :673-681
[8]  
Leiner T, 2000, JMRI-J MAGN RESON IM, V11, P368, DOI 10.1002/(SICI)1522-2586(200004)11:4<368::AID-JMRI4>3.0.CO
[9]  
2-5
[10]   Value of image subtraction in 3D gadolinium-enhanced MR angiography of the renal arteries [J].
Leung, DA ;
Pelkonen, P ;
Hany, TF ;
Zimmermann, G ;
Pfammatter, T ;
Debatin, JF .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (03) :598-602