A highly sensitive technique for measurements of the Kerr electrooptic coefficient in glasses and glass ceramics

被引:12
作者
Vetrov, AA
Lipovskii, AA
Tagantsev, DK
机构
[1] All Russia Sci Ctr, SI Vavilov State Opt Inst, St Petersburg 199034, Russia
[2] St Petersburg State Tech Univ, St Petersburg 195251, Russia
关键词
Physical Chemistry; Measurement Method; Measured Parameter; Optical Quality; Comparative Measurement;
D O I
10.1023/A:1019707107722
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A technique for measuring the Kerr electrooptic coefficient in glassy optical materials and the corresponding laboratory setup are described. The technique proposed, which is based on the application of an alternating voltage of a sound frequency to the samples under characterization, ensures the compensation of the initial birefringence of the samples and shows a high sensitivity (B-min = 10(-17) m/V-2 and Deltan(min)=0.8x10(-8)) at moderate requirements for the optical quality of the samples. The comparative measurement method, the unaltered opto-mechanical part of the setup, and the minimum number of the measured parameters make it possible to eliminate a number of sources of errors and to offer the prospect for developing an automated setup.
引用
收藏
页码:550 / 554
页数:5
相关论文
共 5 条
[1]  
BORELLI NF, 1991, J APPL PHYS, V70, P2774
[2]   Electrooptic glasses and glass-ceramics for elements controlling laser radiation [J].
Karapetyan, GO ;
Lipovskii, AA ;
Loboda, VV ;
Maksimov, LV ;
Svistunov, DV ;
Tagantsev, DK ;
Tatarintsev, BV ;
Vetrov, AA .
LASER OPTICS 2000: CONTROL OF LASER BEAM CHARACTERISTICS AND NONLINEAR METHODS FOR WAVEFRONT CONTROL, 2001, 4353 :23-28
[3]  
KERR J, 1880, PHILOS MAG 5, P157
[4]  
MUSTEL ER, 1970, METODY MODULYATSII S
[5]  
Yariv A., 1984, Optical Waves in Crystals