Evaluation of Bone Heating, Drill Deformation, and Drill Roughness After Implant Osteotomy: Guided Surgery and Classic Drilling Procedure

被引:50
作者
dos Santos, Pamela Leticia [1 ]
Queiroz, Thallita Pereira [2 ,3 ]
Margonar, Rogerio [4 ,5 ]
Gomes de Souza Carvalho, Abrahao Cavalcante [6 ]
Betoni, Walter, Jr. [7 ,8 ]
Rodrigues Rezende, Regis Rocha [9 ]
dos Santos, Paulo Henrique [10 ]
Garcia, Idelmo Rangel, Jr. [11 ]
机构
[1] Univ Sagrado Coracao, Dept Oral Biol, Discipline Oral & Maxillofacial Surg & Integrated, Bauru, SP, Brazil
[2] Univ Ctr Araraquara UNIARA, Discipline Oral & Maxillofacial Surg & Integrated, Araraquara, SP, Brazil
[3] Centro Univ Araraquara, Postgrad Course Dent lmplantol, Araraquara, SP, Brazil
[4] Centro Univ Araraquara, Univ Ctr Araraquara, Discipline Periodontol & Integrated Clin, Araraquara, SP, Brazil
[5] Centro Univ Araraquara, Postgrad Course Dent Implantol, Araraquara, Brazil
[6] Sao Paulo State Univ UNESP, Dent Sch Aracatuba, Dept Surg & Integrated Clin, Discipline Oral & Maxillofacial Surg, Aracatuba, SP, Brazil
[7] Univ Cuiaba UNIC, Dept Oral Surg, Cuiaba, Brazil
[8] Dent Sch Cuiaba, Mato Grosso, Brazil
[9] UniFEB, Univ Ctr Barretos, Sao Paulo, Brazil
[10] UNESP, Dept Dent Mat & Prosthodont, Discipline Dent Mat, Aracatuba, SP, Brazil
[11] UNESP, Dent Sch Aragatuba, Dept Surg, Discipline Oral & Maxillofacial Surg, Aracatuba, SP, Brazil
关键词
bone overheating; dental implants; osteotomy; OSSEOINTEGRATED TITANIUM IMPLANTS; SITE PREPARATION; CELL VIABILITY; WEAR; GENERATION;
D O I
10.11607/jomi.2919
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study evaluated and compared bone heating, drill deformation, and drill roughness after several implant osteotomies in the guided surgery technique and the classic drilling procedure. Materials and Methods: The tibias of 20 rabbits were used. The animals were divided into a guided surgery group (GG) and a control group (CG); subgroups were then designated (G0, G1, G2, G3, and G4, corresponding to drills used 0, 10, 20, 30 and 40 times, respectively). Each animal received 10 sequential osteotomies (5 in each tibia) with each technique. Thermal changes were quantified, drill roughness was measured, and the drills were subjected to scanning electron microscopy. Results: Bone temperature generated by drilling was significantly higher in the GG than in the CG. Drill deformation in the GG and CG increased with drill use, and in the CG a significant difference between GO and groups G3 and G4 was observed. In the GG, a significant difference between GO and all other groups was found. For GG versus CG, a significant difference was found in the 40th osteotomy. Drill roughness in both groups was progressive in accordance with increased use, but there was no statistically significant difference between subgroups or between GG and CG overall. Conclusion: During preparation of implant osteotomies, the guided surgery technique generated a higher bone temperature and deformed drills more than the classic drilling procedure. The increase in tissue temperature was directly proportional to the number of times drills were used, but neither technique generated critical necrosis-inducing temperatures. Drill deformation was directly proportional to the number of times the drills were used. The roughness of the drills was directly proportional to the number of reuses in both groups but tended to be higher in the GG group.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 26 条
[1]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[2]   Descriptive Study of the Longevity of Dental Implant Surgery Drills [J].
Allsobrook, Olive F. L. ;
Leichter, Jonathan ;
Holborow, Douglas ;
Swain, Michael .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2011, 13 (03) :244-254
[3]   Thermal changes observed at implant sites during internal and external irrigation [J].
Benington, IC ;
Biagioni, PA ;
Briggs, J ;
Sheridan, S ;
Lamey, PJ .
CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (03) :293-297
[4]  
BRANEMARK PI, 1985, TISSUE INTEGRATED PR, P199
[5]   Heat production by 3 implant drill systems after repeated drilling and sterilization [J].
Chacon, GE ;
Bower, DL ;
Larsen, PE ;
McGlumphy, EA ;
Beck, FM .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 64 (02) :265-269
[6]  
Cordioli G, 2000, INT J ORAL MAX IMPL, V15, P668
[7]  
Cordioli G, 1997, INT J ORAL MAX IMPL, V12, P186
[8]   Computer-Guided Surgery in Implantology: Review of Basic Concepts [J].
de Almeida, Erika Oliveira ;
Pellizzer, Eduardo Piza ;
Goiatto, Marcelo Coelho ;
Margonar, Rogerio ;
Rocha, Eduardo Passos ;
Freitas, Amilcar Chagas, Jr. ;
Anchieta, Rodolfo Bruniera .
JOURNAL OF CRANIOFACIAL SURGERY, 2010, 21 (06) :1917-1921
[9]   Guided Implant Surgery: What Is the Influence of This New Technique on Bone Cell Viability? [J].
dos Santos, Pamela Leticia ;
Queiroz, Thallita Pereira ;
Margonar, Rogerio ;
Gomes de Souza Carvalho, Abrahao Cavalcante ;
Okamoto, Roberta ;
de Souza Faloni, Ana Paula ;
Garcia Junior, Idelmo Rangel .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2013, 71 (03) :505-512
[10]  
Ellingsen JE, 2004, INT J ORAL MAX IMPL, V19, P659