Wearable Finger Pulse Oximetry for Continuous Oxygen Saturation Measurements During Daily Home Routines of Patients With Chronic Obstructive Pulmonary Disease (COPD) Over One Week: Observational Study

被引:74
作者
Buekers, Joren [1 ,2 ]
Theunis, Jan [1 ]
De Boever, Patrick [1 ,3 ]
Vaes, Anouk W. [4 ]
Koopman, Maud [4 ]
Janssen, Eefje V. M. [4 ]
Wouters, Ernie F. M. [4 ,5 ]
Spruit, Martijn A. [4 ,5 ,6 ,7 ]
Aerts, Jean-Marie [2 ]
机构
[1] Flemish Inst Technol Res VITO Mol, Hlth Unit, Mol, Belgium
[2] Katholieke Univ Leuven, Dept Biosyst, Measure Model & Manage Bioresponses, Kasteelpk Arenberg 30, B-3000 Leuven, Belgium
[3] Hasselt Univ, Ctr Environm Sci, Diepenbeek, Belgium
[4] Ctr Expertise Chron Organ Failure CIRO, Dept Res & Educ, Horn, Netherlands
[5] Maastricht Univ, Dept Resp Med, Med Ctr, Maastricht, Netherlands
[6] Hasselt Univ, Fac Rehabil Sci, Rehabil Res Ctr REVAL, Biomed Res Inst BIOMED, Diepenbeek, Belgium
[7] Maastricht Univ, Sch Nutr & Translat Res Metab NUTRIM, Med Ctr, Maastricht, Netherlands
关键词
COPD; oxygen saturation; finger pulse oximeter; wearable sensor; nocturnal desaturation; telemonitoring; ACUTE EXACERBATIONS; WALK TEST; DESATURATION; SLEEP; HEMODYNAMICS; HYPOXEMIA; VARIABLES; MODERATE; THERAPY; ARMBAND;
D O I
10.2196/12866
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Chronic obstructive pulmonary disease (COPD) patients can suffer from low blood oxygen concentrations. Peripheral blood oxygen saturation (SpO(2)), as assessed by pulse oximetry, is commonly measured during the day using a spot check, or continuously during one or two nights to estimate nocturnal desaturation. Sampling at this frequency may overlook natural fluctuations in SpO(2). Objective: This study used wearable finger pulse oximeters to continuously measure SpO(2) during daily home routines of COPD patients and assess natural SpO(2) fluctuations. Methods: A total of 20 COPD patients wore a WristOx(2) pulse oximeter for 1 week to collect continuous SpO(2) measurements. A SenseWear Armband simultaneously collected actigraphy measurements to provide contextual information. SpO(2) time series were preprocessed and data quality was assessed afterward. Mean SpO(2) , SpO(2) SD, and cumulative time spent with SpO(2) below 90% (CT90) were calculated for every (1) day, (2) day in rest, and (3) night to assess SpO(2) fluctuations. Results: A high percentage of valid SpO(2) data (daytime: 93.27%; nocturnal: 99.31%) could be obtained during a 7-day monitoring period, except during moderate-to-vigorous physical activity (MVPA) (67.86%). Mean nocturnal SpO(2) (89.9%, SD 3.4) was lower than mean daytime SpO(2 )in rest (92.1%, SD 2.9; P<.001). On average, SpO(2) in rest ranged over 10.8% (SD 4.4) within one day. Highly varying CT90 values between different nights led to 50% (10/20) of the included patients changing categories between desaturator and nondesaturator over the course of 1 week. Conclusions: Continuous SpO(2) measurements with wearable finger pulse oximeters identified significant SpO(2 )fluctuations between and within multiple days and nights of patients with COPD. Continuous SpO(2 ) measurements during daily home routines of patients with COPD generally had high amounts of valid data, except for motion artifacts during MVPA. The identified fluctuations can have implications for telemonitoring applications that are based on daily SpO(2) spot checks. CT90 values can vary greatly from night to night in patients with a nocturnal mean SpO(2) around 90%, indicating that these patients cannot be consistently categorized as desaturators or nondesaturators. We recommend using wearable sensors for continuous SpO(2 )measurements over longer time periods to determine the clinical relevance of the identified SpO(2 )fluctuations.
引用
收藏
页数:14
相关论文
共 43 条
[1]   Characterisation of COPD heterogeneity in the ECLIPSE cohort [J].
Agusti, Alvar ;
Calverley, Peter M. A. ;
Celli, Bartolome ;
Coxson, Harvey O. ;
Edwards, Lisa D. ;
Lomas, David A. ;
MacNee, William ;
Miller, Bruce E. ;
Rennard, Steve ;
Silverman, Edwin K. ;
Tal-Singer, Ruth ;
Wouters, Emiel ;
Yates, Julie C. ;
Vestbo, Jorgen .
RESPIRATORY RESEARCH, 2010, 11
[2]   Feature selection from nocturnal oximetry using genetic algorithms to assist in obstructive sleep apnoea diagnosis [J].
Alvarez, Daniel ;
Hornero, Roberto ;
Victor Marcos, J. ;
del Campo, Felix .
MEDICAL ENGINEERING & PHYSICS, 2012, 34 (08) :1049-1057
[3]   Pattern Analysis of Oxygen Saturation Variability in Healthy Individuals: Entropy of Pulse Oximetry Signals Carries Information about Mean Oxygen Saturation [J].
Bhogal, Amar S. ;
Mani, Ali R. .
FRONTIERS IN PHYSIOLOGY, 2017, 8
[4]   Oxygen saturation measurements in telemonitoring of patients with COPD: a systematic review [J].
Buekers, Joren ;
De Boever, Patrick ;
Vaes, Anouk W. ;
Aerts, Jean-Marie ;
Wouters, Emiel F. M. ;
Spruit, Martijn A. ;
Theunis, Jan .
EXPERT REVIEW OF RESPIRATORY MEDICINE, 2018, 12 (02) :113-123
[5]   Changes in telemonitored physiological variables and symptoms prior to exacerbations of chronic obstructive pulmonary disease [J].
Burton, Christopher ;
Pinnock, Hilary ;
McKinstry, Brian .
JOURNAL OF TELEMEDICINE AND TELECARE, 2015, 21 (01) :29-36
[6]  
Casanova C, 2006, RESP CARE, V51, P1416
[7]   Sleep-related O-2 desaturation and daytime pulmonary haemodynamics in COPD patients with mild hypoxaemia [J].
Chaouat, A ;
Weitzenblum, E ;
Kessler, R ;
Charpentier, C ;
Ehrhart, M ;
LeviValensi, P ;
Zielinski, J ;
Delaunois, L ;
Cornudella, R ;
dosSantos, JM .
EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (08) :1730-1735
[8]  
de Jesus JAL, 2011, IEEE ENG MED BIO, P1875, DOI 10.1109/IEMBS.2011.6090532
[9]   Psychophysical parameters of a multidimensional pain scale in newborns [J].
de Oliveira, M. V. M. ;
de Jesus, J. A. L. ;
Tristao, R. M. .
PHYSIOLOGICAL MEASUREMENT, 2012, 33 (01) :39-49
[10]  
Fussell Kevin M, 2003, Respir Care, V48, P115