Evaluation of Polarization Effects of e- Collection Schottky CdTe Medipix3RX Hybrid Pixel Detector

被引:14
作者
Astromskas, Vytautas [1 ]
Gimenez, Eva N. [2 ]
Lohstroh, Annika [1 ]
Tartoni, Nicola [2 ]
机构
[1] Univ Surrey, Guildford GU2 7XH, Surrey, England
[2] Diamond Light Source Ltd, Didcot OX11 0DE, Oxon, England
关键词
Medipix; polarization; Schottky CdTe; semiconductor radiation detectors; tri-phase pixels;
D O I
10.1109/TNS.2016.2516827
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper focuses on the evaluation of operational conditions such as temperature, exposure time and flux on the polarization of a Schottky electron collection CdTe detector. A Schottky e(-) collection CdTe Medipix3RX hybrid pixel detector was developed as a part of the CALIPSO-HIZPAD2 EU project. The 128 X 128 pixel matrix and 0.75 mm thick CdTe sensor bump-bonded to Medipix3RX readout chips enabled the study of the polarization effects. Single and quad module Medipix3RX chips were used which had 128 X 128 and 256 X 256 pixel matrices, respectively. This study reports the sensor-level and pixel-level polarization effects of the detector obtained from a laboratory X-ray source. We report that the sensor-level polarization is highly dependent on temperature, flux and exposure time. Furthermore, the study of pixel-level polarization effects led to identification of a new type of pixel behaviour that is characterised by three distinct phases and, thus, named "tri-phase" (3-P) pixels. The 3-P pixels were the dominant cause of degradation of the flat-field image uniformity under high flux operation. A new method of identifying the optimum operational conditions that utilises a criterion related to the 3-P pixels is proposed. A generated optimum operational conditions chart under the new method is reported. The criterion is used for bias voltage reset depolarization of the detector. The method successfully represented the dependency of polarization on temperature, flux and exposure time and was reproducible for multiple sensors. Operating the detector under the 3-P pixel criterion resulted in the total efficiency not falling below 95%.
引用
收藏
页码:252 / 258
页数:7
相关论文
共 9 条
  • [1] Pixel sensitivity variations in a CdTe-Medipix2 detector using poly-energetic x-rays
    Aamir, R.
    Lansley, S. P.
    Zainon, R.
    Fiederle, M.
    Fauler, A.
    Greiffenberg, D.
    Butler, P. H.
    Butler, A. P. H.
    [J]. JOURNAL OF INSTRUMENTATION, 2011, 6
  • [2] The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging
    Ballabriga, R.
    Alozy, J.
    Blaj, G.
    Campbell, M.
    Fiederle, M.
    Frojdh, E.
    Heijne, E. H. M.
    Llopart, X.
    Pichotka, M.
    Procz, S.
    Tlustos, L.
    Wong, W.
    [J]. JOURNAL OF INSTRUMENTATION, 2013, 8
  • [3] TIME-DEPENDENT POLARIZATION OF CDTE GAMMA-RAY DETECTORS
    BELL, RO
    ENTINE, G
    SERREZE, HB
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1974, 117 (01): : 267 - 271
  • [4] Comparison of three types of XPAD3.2/CdTe single chip hybrids for hard X-ray applications in material science and biomedical imaging
    Buton, C.
    Dawiec, A.
    Graber-Bolis, J.
    Arnaud, K.
    Berar, J. F.
    Blanc, N.
    Boudet, N.
    Clemens, J. C.
    Debarbieux, F.
    Delpierre, P.
    Dinkespiler, B.
    Gastaldi, T.
    Hustache, S.
    Morel, C.
    Pangaud, P.
    Perez-Ponce, H.
    Vigeolas, E.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 758 : 44 - 56
  • [5] Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical Applications
    Del Sordo, Stefano
    Abbene, Leonardo
    Caroli, Ezio
    Mancini, Anna Maria
    Zappettini, Andrea
    Ubertini, Pietro
    [J]. SENSORS, 2009, 9 (05) : 3491 - 3526
  • [6] Farella I., 2008, P IEEE NUCL SCI S OC, V56, P154, DOI DOI 10.1109/NSSMIC.2008.4775144
  • [7] Stability issues of high-energy resolution diode type CdTe nuclear radiation detectors in a long-term operation
    Niraula, M
    Nakamura, A
    Aoki, T
    Tomita, Y
    Hatanaka, Y
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 491 (1-2) : 168 - 175
  • [8] Merlin: a fast versatile readout system for Medipix3
    Plackett, R.
    Horswell, I.
    Gimenez, E. N.
    Marchal, J.
    Omar, D.
    Tartoni, N.
    [J]. JOURNAL OF INSTRUMENTATION, 2013, 8
  • [9] Improvement of the Productivity in the THM Growth of CdTe Single Crystal as Nuclear Radiation Detector
    Shiraki, Hiroyuki
    Funaki, Minoru
    Ando, Yukio
    Kominami, Shinya
    Amemiya, Kensuke
    Ohno, Ryoichi
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 395 - 399