Design and development of a new acetabular cup prosthesis

被引:7
作者
Mathias, Martin J. [1 ]
Tabeshfar, Kamran [1 ]
机构
[1] Bournemouth Univ, Bournemouth BH1 3NA, Dorset, England
来源
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS | 2006年 / 26卷 / 08期
基金
英国工程与自然科学研究理事会;
关键词
acetabular cup; fatigue; prosthesis; finite element; ceramics; HIP; WEAR; ARTHROPLASTY; AGE;
D O I
10.1016/j.msec.2005.08.003
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Historically, the main objective of Total Hip Replacements (THR) was to relieve pain and increase quality of life in the elderly and not normally intended for young active patients. With the current trend of performing THR on younger patients, the requirement is not only the increased longevity of the prosthesis but also increased performance for more strenuous activities. The acetabular cup design has far more impact on long-term survival of the THR than the femoral component. Optimising the acetabular cup produces a complex problem where the individual design factors have massive impact on the system. Changes in stress values, even caused by initial primary fixation during the operation, can result in stresses being transferred unfavourably. For optimisation of the acetabular cup, the properties of the natural hip must be retained by minimising both remodelling and bone resorption. This work has produced a novel acetabular cup prostheses with a ceramic on ceramic bearing surface that should last longer and perform better, thus reducing the necessity for costly and debilitating revisions. In addition, the use of the proposed novel anatomically orientated mechanical testing methods gives fast results and can also allow further insight into fatigue of alumina bearing couples in THR. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:1428 / 1433
页数:6
相关论文
共 30 条
  • [1] Limitations of artificial hip joint mobility due to wear and ceramic cup design
    Bader, R
    Steinhauser, E
    Willmann, G
    Gradinger, R
    [J]. BIOCERAMICS, 2000, 192-1 : 549 - 552
  • [2] HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 969 - 990
  • [3] Bizot P, 2000, CLIN ORTHOP RELAT R, P68
  • [4] BOBYN JD, 1994, CLIN ORTHOP RELAT R, V298, P27
  • [5] BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
  • [6] *BRIT STAND I, 1994, 6474 ISO BRIT STAND
  • [7] *BSI, 1993, 72067 ISO BRIT STAND
  • [8] *BSL, 1998, 7206 ISO BRIT STAND
  • [9] High strength and toughness alumina matrix composites by transformation toughening and 'in situ' platelet reinforcement (ZPTA) - The new generation of bioceramics
    Burger, W
    Richter, HG
    [J]. BIOCERAMICS, 2000, 192-1 : 545 - 548
  • [10] CHRISTEL PS, 1992, CLIN ORTHOP RELAT R, P10