ESTIMATION OF CENTRAL AORTIC BLOOD PRESSURE FROM NON-INVASIVE CUFF PRESSURE OSCILLATION SIGNALS VIA SYSTEM IDENTIFICATION

被引:0
作者
Ghasemi, Zahra [1 ]
Kim, Chang-Sei [1 ]
Ginsberg, Eric [2 ]
Duell, John [2 ]
Gupta, Anuj [2 ]
Hahn, Jin-Oh [1 ]
机构
[1] Univ Maryland, College Pk, MD 20742 USA
[2] Univ Maryland, Baltimore, MD 21201 USA
来源
PROCEEDINGS OF THE ASME 9TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2016, VOL 1 | 2017年
基金
美国国家科学基金会;
关键词
WAVE-FORM; RADIAL TONOMETRY; RECONSTRUCTION; HYPERTENSION; VALIDATION; EXERCISE;
D O I
暂无
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
This paper presents a model-based system identification approach to estimation of central aortic blood pressure waveform from non-invasive cuff pressure oscillation signals. First, we developed a mathematical model that can reproduce the relationship between central aortic blood pressure waveform and non-invasive cuff pressure oscillation signals at diametric locations by combining models to represent wave propagation in the artery, arterial pressure-volume relationship, and mechanics of the occlusive cuff. Second, we formulated the problem of estimating central aortic blood pressure waveform from non-invasive cuff pressure oscillation signals into a system identification problem. Third, we showed the proof-of-concept of the approach using simulated central aortic blood pressure waveform and cuff pressure oscillation signals. Finally, we illustrated the feasibility of the approach using central aortic blood pressure waveform and cuff pressure oscillation signals collected from a human subject. We showed that the proposed approach could estimate central aortic blood pressure waveform with accuracy: the root-mean-squared error associated with the central aortic blood pressure waveform was 1.7 mmHg (amounting to 1.6 % of the underlying mean blood pressure) while the errors associated with central aortic systolic and pulse pressures were -0.4 mmHg and -1.5 mmHg (amounting to -0.3 % and -1.4 % of the underlying mean blood pressure).
引用
收藏
页数:7
相关论文
共 26 条
  • [1] Abutaleb A.S., 2010, Math. Probl. Eng, V2010, P1
  • [2] Abutaleb A. S., 2010, MATH PROBL ENG, V2010, P27
  • [3] Chen CH, 1997, CIRCULATION, V95, P1827
  • [4] MECHANICS OF THE OCCLUSIVE ARM CUFF AND ITS APPLICATION AS A VOLUME SENSOR
    DRZEWIECKI, G
    BANSAL, V
    KARAM, E
    HOOD, R
    APPLE, H
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (07) : 704 - 708
  • [5] THEORY OF THE OSCILLOMETRIC MAXIMUM AND THE SYSTOLIC AND DIASTOLIC DETECTION RATIOS
    DRZEWIECKI, G
    HOOD, R
    APPLE, H
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) : 88 - 96
  • [6] Subject-specific estimation of central aortic blood pressure via system identification: preliminary in-human experimental study
    Fazeli, Nima
    Kim, Chang-Sei
    Rashedi, Mohammad
    Chappell, Alyssa
    Wang, Shaohua
    MacArthur, Roderick
    McMurtry, M. Sean
    Finegan, Barry
    Hahn, Jin-Oh
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2014, 52 (10) : 895 - 904
  • [7] Effects of a fixed-dose ACE inhibitor-diuretic combination on ambulatory blood pressure and arterial properties in isolated systolic hypertension
    Ferguson, Joanne M.
    Minas, Jack
    Siapantas, Siky
    Komesaroff, Paul A.
    Sudhir, Krishnankutty
    [J]. JOURNAL OF CARDIOVASCULAR PHARMACOLOGY, 2008, 51 (06) : 590 - 595
  • [8] Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometry
    Fetics, B
    Nevo, E
    Chen, CH
    Kass, DA
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (06) : 698 - 706
  • [9] Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform
    Gallagher, D
    Adji, A
    O'Rourke, MF
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2004, 17 (11) : 1059 - 1067
  • [10] Modeling and 2-sensor blind identification of human cardiovascular system
    Hahn, Jin-Oh
    Reisner, Andrew T.
    Asada, H. Harry
    [J]. CONTROL ENGINEERING PRACTICE, 2009, 17 (11) : 1318 - 1328