Economical helium bath cryopump:: design and testing

被引:0
作者
Musilová, V [1 ]
Dupák, J [1 ]
Hanzelka, P [1 ]
Králík, T [1 ]
Urban, P [1 ]
机构
[1] Acad Sci Czech Republ, Inst Sci Instruments, CS-61264 Brno, Czech Republic
关键词
cryopump; molecular flow; radiative heat transfer; cryogenic optimisation; Monte Carlo particle tracing;
D O I
10.1016/j.vacuum.2003.12.053
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A small UHV helium bath cryopump with low cryoliquid consumption has been designed, manufactured and tested. The cryopump is designed to keep UHV in electron optical devices without generation of disturbing electromagnetic fields or vibrations. The outer volume of its shell is 151, the filling volumes of liquid helium (LHe) and liquid nitrogen (LN2) are 3.01 and 3.41, respectively. Operating times between cryoliquid fillings are 35 days for LHe and 6 days for LN2 A pump speed of 251/s was measured for the He gas. A lowest pressure of 1.5 x 10(-7) Pa was achieved in a test chamber and the pumping tests indicated that an even lower pressure would be achieved in a well-degassed chamber. Good agreement between the calculated and measured values both of the pump speed and the cryoliquid evaporation has been found. The system of chevron-type baffles was designed and checked by Monte Carlo simulations of molecular flow and radiative heat transfer. The pump prototype was successfully used to evacuate an operating cryostat of a NMR spectrometer. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:77 / 83
页数:7