Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [41] Agreement of Corneal Diameter Measurements Obtained by a Swept-source Biometer and a Scheimpflug-based Topographer
    Salouti, Ramin
    Nowroozzadeh, Mohammad Hossein
    Tajbakhsh, Zahra
    Bagheri, Mansoureh
    Ghoreyshi, Maryam
    Azizzadeh, Monir
    Razeghinejad, Mohammad Reza
    CORNEA, 2017, 36 (11) : 1373 - 1376
  • [42] Agreement between swept-source OCT and Scheimpflug biometers in ocular measurements
    Mclintock, Cameron
    Mckelvie, James
    Niyazmand, Hamed
    Uprety, Samir
    BMC OPHTHALMOLOGY, 2025, 25 (01)
  • [43] Comparison of Central Corneal Thickness Measurements with a Rotating Scheimpflug Camera, a Specular Microscope, Optical Low-Coherence Reflectometry, and Ultrasound Pachymetry in Keratoconic Eyes
    Cinar, Yasin
    Cingu, Abdullah Kursat
    Turkcu, Fatih Mehmet
    Cinar, Tuba
    Sahin, Alparslan
    Yuksel, Harun
    Ari, Seyhmus
    SEMINARS IN OPHTHALMOLOGY, 2015, 30 (02) : 105 - 111
  • [44] Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices
    Feng, Yue
    Nitter, Tore
    Bertelsen, Geir
    Stojanovic, Aleksandar
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 52 (08) : 800 - 810
  • [45] Corneal densitometry measurements comparison between anterior segment OCT and scheimpflug imaging
    Lupardi, Enrico
    Moramarco, Antonio
    Cassini, Federico
    Febbraro, Simone
    Savini, Giacomo
    Fontana, Luigi
    INTERNATIONAL OPHTHALMOLOGY, 2024, 44 (01)
  • [46] Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth
    Bedei, Andrea
    Appolloni, Ilaria
    Madesani, Antonio
    Pietrelli, Alessia
    Franceschi, Sara
    Barabesi, Lucio
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2012, 22 : S29 - S32
  • [47] Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
    Mohammad Naderan
    Saeed Shoar
    Morteza Naderan
    Mohammad Amin Kamaleddin
    Mohammad Taher Rajabi
    International Journal of Ophthalmology, 2015, (02) : 275 - 280
  • [48] Repeatability, Reproducibility, and Agreement Between Three Different Scheimpflug Systems in Measuring Corneal and Anterior Segment Biometry
    Hernandez-Camarena, Julio C.
    Chirinos-Saldana, Patricia
    Navas, Alejandro
    Ramirez-Miranda, Arturo
    de la Mota, Angie
    Jimenez-Corona, Aida
    Graue-Hernandez, Enrique O.
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (09) : 616 - U44
  • [49] Agreement between optical coherence and Scheimpflug tomography: Vault measurements and reproducibility after implantable collamer lens implantation
    Almorin-Fernandez-Vigo, I.
    Sanchez-Guillen, I.
    Fernandez-Vigo, J. I.
    De-Pablo-Gomez-De-Liano, L.
    Kudsieh, B.
    Fernandez-Vigo, J. A.
    Macarro-Merino, A.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2021, 44 (09): : 1370 - 1380
  • [50] Comparison of corneal astigmatism measurements of 2 optical biometer models for tonic intraocular lens selection
    Srivannaboon, Sabong
    Chirapapaisan, Chareenun
    Chonpimai, Pratuangsri
    Koodkaew, Sunisa
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (02) : 364 - 371