Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [31] Anterior Surface-Based Keratometry Compared With Scheimpflug Tomography-Based Total Corneal Astigmatism
    Tonn, Bastian
    Klaproth, Oliver Klaus
    Kohnen, Thomas
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (01) : 291 - 298
  • [32] Factors affecting central corneal thickness measurement agreement between Scheimpflug imaging and ultrasound pachymetry in keratoconus
    Sorkin, Nir
    Ohri, Anamika
    Jung, Hyejung
    Haines, Lacey
    Sorbara, Luigina
    Mimouni, Michael
    Singal, Neera
    Hatch, Wendy
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2021, 105 (10) : 1371 - 1375
  • [33] Analysis of differences between keratometric astigmatism and total corneal astigmatism measured by IOLMaster 700
    Chunxia Yu
    Xiaoling Fang
    Jinhua Tao
    Yiping Dong
    Shenyu Ben
    Yulan Wang
    International Ophthalmology, 45 (1)
  • [34] Corneal higher-order aberrations measurements: Comparison between Scheimpflug and dual Scheimpflug-Placido technology in keratoconic eyes
    Piccinini, Andre L.
    Golan, Oren
    Torres-Netto, Emilio A.
    Hafezi, Farhad
    Randleman, J. Bradley
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (07) : 985 - 991
  • [35] A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug plus Placido topographer and rotating prism auto-keratorefractor
    Prakash, Gaurav
    Srivastava, Dhruv
    Choudhuri, Sounak
    INTERNATIONAL OPHTHALMOLOGY, 2015, 35 (06) : 869 - 880
  • [36] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    EYE AND VISION, 2022, 9 (01)
  • [37] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Robert Herber
    Janine Lenk
    Lutz E. Pillunat
    Frederik Raiskup
    Eye and Vision, 9
  • [38] Agreement Between Swept-Source Optical Biometry and Scheimpflug-based Topography Measurements of Anterior Segment Parameters
    Ozyol, Pelin
    Ozyol, Erhan
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 169 : 73 - 78
  • [39] Posterior and total corneal astigmatism measured with optical coherence tomography-based biometer and dual Scheimpflug analyzer
    Wang, Li
    Cao, Danmin
    Vilar, Cesar
    Koch, Douglas D.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (12) : 1652 - 1658
  • [40] Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer
    Kim, Eric J.
    de Oca, Ildamaris Montes
    Wang, Li
    Weikert, Mitchell P.
    Koch, Douglas D.
    Khandelwal, Sumitra S.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (12) : 2731 - 2738