Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
|
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [21] Effect of 2% fluorescein on Scheimpflug central corneal thickness measurements
    Briggs, Stella
    Bin Moammar, Marwa
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (02) : 239 - 242
  • [22] Placido disk-based topography versus high-resolution rotating Scheimpflug camera for corneal power measurements in keratoconic and post-LASIK eyes: reliability and agreement
    Penna, Rachele R.
    de Sanctis, Ugo
    Catalano, Martina
    Brusasco, Luca
    Grignolo, Federico M.
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2017, 10 (03) : 453 - 460
  • [23] The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging
    Zhao, Yune
    Chen, Ding
    Savini, Giacomo
    Wang, Qing
    Zhang, Hongfang
    Jin, Yili
    Song, Benhao
    Ning, Rui
    Huang, Jinhai
    Mei, Chenyang
    EYE AND VISION, 2020, 7 (01)
  • [24] Agreement between Corneal Thickness Measurements Using Pentacam Scheimpflug Camera, Noncontact Specular Microscopy, and Ultrasonographic Pachymetry in Diabetic Patients
    Altay, Yesim
    Balta, Ozgur
    Demirok, Gulizar
    Burcu, Ayse
    Balta, Oyku Bezen
    Ornek, Firdevs
    CURRENT EYE RESEARCH, 2017, 42 (02) : 187 - 194
  • [25] The effect of posterior corneal flat meridian and astigmatism amount on the total corneal astigmatism estimated from anterior corneal measurements
    Eom, Youngsub
    Kang, Su-Yeon
    Kim, Hyo Myung
    Song, Jong Suk
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 252 (11) : 1769 - 1777
  • [26] Comparison of Multicolored Spot Reflection Topographer and Scheimpflug-Placido System in Corneal Power and Astigmatism Measurements With Normal and Post-refractive Patients
    Lu, Weicong
    Miao, Yaxin
    Li, Yue
    Hu, Xiuli
    Hu, Qingjie
    Huang, Jinhai
    JOURNAL OF REFRACTIVE SURGERY, 2019, 35 (06) : 370 - +
  • [27] Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients
    Asawaworarit, Rosepon
    Satitpitakul, Vannarut
    Taweekitikul, Parichart
    Pongpirui, Krit
    PLOS ONE, 2022, 17 (05):
  • [28] Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System
    Zhang, Yaqin
    Dong, Jing
    Zhang, Suhua
    Sun, Bin
    Wang, Xiaoliang
    Tang, Maolong
    Wang, Xiaogang
    JOURNAL OF OPHTHALMOLOGY, 2020, 2020
  • [29] Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam
    Li, Xiaochun
    Cao, Xiaoguang
    Bao, Yongzhen
    BIOMED RESEARCH INTERNATIONAL, 2022, 2022
  • [30] Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    Kawamorita, Takushi
    Uozato, Hiroshi
    Kamiya, Kazutaka
    Bax, Leon
    Tsutsui, Kenta
    Aizawa, Daisuke
    Shimizu, Kimiya
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) : 127 - 133