Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras

被引:2
|
作者
Savini, Giacomo [1 ]
Noeser, Kristian [2 ]
Schiano-Lomoriello, Domenico [1 ]
Ducoli, Pietro [1 ]
机构
[1] Fdn GB Bietti, IRCCS, Via Livenza 3, I-00198 Rome, Italy
[2] Reg Hosp Randers, Dept Ophthalmol, Randers, Denmark
关键词
corneal astigmatism; Scheimpflug camera; vector analysis; TORIC INTRAOCULAR LENSES; POWER CALCULATION; CATARACT-SURGERY; REPEATABILITY; TOPOGRAPHY; REPRODUCIBILITY; TOMOGRAPHY; PENTACAM; SYSTEMS;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To investigate agreement between rotating Scheimpflug camera (Pentacam HR, Oculus) and rotating Scheimpflug camera combined with Placido disc corneal topography (Sirius, CSO) in measuring total corneal astigmatism (TCA). Methods: In this observational study, all patients undergoing cataract surgery with preoperative measurement of TCA by both devices and a validated corneal topographer (Keratron, Optikon 2000) were retrospectively evaluated. Astigmatism analysis was performed with and without vector analysis separately in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Vector analysis was performed using the Naeser polar system. Results: In 130 eyes of 130 subjects, nonvectorial analysis revealed that the mean TCA values obtained with the Sirius were higher than the corresponding values given by the Pentacam HR in all subgroups, although the difference was statistically significant only in eyes with against-the-rule astigmatism (P = 0.0009). This finding was confirmed by vector analysis. A TCA magnitude difference greater than 0.5 diopters was observed in 20.8% of cases, and a TCA axis difference greater than 10 degrees was observed in 45.4% of cases. Axis differences dropped to 18.5% when only eyes with astigmatism >0.75 diopters were analyzed and 3 measurements were averaged. The mean difference in the meridional and torsional power of TCA was close to zero in all subgroups, but with relatively large standard deviations (approximately 0.5 D). Conclusions: Agreement between both devices in measuring TCA is only moderate with respect to both magnitude and axis orientation.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 50 条
  • [1] Agreement Between Placido Topography and Scheimpflug Tomography for Corneal Astigmatism Assessment
    Delrivo, Marianella
    Ruisenor Vazquez, Pablo R.
    David Galletti, Jonatan
    Garibotto, Martin
    Fuentes Bonthoux, Fernando
    Pfoertner, Tomas
    Gaston Galletti, Jeremias
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (01) : 49 - 53
  • [2] Analysis of total corneal astigmatism with a rotating Scheimpflug camera in keratoconus
    Kim, Jinho
    Whang, Woong-Joo
    Kim, Hyun-Seung
    BMC OPHTHALMOLOGY, 2020, 20 (01)
  • [3] Agreement Between Internal Astigmatism and Posterior Corneal Astigmatism in Pseudophakic Eyes
    Feizi, Sepehr
    Delfazayebaher, Siamak
    Javadi, Mohammad Ali
    JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (06) : 379 - +
  • [4] Accuracy of Total Corneal Astigmatism Measurements With a Scheimpflug Imager and a Color Light-Emitting Diode Corneal Topographer
    Klijn, Stijn
    Reus, Nicolaas J.
    van der Sommen, Charlotte M.
    Sicam, Victor Arni D. P.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 167 : 72 - 78
  • [5] Measurements of Anterior and Posterior Corneal Curvatures with OCT and Scheimpflug Biometers in Patients with Low Total Corneal Astigmatism
    Muzyka-Wozniak, Maria
    Oleszko, Adam
    Grzybowski, Andrzej
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (23)
  • [6] Precision of Corneal Thickness Measurements Obtained Using the Scheimpflug-Placido Imaging and Agreement with Ultrasound Pachymetry
    Huang, Jinhai
    Savini, Giacomo
    Wang, Chengfang
    Lu, Weicong
    Gao, Rongrong
    Li, Yuanguang
    Wang, Qinmei
    Zhao, Yune
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [7] Agreement of corneal measurements between dual rotating Scheimpflug-Placido system and Placido-based topography device in normal and keratoconus eyes
    Ortiz-Toquero, Sara
    Zuniga, Virginia
    Rodriguez, Guadalupe
    de Juan, Victoria
    Martin, Raul
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) : 1198 - 1206
  • [8] Agreement and reliability in measuring central corneal thickness with a rotating Scheimpflug-Placido system and ultrasound pachymetry
    Maresca, Nunzio
    Zeri, Fabrizio
    Palumbo, Paolo
    Calossi, Antonio
    CONTACT LENS & ANTERIOR EYE, 2014, 37 (06) : 442 - 446
  • [9] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Savini, Giacomo
    Taroni, Leonardo
    Schiano-Lomoriello, Domenico
    Hoffer, Kenneth J.
    EYE, 2021, 35 (01) : 307 - 315
  • [10] Agreement of Central Corneal Thickness Measurements between Scheimpflug Photography and Optical Low-Coherence Reflectometry in Children
    Hashemi, Hassan
    Nabovati, Payam
    Khabazkhoob, Mehdi
    Emamian, Mohammad Hassan
    Yekta, Abbasali
    Fotouhi, Akbar
    SEMINARS IN OPHTHALMOLOGY, 2020, 35 (04) : 252 - 256