Home sphygmomanometers: validation versus accuracy

被引:42
作者
Akpolat, Tekin [1 ]
Dilek, Melda
Aydogdu, Turkan
Adibelli, Zelal
Erdem, Dilek Gurgenyatagi [2 ]
Erdem, Emre [2 ]
机构
[1] Ondokuz Mayis Univ, Sch Med, Dept Nephrol, Tip Fak,Nefrol Bilim Dali, TR-55139 Samsun, Turkey
[2] Ondokuz Mayis Univ, Sch Med, Dept Internal Med, TR-55139 Samsun, Turkey
关键词
accuracy; blood pressure; home sphygmomanometers; hypertension; validation; BLOOD-PRESSURE-MEASUREMENT; EDUCATION-PROGRAM RECOMMENDATIONS; EUROPEAN-SOCIETY; HYPERTENSION; MONITORS; DEVICES; PREVALENCE; MANAGEMENT;
D O I
10.1097/MBP.0b013e3283262f31
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The validation and accuracy of sphygmomanometers are important issues in the home sphygmomanometer market and in clinical practice. The aims of this study are to assess the frequency of validated home sphygmomanometers and to evaluate the relationship between the validation and accuracy of the devices. Methods Eight hundred and ninety-six home sphygmomanometers were brought by individuals to the University Hospital Hypertension Clinic for this study, and 870 (97%) of the devices were in adequate working condition and suitable for analysis. The relationship between accuracy and validation was investigated only in automated sphygmomanometers (n = 554). Results Both the manufacturer and model were known in 500 devices (90%), and these devices were eligible for evaluation of validation. We found 74 devices listed in the dabl Educational Trust website. Twenty-two of the 74 (30%) devices were validated. Four hundred (72%) of the 554 automated sphygmomanometers were inaccurate. The frequency of accuracy was higher among validated devices compared with nonvalidated devices (68 vs. 15%) (P < 0.01). Conclusion Our study showed that the frequency of accurate devices was higher among validated sphygmomanometers compared with nonvalidated sphygmomanometers. To our knowledge, our study is the first to demonstrate the relationship between accuracy and validation of home sphygmomanometers. The frequency of device-related errors can be decreased by training patients and supervising the blood pressure device market. Nonprofit organizations can help patients to overcome some of the problems in the blood pressure device market. Blood Press Monit 14:26-31 (C) 2009 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:26 / 31
页数:6
相关论文
共 25 条
  • [1] Sphygmomanometer calibration: a survey of one inner-city primary care group
    Ashworth, M
    Gordon, K
    Baker, G
    Deshmukh, A
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (04) : 259 - 262
  • [2] *ASS ADV MED INSTR, 1993, ANSI AAMI SP, V10
  • [3] The dabl educational trust device equivalence procedure
    Atkins, Neil
    O'Brien, Eoin
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (04) : 245 - 249
  • [4] CAMPBELL NRC, 1990, CAN MED ASSOC J, V143, P19
  • [5] Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure
    Chobanian, AV
    Bakris, GL
    Black, HR
    Cushman, WC
    Green, LA
    Izzo, JL
    Jones, DW
    Materson, BJ
    Oparil, S
    Wright, JT
    Roccella, EJ
    [J]. HYPERTENSION, 2003, 42 (06) : 1206 - 1252
  • [6] Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care
    Coleman, AJ
    Steel, SD
    Ashworth, M
    Vowler, SL
    Shennan, A
    [J]. BLOOD PRESSURE MONITORING, 2005, 10 (04) : 181 - 188
  • [7] Prevalence of home blood pressure measurement among selected hypertensive patients: Results of a multicenter survey from six hospital outpatient hypertension clinics in Italy
    Cuspidi, C
    Meani, S
    Lonati, L
    Fusi, V
    Magnaghi, G
    Garavelli, G
    Palumbo, G
    Pini, C
    Vaccarella, A
    Parati, G
    Leonetti, G
    Zanchetti, A
    [J]. BLOOD PRESSURE, 2005, 14 (04) : 251 - 256
  • [8] Self-measurement of blood pressure at home: Is it reliable?
    Dilek, Melda
    Adibelli, Zelal
    Aydogdu, Turkan
    Koksal, Ali Riza
    Cakar, Burcu
    Akpolat, Tekin
    [J]. BLOOD PRESSURE, 2008, 17 (01) : 34 - 41
  • [9] Limitations of current validation protocols for home blood pressure monitors for individual patients
    Gerin, W
    Schwartz, AR
    Schwartz, JE
    Pickering, TG
    Davidson, KW
    Bress, J
    O'Brien, E
    Atkins, N
    [J]. BLOOD PRESSURE MONITORING, 2002, 7 (06) : 313 - 318
  • [10] A survey of validated automated home blood pressure monitors available for the Internet shopper
    Graves, JW
    [J]. BLOOD PRESSURE MONITORING, 2005, 10 (02) : 103 - 107