IMPLANTED CARDIAC DEFIBRILLATOR CARE IN RADIATION ONCOLOGY PATIENT POPULATION

被引:61
作者
Gelblum, Daphna Y. [1 ]
Amols, Howard [2 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, Commack, NY 11725 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Med Phys, Commack, NY 11725 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2009年 / 73卷 / 05期
关键词
Implanted cardiac defibrillator; Radiotherapy; Patient monitoring; CARDIOVERTER-DEFIBRILLATOR; PACEMAKERS;
D O I
10.1016/j.ijrobp.2008.06.1903
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To review the experience of a large cancer center with radiotherapy (RT) patients bearing implantable cardiac defibrillators (ICDs) to propose some preliminary care guidelines as we learn more about the devices and their interaction with the therapeutic radiation environment. Methods and Materials: We collected data on patients with implanted ICDs treated with RT during a 2.5-year period at any of the five Memorial Sloan-Kettering clinical campuses. Information regarding the model, location, and dose detected from the device, as well as the treatment fields, fraction size, and treatment energy was collected. During this time, a new management policy for these patients had been implemented requiring treatment with low-energy beams (6 MV) and close surveillance of the patients in partnership with their electrophysiologist, as they received RT. Results: During the study period, 33 patients were treated with an ICD in place. One patient experienced a default of the device to its initial factory setting that was detected by the patient hearing an auditory signal from the device. This patient had initially been treated with a 15-MV beam. After this episode, his treatment was replanned to be completed with 6-MV photons, and he experienced no further events. Conclusion: Patients with ICDs; and other implanted computer-controlled devices will be encountered more frequently in the RT department, and proper management is important. We present a policy for the safe treatment of these patients in the radiation oncology environment. (C) 2009 Elsevier Inc.
引用
收藏
页码:1525 / 1531
页数:7
相关论文
共 22 条
  • [1] Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure
    Bardy, GH
    Lee, KL
    Mark, DB
    Poole, JE
    Packer, DL
    Boineau, R
    Domanski, M
    Troutman, C
    Anderson, J
    Johnson, G
    McNulty, SE
    Clapp-Channing, N
    Davidson-Ray, LD
    Fraulo, ES
    Fishbein, DP
    Luceri, RM
    Ip, JH
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) : 225 - 237
  • [2] BRADLEY PD, 1998, IEEE T NUCL SCI
  • [3] Cost-effectiveness in a flat world - Can ICDs help the United States get rhythm?
    Goldman, L
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2005, 353 (14) : 1513 - 1515
  • [4] *GUID CORP CARD RH, 2006, GUID CO LIT RAD TOL
  • [5] PHOTONEUTRONS FROM MEDICAL LINEAR ACCELERATORS - RADIOBIOLOGICAL MEASUREMENTS AND RISK ESTIMATES
    HALL, EJ
    MARTIN, SG
    AMOLS, H
    HEI, TK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (01): : 225 - 230
  • [6] Implantable cardiac defibrillators may be damaged by radiation therapy
    Hoecht, S
    Rosenthal, P
    Sancar, D
    Behrens, S
    Hinkelbein, W
    Hoeller, U
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 2002, 20 (08) : 2212 - 2213
  • [7] Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators
    Hurkmans, CW
    Scheepers, E
    Springorum, BGF
    Uiterwaal, H
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (01): : 282 - 289
  • [8] Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy
    Kadish, A
    Dyer, A
    Daubert, JP
    Quigg, R
    Estes, NAM
    Anderson, KP
    Calkins, H
    Hoch, D
    Goldberger, J
    Shalaby, A
    Sanders, WE
    Schaechter, A
    Levine, JH
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (21) : 2151 - 2158
  • [9] Pacemaker and ICD generator reliability - Meta-analysis of device registries
    Maisel, WH
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16): : 1929 - 1934
  • [10] Pacemaker and ICD generator malfunctions - Analysis of food and drug administration annual reports
    Maisel, WH
    Moynahan, M
    Zuckerman, BD
    Gross, TP
    Tovar, OH
    Tillman, DB
    Schultz, DB
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 295 (16): : 1901 - 1906