Measurement of the micrometer diameters of focused laser-beam radiation with a thin-wire bolometer

被引:0
|
作者
Kuz'michev, VM
Safronov, BV
Pogorelov, SV
Balkashin, VP
机构
关键词
Radiation; Physical Chemistry; Analytical Chemistry; Laser Radiation; Systematic Error;
D O I
10.1023/A:1015929920117
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The profile distributions of a focused Gaussian beam of 63-mum laser radiation for E- and H-polarizations were recorded by a thin-wire bolometer 3 mum in diameter The profiles were used to determine the beam diameters, which were D-E = (8.16 +/- 0.12) mum and D-H = (8.00 +/- 0.12) mum upon elimination of the dominant systematic errors.
引用
收藏
页码:264 / 267