Simultaneous derivatization/preconcentration of volatile aldehydes with a miniaturized fiber-packed sample preparation device designed for gas chromatographic analysis

被引:54
作者
Saito, Yoshihiro [1 ]
Ueta, Ikuo [1 ]
Ogawa, Mitsuhiro [1 ]
Jinno, Kiyokatsu [1 ]
机构
[1] Toyohashi Univ Technol, Sch Mat Sci, Toyohashi, Aichi 4418580, Japan
基金
日本学术振兴会;
关键词
derivatization; extraction; fiber; GC; air/gases; aldehydes; SOLID-PHASE MICROEXTRACTION; ROD HETEROCYCLIC POLYMER; LIQUID-CHROMATOGRAPHY; DYNAMIC EXTRACTION; CAPILLARY-COLUMNS; FIBROUS MATERIALS; STATIONARY-PHASE; NEEDLE TRAP; SEPARATION; AIR;
D O I
10.1007/s00216-006-0509-z
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel in-needle sample preparation device has been developed for the determination of volatile aldehydes in gaseous samples. The needle device is designed for the gas chromatographic (GC) analysis of aldehydes and ketones commonly found in typical in-house environments. In order to prepare the extraction device, a bundle of polymer-coated filaments was longitudinally packed into a specially designed needle. Derivatization reactions were prompted by 2,4-dinitrophenylhydrazine (NDPH) included in the needle, and so the aldehydes and ketones were derivatized to the corresponding hydrazones and extracted with the extraction needle. A reproducible extraction needle preparation process was established, along with a repeatable derivatization/extraction process that ensures the successful determination of aldehydes. The storage performance of the extraction needle was also evaluated at room temperature for three days. The results demonstrate the successful application of the fiber-packed extraction device to the preparation of a gaseous sample of aldehydes, and the future possibility of applying the extraction device to the analysis of in-house environments.
引用
收藏
页码:725 / 732
页数:8
相关论文
共 44 条
[1]   Surface derivatization of poly(p-phenylene terephthalamide) fiber designed for novel separation and extraction media [J].
Abe, A ;
Saito, Y ;
Imaizumi, M ;
Ogawa, M ;
Takeichi, T ;
Jinno, K .
JOURNAL OF SEPARATION SCIENCE, 2005, 28 (17) :2413-2418
[2]   2,4-DINITROPHENYLHYDRAZINE-COATED SILICA-GEL CARTRIDGE METHOD FOR DETERMINATION OF FORMALDEHYDE IN AIR - IDENTIFICATION OF AN OZONE INTERFERENCE [J].
ARNTS, RR ;
TEJADA, SB .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1989, 23 (11) :1428-1430
[3]   SOLID-PHASE MICROEXTRACTION WITH THERMAL-DESORPTION USING FUSED-SILICA OPTICAL FIBERS [J].
ARTHUR, CL ;
PAWLISZYN, J .
ANALYTICAL CHEMISTRY, 1990, 62 (19) :2145-2148
[4]   Sorptive sample preparation - a review [J].
Baltussen, E ;
Cramers, CA ;
Sandra, PJF .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 373 (1-2) :3-22
[5]   Needle-type concentrator and its application to the determination of pollutants [J].
Berezkin, VG ;
Makarov, ED ;
Stolyarov, BV .
JOURNAL OF CHROMATOGRAPHY A, 2003, 985 (1-2) :63-65
[6]   Automated in-tube solid-phase microextraction coupled to high-performance liquid chromatography [J].
Eisert, R ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 1997, 69 (16) :3140-3147
[7]   Miniaturized solid-phase extraction with resin disks [J].
Fritz, JS ;
Masso, JJ .
JOURNAL OF CHROMATOGRAPHY A, 2001, 909 (01) :79-85
[8]  
Jinno K, 2001, ELECTROPHORESIS, V22, P3785, DOI 10.1002/1522-2683(200109)22:17<3785::AID-ELPS3785>3.0.CO
[9]  
2-U
[10]  
Jönsson JÅ, 2001, J SEP SCI, V24, P495, DOI 10.1002/1615-9314(20010801)24:7<495::AID-JSSC495>3.0.CO