A prospective, randomised comparison of single and three piece acrylic foldable intraocular lenses

被引:49
作者
Nejima, R
Miyata, K
Honbou, M
Tokunaga, T
Tanabe, T
Sato, M
Oshika, T
机构
[1] Univ Tsukuba, Inst Clin Med, Dept Ophthalmol, Tsukuba, Ibaraki 3058575, Japan
[2] Miyata Eye Hosp, Miyazaki, Japan
关键词
D O I
10.1136/bjo.2003.037663
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims: To compare the postoperative performance of single and three piece acrylic foldable intraocular lenses (IOLs). Methods: 20 patients underwent bilateral cataract surgery with a single piece SA30AL IOL in one eye and a three piece MA30BA IOL in the other eye. The eyes were randomly assigned to either a single or three piece lens. The amount of IOL decentration and tilt, area of anterior capsule opening, and degree of posterior capsule opacification were measured using the Scheimpflug anterior segment analysis system (Nidek EAS-1000). Visual acuity and contrast sensitivity were examined. Measurements were performed by masked examiners before and 1 day, 1 week, 1, 3, 6, and 18 months after surgery. Results: There were no significant differences between the two groups (p>0.05, paired t test) in the amount of IOL decentration, IOL tilt, area of anterior capsule opening, degree of posterior capsule opacification, best corrected visual acuity, and contrast sensitivity throughout the 18 month follow up period. Conclusion: The single and three piece acrylic foldable IOLs are equally stable in the eye after surgery.
引用
收藏
页码:746 / 749
页数:4
相关论文
共 35 条
  • [1] Quantification of posterior capsule opacification with round and sharp edge intraocular lenses
    Auffarth, GU
    Golescu, A
    Becker, KA
    Völcker, HE
    [J]. OPHTHALMOLOGY, 2003, 110 (04) : 772 - 780
  • [2] Preliminary results of cataract extraction with implantation of a single-piece AcrySof intraocular lens
    Caporossi, A
    Casprini, F
    Tosi, GM
    Baiocchi, S
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (04) : 652 - 655
  • [3] Comparison of AcrySof MA30BA and Sensar AR40 acrylic intraocular lenses
    Casprini, F
    Tosi, GM
    Quercioli, PP
    Caporossi, A
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) : 1130 - 1134
  • [4] Clinical performance of Alcon SA30AL and SA60AT single-piece acrylic intraocular lenses
    Davison, JA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07) : 1112 - 1123
  • [5] Scanning electron microscopic and histologic evaluation of the AcrySof SA30AL acrylic intraocular lens - Manufacturing quality and morphology in the capsular bag
    Escobar-Gomez, M
    Apple, DJ
    Vargas, LG
    Werner, L
    Arthur, SN
    Pandey, SK
    Izak, AM
    Schmidbauer, JM
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) : 164 - 169
  • [6] Relationship of AcrySof acrylic and PhacoFlex silicone intraocular lenses to visual acuity and posterior capsule opacification
    Halpern, MT
    Covert, D
    Battista, C
    Weinstein, AJ
    Levinson, RD
    Yan, L
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (04) : 662 - 669
  • [7] Anterior capsule contraction and intraocular lens dislocation after implant surgery in eyes with retinitis pigmentosa
    Hayashi, K
    Hayashi, H
    Matsuo, K
    Nakao, F
    Hayashi, F
    [J]. OPHTHALMOLOGY, 1998, 105 (07) : 1239 - 1243
  • [8] In vivo quantitative measurement of posterior capsule opacification after extracapsular cataract surgery
    Hayashi, K
    Hayashi, H
    Nakao, F
    Hayashi, F
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 125 (06) : 837 - 843
  • [9] Decentration and tilt of polymethyl methacrylate, silicone, and acrylic soft intraocular lenses
    Hayashi, K
    Harada, M
    Hayashi, H
    Nakao, F
    Hayashi, F
    [J]. OPHTHALMOLOGY, 1997, 104 (05) : 793 - 798
  • [10] Comparison of decentration and tilt between one piece and three piece polymethyl methacrylate intraocular lenses
    Hayashi, K
    Hayashi, H
    Nakao, F
    Hayashi, F
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (04) : 419 - 422