Randomized comparison of split tip versus step tip high-flow hemodialysis catheters

被引:54
作者
Trerotola, SO
Kraus, M
Shah, H
Namyslowski, J
Johnson, MS
Stecker, MS
Ahmad, I
McLennan, G
Patel, NH
O'Brien, E
Lane, KA
Ambrosius, WT
机构
[1] Indiana Univ, Sch Med, Dept Radiol, Indianapolis, IN 46202 USA
[2] Indiana Univ, Sch Med, Dept Med Nephrol, Indianapolis, IN 46202 USA
[3] Indiana Univ, Sch Med, Dept Med Biostat, Indianapolis, IN 46202 USA
关键词
dialysis catheters; catheterization; central venous access; thrombosis; fibrin sheaths; polyurethane catheter;
D O I
10.1046/j.1523-1755.2002.00416.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background. Our purpose was to compare the function and complications of two high-flow polyurethane hemodialysis catheters. Methods, This prospective, randomized trial compared the Ash-Split (MedComp) and Opti-Flow (Bard Access Systems) catheters. All patients referred for tunneled hemodialysis catheter placement were offered entry in the study, provided they met inclusion criteria. Catheters were placed by interventional radiologists using ultrasound and fluoroscopic guidance. Procedure time and initial complications were recorded. Effective (QbEff) catheter flow rates and recirculation were studied at baseline, one month, three and six months using ultrasonic dilution (Transonic) at various pump speeds (Qb). Episodes of catheter malfunction and infection were recorded. Catheter removal or six months was the study endpoint. Results. A total of 132 patients were enrolled in the trial. The groups did not differ as to age, sex distribution, height or weight (P > 0.05). Initial complications included kinking resulting in catheter failure (Optiflow N = 3), and tunnel bleeding (Optiflow N = 1: Ash N = 3). Adjusted mean flow rates (QbEff) at Qb300 were 299 mL/min Ash and 305 mL/min Optiflow (P = 0.06), at Qb400 were 365 mL/min Ash and 382 mL/min Optiflow (P = 0.01), and at QbMax were 414 mL/min Ash and 433 mL/min Optiflow (P = 0.03). Recirculation was significantly higher with the Optiflow catheter at most measurement points. Total late complications were lower in the Ash group (P = 0.04), and catheter survival was significantly higher in the Ash group (P = 0.02). Conclusions. Both catheters can deliver flow rates well beyond those recommended by the Dialysis Outcomes Quality Initiative. While the Optiflow delivered higher flow rates at some measurement points, this was offset by higher recirculation. The Ash catheter showed a long-term survival advantage and fewer late complications.
引用
收藏
页码:282 / 289
页数:8
相关论文
共 28 条
[1]  
[Anonymous], 2001, AM J KIDNEY DIS, V37, pS137, DOI DOI 10.1016/S0272-6386(01)70007-8
[2]   Adequacy of haemodialysis with cuffed central-vein catheters [J].
Atherikul, K ;
Schwab, SJ ;
Conlon, PJ .
NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (03) :745-749
[3]  
Beathard GA, 1999, J AM SOC NEPHROL, V10, P1045
[4]   INTERNAL JUGULAR-VEIN CANNULATION USING 2 SILASTIC CATHETERS - A NEW, SIMPLE AND SAFE LONG-TERM VASCULAR ACCESS FOR EXTRACORPOREAL TREATMENT [J].
CANAUD, B ;
BERAUD, JJ ;
JOYEUX, H ;
MION, C .
NEPHRON, 1986, 43 (02) :133-138
[5]   INFECTIOUS COMPLICATIONS OF INDWELLING LONG-TERM CENTRAL VENOUS CATHETERS [J].
CLARKE, DE ;
RAFFIN, TA .
CHEST, 1990, 97 (04) :966-972
[6]   Replacement of failing tunneled hemodialysis catheters through pre-existing subcutaneous tunnels: A comparison of catheter function and infection rates for de novo placements and over-the-wire exchanges [J].
Duszak, R ;
Haskal, ZJ ;
Thomas-Hawkins, C ;
Soulen, MC ;
Baum, RA ;
Shlansky-Goldberg, RD ;
Cope, C .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 9 (02) :321-327
[7]   Exchange of poorly functioning tunneled permanent hemodialysis catheters [J].
Garofalo, RS ;
Zaleski, GX ;
Lorenz, JM ;
Funaki, B ;
Rosenblum, JD ;
Leef, JA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (01) :155-158
[8]   Percutaneous fibrin sheath stripping versus transcatheter urokinase infusion for malfunctioning well-positioned tunneled central venous dialysis catheters: A prospective, randomized trial [J].
Gray, RJ ;
Levitin, A ;
Buck, D ;
Brown, LC ;
Sparling, YH ;
Jablonski, KA ;
Fessahaye, A ;
Gupta, AK .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (09) :1121-1129
[9]   Transvenous removal of fibrin sheaths from tunneled hemodialysis catheters [J].
Haskal, ZJ ;
Leen, VH ;
ThomasHawkins, C ;
ShlanskyGoldberg, RD ;
Baum, RA ;
Soulen, MC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (04) :513-517
[10]  
HERNANDEZ D, 1993, NEPHROL DIAL TRANSPL, V8, P227