Ceramic Materials as Bearing Surfaces for Total Hip Arthroplasty

被引:63
作者
D'Antonio, James A. [1 ]
Sutton, Kate [1 ]
机构
[1] Acad Resources, Stryker Orthopaed, Mahwah, NJ USA
关键词
FEMORAL-HEAD; POLYETHYLENE WEAR; ON-POLYETHYLENE; COBALT-CHROMIUM; SURVIVORSHIP; SIMULATOR; DAMAGE;
D O I
10.5435/00124635-200902000-00002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
During the past decade, advances in total hip arthroplasty component design have produced implants with reliable clinical results in regard to fixation. The foremost unresolved challenge has been the development of bearing surfaces that can withstand the higher demands of younger and more active patients. New alternative bearings with superior wear characteristics that minimize debris include ceramic-on-ceramic, metal-on-metal, and highly cross-linked polyethylenes in combination with ceramic or metal. Alumina-on-alumina ceramic bearings are extremely hard and scratch resistant and provide superior lubrication and wear resistance compared with other bearing surfaces in clinical use. Survivorship revision for any reason for the alumina ceramic bearings at 10 years was significantly higher compared with metal-on-polyethylene. Bearings currently being studied because of their encouraging wear performance in the laboratory are an alumina matrix (82% alumina, 17% zirconia, 0.3% chromium oxide), zirconium oxide, and ceramic-on-cobalt-chromium.
引用
收藏
页码:63 / 68
页数:6
相关论文
共 32 条
  • [11] DENNIS DA, 2006, BIOC ALT BEAR TOT JO
  • [12] A novel low wearing differential hardness, ceramic-on-metal hip joint prosthesis
    Firkins, PJ
    Tipper, JL
    Ingham, E
    Stone, MH
    Farrar, R
    Fisher, J
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (10) : 1291 - 1298
  • [13] Garino J., 2006, Semin Arthroplasty, V17, P113, DOI [10.1053/j.sart.2006.09.004, DOI 10.1053/J.SART.2006.09.004]
  • [14] Comparison of the cytotoxicity of clinically relevant cobalt-chromium and alumina ceramic wear particles in vitro
    Germain, MA
    Hatton, A
    Williams, S
    Matthews, JB
    Stone, MH
    Fisher, J
    Ingham, E
    [J]. BIOMATERIALS, 2003, 24 (03) : 469 - 479
  • [15] JAFFE WL, 2007, 74 ANN M P AM ASS OR, P418
  • [16] JARRETT CA, 2006, 16 ANN M AM ASS HIP, P20
  • [17] Damage of oxinium femoral heads subsequent to hip arthroplasty dislocation - Three retrieval case studies
    Kop, Alan M.
    Whitewood, Colin
    Johnston, Don J. L.
    [J]. JOURNAL OF ARTHROPLASTY, 2007, 22 (05) : 775 - 779
  • [18] Kuntz M., 2006, Semin Arthroplasty, V17, P141
  • [19] Zirconia femoral head fractures - A clinical and retrieval analysis
    Masonis, JL
    Bourne, RB
    Ries, MD
    McCalden, RW
    Salehi, A
    Kelman, DC
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (07) : 898 - 905
  • [20] Quantification of third-body damage and its effect on UHMWPE wear with different types of femoral head
    Minakawa, H
    Stone, MH
    Wroblewski, BM
    Lancaster, JG
    Ingham, E
    Fisher, J
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (05): : 894 - 899