New Bioabsorbable Septal Repair Implant for Percutaneous Closure of a Patent Foramen Ovale: Short-Term Results of a Single-Centre Experience

被引:10
作者
Van den Branden, B. J. L. [1 ]
Post, M. C. [1 ]
Jaarsma, W. [1 ]
ten Berg, J. M. [1 ]
Suttorp, M. J. [1 ]
机构
[1] St Antonius Hosp, Dept Cardiol, NL-3435 CM Nieuwegein, Netherlands
关键词
congenital heart defect; catheterization; efficacy; TRANSCATHETER CLOSURE; PARADOXICAL EMBOLISM; FOLLOW-UP; DEVICE; OCCLUDER; AMPLATZER; EVENTS; STROKE; RISK; COMPLICATIONS;
D O I
10.1002/ccd.21998
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Permanent implants for closure of a patent foramen ovale (PFO) have a number of possible disadvantages including erosions, thrombus formation, and allergic reactions. The incidence of adverse events may be lower using a bioabsorbable device. Objective: To evaluate the short-term safety and efficacy of a new bioabsorbable closure device. Methods: All 35 consecutive patients (21 female, mean age 47.9 +/- 10.8 years), who underwent a percutaneous PFO closure between November 2007 and July 2008, were included. All complications were reported. The efficacy was based on the residual shunting the day after implant and at 1 month follow-up and was graded as minimal, moderate, or severe, using contrast transthoracic echocardiography with the Valsalva manoeuvre. Results: The only in-hospital complication was a surgical device retrieval from the femoral vein. Four patients developed a minimal inguinal haematoma. One day after closure, residual shunting was present in 56% of the patients (minimal 27%, moderate 23% and severe 6%). At I month follow-up (n = 33), one patient developed a transient neurological deficit and three patients suffered from paroxysmal atrial fibrillation. A residual shunt at 1 month was present in 45% of the patients (minimal 30%, moderate 12%, and severe 3%). Conclusions: Percutaneous PFO closure using the bioabsorbable closure device seems to be safe. However, a high rate of residual shunting is present at 1 month follow-up. Long-term follow-up data are necessary to evaluate the efficacy and safety of this device. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:286 / 290
页数:5
相关论文
共 24 条
[1]  
Alaeddini Jamshid, 2006, J Invasive Cardiol, V18, P365
[2]   Erosion of amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk [J].
Amin, Z ;
Hijazi, ZM ;
Bass, JL ;
Cheatham, JP ;
Hellenbrand, WE ;
Kleinman, CS .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) :496-502
[3]   Transcatheter closure of patent foramen ovale (PFO) in patients with paradoxical embolism - Periprocedural safety and mid-term follow-up results of three different device occluder systems [J].
Braun, M ;
Gliech, V ;
Boscheri, A ;
Schoen, S ;
Gahn, G ;
Reichmann, H ;
Haass, M ;
Schraeder, R ;
Strasser, RH .
EUROPEAN HEART JOURNAL, 2004, 25 (05) :424-430
[4]   Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder [J].
Divekar, A ;
Gaamangwe, T ;
Shaikh, N ;
Raabe, M ;
Ducas, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (08) :1213-1218
[5]   Interventional atrial septal defect closure using a totally bioresorbable occluder matrix - Development and preclinical evaluation of the BioSTAR device [J].
Jux, Christian ;
Bertram, Harald ;
Wohlsein, Peter ;
Bruegmann, Michael ;
Paul, Thomas .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (01) :161-169
[6]  
Jux Christian, 2003, J Interv Cardiol, V16, P149, DOI 10.1046/j.1540-8183.2003.08027.x
[7]   Transcatheter closure versus medical therapy of patent foramen ovale and presumed paradoxical thromboemboli - A systematic review [J].
Khairy, P ;
O'Donnell, CP ;
Landzberg, MJ .
ANNALS OF INTERNAL MEDICINE, 2003, 139 (09) :753-760
[8]   Outcome after percutaneous closure of a patent foramen ovale using the Intrasept™ device:: A multi-centre study [J].
Luermans, Justin G. L. M. ;
Post, Martijn C. ;
Schraeder, Rainer ;
Sluysmans, Thierry ;
Vydt, Tom ;
Vermeersch, Paul ;
Chessa, Massimo ;
Onorato, Eustaquio ;
Goy, Jean-Jacques ;
Budts, Werner I. H. L. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 71 (06) :822-828
[9]   Percutaneous transcatheter closure of patent foramen ovale in patients with paradoxical embolism [J].
Martín, F ;
Sánchez, PL ;
Doherty, E ;
Colon-Hernandez, PJ ;
Delgado, G ;
Inglessis, I ;
Scott, N ;
Hung, J ;
King, MEE ;
Buonanno, F ;
Demirjian, Z ;
de Moor, M ;
Palacios, IF .
CIRCULATION, 2002, 106 (09) :1121-1126
[10]   RECURRENT CEREBROVASCULAR EVENTS IN PATIENTS WITH PATENT FORAMEN OVALE, ATRIAL SEPTAL ANEURYSM, OR BOTH AND CRYPTOGENIC STROKE OR TRANSIENT ISCHEMIC ATTACK [J].
MAS, JL ;
ZUBER, M .
AMERICAN HEART JOURNAL, 1995, 130 (05) :1083-1088