Occupational exposure to Am-Be neutron calibration source mounted in OB26 shielding container

被引:1
作者
Szewczak, Kamil [1 ]
Jednorog, Slawomir [2 ]
机构
[1] Cent Lab Radiol Protect, PL-03194 Warsaw, Poland
[2] EURATOM, Inst Plasma Phys & Laser Microfus, PL-01497 Warsaw, Poland
关键词
occupational exposure; active instruments measurements; neutron source calibration; IRRADIATOR; PURPOSES;
D O I
10.2478/nuka-2014-0016
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Laboratory for Dosimetric and Radon Instruments Calibration which is a part of Central Laboratory for Radiological Protection (CLRP) in Warsaw is equipped with Am-241-Be neutron calibration source with activity of 185 GBq since 1999. The capsule was mounted in the OB26 type shielding container. The control room is separated from the above source by a concrete wall of 0.5 m in thickness. The calibration hall is adjacent to one side of the office room. To comply with the requirements of the radiological protection system, the occupational exposure of persons that are working both in the office and control room needs to be assessed. Two methods were involved for ambient dose equivalent rate determination. The active instrument measurements (AIMs) performed with the Berthold LB6411 neutron probe and the Monte Carlo simulation method (MCS) based on MCNP5 code. These estimations were completed for five reference points. Additionally the gamma radiation component was measured by RSS131 ionisation chamber. An increased value of the ambient dose equivalent rate from neutrons was observed in two reference positions. The first observation was done in the control room while the second one in the office room. Expected individual dose equivalents were evaluated based on the results of the AIM and on the expected working time in particular reference points. The annual individual dose equivalent associated with calibration activities using mentioned neutron source was estimated at maximum 0.8 mSv.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 11 条
[1]  
Berthold Technologies, 1996, NEUTR PROB LB6411 OP
[2]  
Blizard E.P., 1958, RAD ATTENUATION CHAR
[3]  
International Organization for Standardization, 2001, 85291 ISO
[4]   Radionuclide neutron sources in calibration laboratory-neutron and gamma doses and their changes in time [J].
Jozefowicz, K. ;
Golnik, N. ;
Tulik, P. ;
Zielczynski, M. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :134-137
[5]   Standard fields of old neutron sources -: Parameters and traceability [J].
Józefowicz, K ;
Golnik, N ;
Zielczynski, M .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :107-110
[6]   NEUTRON FIELD CHARACTERISATION OF THE OB26 CRNA IRRADIATOR IN VIEW OF ITS USE FOR CALIBRATION PURPOSES [J].
Mazrou, Hakim ;
Sidahmed, Tassadit ;
Allab, Malika .
RADIATION PROTECTION DOSIMETRY, 2010, 141 (02) :114-126
[7]   Monte-Carlo investigation of radiation beam quality of the CRNA neutron irradiator for calibration purposes [J].
Mazrou, Hakim ;
Sidahmed, Tassadit ;
Allab, Malika .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (10) :1915-1921
[8]  
Mcconn Jr R.J., 2011, COMPENDIUM MAT COMPO
[9]  
Murata I., 2014, PROGR NUCL SCI TECHN, V4, P345, DOI DOI 10.15669/PNST.4.345
[10]   Neutron calibration facilities [J].
Schuhmacher, H .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :33-42