Concurrent Validity and Test-retest Reliability of the OPTOGait Photoelectric Cell System for the Assessment of Spatio-temporal Parameters of the Gait of Young Adults

被引:96
作者
Lee, Myung Mo [1 ]
Song, Chang Ho [1 ]
Lee, Kyoung Jin [1 ]
Jung, Sang Woo [1 ]
Shin, Doo Chul [1 ]
Shin, Seung Ho [1 ]
机构
[1] Sahmyook Univ, Coll Hlth Sci, Dept Phys Therapy, Seoul 139742, South Korea
关键词
Locomotion; Measurement; Analysis; GAITRITE(R) WALKWAY SYSTEM; INTERRATER RELIABILITY; QUANTIFICATION; PEOPLE; RATER;
D O I
10.1589/jpts.26.81
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
[Purpose] The purpose of this study was to investigate the concurrent validity and test-retest reliability recently introduced OPTOGait Photoelectric Cell System for the assessment of spatio-temporal parameters of gait. [Subjects] Twenty healthy young adults (mean age = 27.35, SD = 7.4) were asked to walk 3 times on walkway at a comfortable speed. [Methods] Concurrent validity was assessed by comparing data obtained using the OPTOGait and GAITRite systems, and reliability was assessed by comparing data from the first and third OPTOGait sessions. [Results] Concurrent validity, as identified by intra-class correlation coefficients (ICC (2, 1) = 0.929-0.998), coefficients of variation (CVME = 0.32-11.30%), and 95% limits of agreement, showed high levels of correlation. In addition, the test-retest reliability of the OPTOGait Photoelectric Cell System was demonstrated as showing a high level of correlation with all spatio-temporal parameters by intra-class correlation coefficients (ICC (3, 1) = 0.785-0.952), coefficients of variation (CVME = 1.66-4.06%), 95% limits of agreement, standard error of measurement (SEM = 2.17-5.96%), and minimum detectable change (MDC95% = 6.01-16.52%). [Conclusion] The OPTOGait Photoelectric Cell System has strong concurrent validity along with relative and absolute test-retest reliabilities. This portable system with easy-to-use features can be used for clinical assessments or research purposes as an objective means of assessing gait.
引用
收藏
页码:81 / 85
页数:5
相关论文
共 28 条
[11]   CLINICAL-SIGNIFICANCE - A STATISTICAL APPROACH TO DEFINING MEANINGFUL CHANGE IN PSYCHOTHERAPY-RESEARCH [J].
JACOBSON, NS ;
TRUAX, P .
JOURNAL OF CONSULTING AND CLINICAL PSYCHOLOGY, 1991, 59 (01) :12-19
[12]  
JUDGE JO, 1993, ARCH PHYS MED REHAB, V74, P400
[13]   RELIABILITY OF OBSERVATIONAL KINEMATIC GAIT ANALYSIS [J].
KREBS, DE ;
EDELSTEIN, JE ;
FISHMAN, S .
PHYSICAL THERAPY, 1985, 65 (07) :1027-1033
[14]   Test-retest reliability of the GAITRite system in people with stroke undergoing rehabilitation [J].
Kuys, Suzanne S. ;
Brauer, Sandra G. ;
Ada, Louise .
DISABILITY AND REHABILITATION, 2011, 33 (19-20) :1848-1853
[15]   Validity of the Optogait photoelectric system for the assessment of spatiotemporal gait parameters [J].
Lienhard, Karin ;
Schneider, David ;
Maffiuletti, Nicola A. .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (04) :500-504
[16]   Intra-rater and inter-rater reliability of gait measurements with CODA mpx30 motion analysis system [J].
Maynard, V ;
Bakheit, AMO ;
Oldham, J ;
Freeman, J .
GAIT & POSTURE, 2003, 17 (01) :59-67
[17]   The validity and reliability of the GAITRite system's measurements: A preliminary evaluation [J].
McDonough, AL ;
Batavia, M ;
Chen, FC ;
Kwon, S ;
Ziai, J .
ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2001, 82 (03) :419-425
[18]   Reliability of the GAITRite® walkway system for the quantification of temporo-spatial parameters of gait in young and older people [J].
Menz, HB ;
Latt, MD ;
Tiedemann, A ;
Kwan, MMS ;
Lord, SR .
GAIT & POSTURE, 2004, 20 (01) :20-25
[19]   Outcome measures to assess walking ability following stroke: a systematic review of the literature [J].
Mudge, Suzie ;
Stott, N. Susan .
PHYSIOTHERAPY, 2007, 93 (03) :189-200
[20]  
Portney L, 2000, FDN CLIN RES APPL PR, P579