Accuracy of bedside point of care testing in critical emergency department patients

被引:9
作者
McIntosh, Braden W. [1 ]
Vasek, Jerina [2 ]
Taylor, Maria [1 ]
Le Blanc, Deborah [1 ]
Thode, Henry C. [1 ]
Singer, Adam J. [1 ]
机构
[1] SUNY Stony Brook, Dept Emergency Med, Stony Brook, NY 11794 USA
[2] SUNY Stony Brook, Dept Clin Pathol, Stony Brook, NY 11794 USA
关键词
Point-of-care testing; Accuracy; Emergency department; OF-CARE; ELECTROLYTE;
D O I
10.1016/j.ajem.2017.09.018
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Point-of-care (POC) testing reduces laboratory turn-around having the potential to improve timely diagnosis and management. We compared the accuracy of nurse performed POC and core laboratory testing and determined whether deviations between the two were clinically meaningful. Methods: We performed a prospective, observational study on a convenience sample of 50 critical care ED patients in whom a POC chemistry and hematocrit was ordered. Blood samples were divided into 2 aliquots; one sample was tested by the treating nurse using a handheld POC device and the other sample was tested in the core laboratory. Paired comparisons of test results were performed using Pearson's correlation coefficients, Lin concordance coefficients, and Bland Altman plots. Results: Mean patient age was 67, 50% were male, 82% were admitted. Pearson's correlation and Lin concordance coefficients were excellent (0.84-1.00) for all 8 analytes. Mean (95%CI) paired differences between POC and core laboratory measurements were Na+ 0.30 (-0.22 to 0.82) mmol/L, K+ - 0.12 (-0.14 to -0.09) mmol/L, Cl- 2.10 (1.41 to 2.78) mmol/L, TCO2-1.68 (-2.06 to-1.30) mmol/L, glucose 2.46 (1.46 to 3.46) mg/dL, BUN, 1.69 (0.95 to 2.42) mg/dL, creatinine 0.13 (0.08 to 0.17) mg/dL, and hematocrit -0.39 (-0.93 to 0.15) %. In 3 of 400 measurements, the difference between POC and core lab exceeded the maximal clinically acceptable deviation based on physician surveys. Conclusions: Bedside POC by ED nurses is reliable and accurate and does not deviate significantly from core laboratory testing by trained technicians. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:567 / 570
页数:4
相关论文
共 12 条
[1]  
Asirvatham Jaya R, 2013, N Am J Med Sci, V5, P255, DOI 10.4103/1947-2714.110426
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   Electrolytes assessed by point-of-care testing - Are the values comparable with results obtained from the central laboratory? [J].
Chacko, Binila ;
Peter, John Victor ;
Patole, Shalom ;
Fleming, Jude Joseph ;
Selvakumar, Ratnasamy .
INDIAN JOURNAL OF CRITICAL CARE MEDICINE, 2011, 15 (01) :24-29
[4]  
Forsman RW, 1996, CLIN CHEM, V42, P813
[5]   Measurement of whole-blood potassium - Is it clinically safe? [J].
Hawkins, R .
CLINICAL CHEMISTRY, 2003, 49 (12) :2105-2106
[6]   Discrepancy between visual and spectrophotometric assessment of sample haemolysis [J].
Hawkins, R .
ANNALS OF CLINICAL BIOCHEMISTRY, 2002, 39 :521-522
[7]  
Jarvis PRE, 2016, CLIN EXP EMERG MED, V3, P63
[8]   Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers [J].
Leino, Aila ;
Kurvinen, Kaisa .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (07) :1187-1191
[9]   A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY [J].
LIN, LI .
BIOMETRICS, 1989, 45 (01) :255-268
[10]   Point-of-care testing in the overcrowded emergency department - can it make a difference? [J].
Rooney, Kevin D. ;
Schilling, Ulf Martin .
CRITICAL CARE, 2014, 18 (06)