The syringe potentiometer: a low-cost device for pneumotachograph calibration

被引:4
作者
Cross, Troy J. [1 ,2 ]
Kelley, Eli F. [1 ]
Hardy, Tim A. [3 ]
Isautier, Jennifer M. J. [1 ]
Johnson, Bruce D. [1 ]
机构
[1] Mayo Clin, Dept Cardiovasc Dis, Rochester, MN 55902 USA
[2] Griffith Univ, Menzies Hlth Inst Queensland, Brisbane, Qld, Australia
[3] Univ Leeds, Sch Biomed Sci, Leeds, W Yorkshire, England
关键词
calibration; pneumotachograph; potentiometer; respiratory airflow; syringe; ACCURACY;
D O I
10.1152/japplphysiol.00150.2019
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this report was to 1) detail the construction of a low-cost device that provides a "reference" flow waveform for pneumotachograph (PNT) calibration, i.e., the "syringe potentiometer" (SP), and to compare the 2) accuracy and 3) practical performance of "calibration curves" obtained with the SP device to other more established methods of PNT calibration, i.e., the weighted averaging (WA) and polynomial least-squares (PolyLS) methods. Volume and flow waveforms obtained via the SP device were validated against a motion capture system and were deemed accurate surrogates of actual syringe volume and flows. The SP device was used to construct a calibration curve of a PNT by dividing the flow waveform of the SP by the analog output of the PNT amplifier. A total of 187 inspiratory and 187 expiratory strokes were collected. When the entire data set of expiratory strokes was used, the SP, WA, and PolyLS methods together demonstrated acceptable volume and flow errors as per American Thoracic Society/European Respiratory Society recommendations (less than +/- 3.5% and less than +/- 5.0% errors, respectively). The "practical" performance of each method was assessed with a nested subsampling procedure, whereby volume and flow errors were evaluated as the number of strokes was increased (in blocks of 5 strokes). To this end, the SP method demonstrated practical performance superior to that of the WA and PolyLS approaches, whereby acceptable volume and flow errors were achieved after only 5 calibration strokes; the WA and PolyLS methods required 15 and 20 strokes, respectively, to achieve the same level of volume and flow accuracy. NEW & NOTEWORTHY This report describes the construction and validation of a low-cost device for the purposes of pneumotachograph (PNT) calibration: the "syringe potentiometer" (SP). Calibration of a PNT with the SP device yielded acceptable volume and flow errors (<3.5% and 5%, respectively) across a wide range of flows (<0.5 to 15 L/s). The SP device offered superior "practical performance" over other established PNT calibration methods, whereby acceptable volume and flow errors were achieved after only five calibration strokes.
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [21] Simple low-cost construction and calibration of accurate pneumotachographs for monitoring mechanical ventilation in low-resource settings
    Farre, Ramon
    Rodriguez-Lazaro, Miguel A.
    Gozal, David
    Trias, Gerard
    Solana, Gorka
    Navajas, Daniel
    Otero, Jorge
    FRONTIERS IN MEDICINE, 2022, 9
  • [22] Excalibur: An Accurate, Scalable, and Low-Cost Calibration Tool for Sensing Devices
    Li, Chia-Chi
    Dezfouli, Behnam
    IEEE SENSORS JOURNAL, 2019, 19 (23) : 11323 - 11336
  • [23] Low-Cost Eye Tracking Calibration: A Knowledge-Based Study
    Garde, Gonzalo
    Larumbe-Bergera, Andoni
    Bossavit, Benoit
    Porta, Sonia
    Cabeza, Rafael
    Villanueva, Arantxa
    SENSORS, 2021, 21 (15)
  • [24] Low-Cost Imaging Photometer and Calibration Method for Road Tunnel Lighting
    Cattini, Stefano
    Rovati, Luigi
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2012, 61 (05) : 1181 - 1192
  • [25] Low-Cost and Efficient Thermal Calibration Scheme for MEMS Triaxial Accelerometer
    Xu, Tongxu
    Xu, Xiang
    Xu, Dacheng
    Zou, Zelan
    Zhao, Heming
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2021, 70
  • [26] In chamber calibration and performance evaluation of air quality low-cost sensors
    Agrawal, Deepali
    Saini, Anil Kumar
    Rai, Aakash Chand
    Kala, Prateek
    ATMOSPHERIC POLLUTION RESEARCH, 2024, 15 (12)
  • [27] Chromatic aberration correction: an enhancement to the calibration of low-cost digital dermoscopes
    Wighton, Paul
    Lee, Tim K.
    Lui, Harvey
    McLean, David
    Atkins, M. Stella
    SKIN RESEARCH AND TECHNOLOGY, 2011, 17 (03) : 339 - 347
  • [28] Low-Cost Inertial Measurement Unit Calibration With Nonlinear Scale Factors
    Zhang, Xin
    Zhou, Changle
    Chao, Fei
    Lin, Chih-Min
    Yang, Longzhi
    Shang, Changjing
    Shen, Qiang
    IEEE TRANSACTIONS ON INDUSTRIAL INFORMATICS, 2022, 18 (02) : 1028 - 1038
  • [29] A new static method of calibration for low-cost laser triangulation systems
    Zhao, Yihui
    Supri, Taufik Bin Mohamed
    Yang, Song
    Qin, Yi
    MEASUREMENT, 2020, 156 (156)
  • [30] Addressing Low-Cost Methane Sensor Calibration Shortcomings with Machine Learning
    Kiplimo, Elijah
    Riddick, Stuart N.
    Mbua, Mercy
    Upreti, Aashish
    Anand, Abhinav
    Zimmerle, Daniel J.
    ATMOSPHERE, 2024, 15 (11)