The syringe potentiometer: a low-cost device for pneumotachograph calibration

被引:4
作者
Cross, Troy J. [1 ,2 ]
Kelley, Eli F. [1 ]
Hardy, Tim A. [3 ]
Isautier, Jennifer M. J. [1 ]
Johnson, Bruce D. [1 ]
机构
[1] Mayo Clin, Dept Cardiovasc Dis, Rochester, MN 55902 USA
[2] Griffith Univ, Menzies Hlth Inst Queensland, Brisbane, Qld, Australia
[3] Univ Leeds, Sch Biomed Sci, Leeds, W Yorkshire, England
关键词
calibration; pneumotachograph; potentiometer; respiratory airflow; syringe; ACCURACY;
D O I
10.1152/japplphysiol.00150.2019
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The purpose of this report was to 1) detail the construction of a low-cost device that provides a "reference" flow waveform for pneumotachograph (PNT) calibration, i.e., the "syringe potentiometer" (SP), and to compare the 2) accuracy and 3) practical performance of "calibration curves" obtained with the SP device to other more established methods of PNT calibration, i.e., the weighted averaging (WA) and polynomial least-squares (PolyLS) methods. Volume and flow waveforms obtained via the SP device were validated against a motion capture system and were deemed accurate surrogates of actual syringe volume and flows. The SP device was used to construct a calibration curve of a PNT by dividing the flow waveform of the SP by the analog output of the PNT amplifier. A total of 187 inspiratory and 187 expiratory strokes were collected. When the entire data set of expiratory strokes was used, the SP, WA, and PolyLS methods together demonstrated acceptable volume and flow errors as per American Thoracic Society/European Respiratory Society recommendations (less than +/- 3.5% and less than +/- 5.0% errors, respectively). The "practical" performance of each method was assessed with a nested subsampling procedure, whereby volume and flow errors were evaluated as the number of strokes was increased (in blocks of 5 strokes). To this end, the SP method demonstrated practical performance superior to that of the WA and PolyLS approaches, whereby acceptable volume and flow errors were achieved after only 5 calibration strokes; the WA and PolyLS methods required 15 and 20 strokes, respectively, to achieve the same level of volume and flow accuracy. NEW & NOTEWORTHY This report describes the construction and validation of a low-cost device for the purposes of pneumotachograph (PNT) calibration: the "syringe potentiometer" (SP). Calibration of a PNT with the SP device yielded acceptable volume and flow errors (<3.5% and 5%, respectively) across a wide range of flows (<0.5 to 15 L/s). The SP device offered superior "practical performance" over other established PNT calibration methods, whereby acceptable volume and flow errors were achieved after only five calibration strokes.
引用
收藏
页码:1150 / 1162
页数:13
相关论文
共 50 条
  • [1] Laboratory Calibration and Performance Evaluation of Low-Cost Capacitive and Very Low-Cost Resistive Soil Moisture Sensors
    Adla, Soham
    Rai, Neeraj Kumar
    Karumanchi, Sri Harsha
    Tripathi, Shivam
    Disse, Markus
    Pande, Saket
    SENSORS, 2020, 20 (02)
  • [2] Calibration Stability and Localization Accuracy of a Low-Cost and Portable Breast Microwave Sensing Device
    Fontaine, Gabrielle
    Jelani, Awab
    Pistorius, Stephen
    2024 IEEE MTT-S INTERNATIONAL MICROWAVE BIOMEDICAL CONFERENCE, IMBIOC 2024, 2024, : 48 - 50
  • [3] A Bezier based state calibrating method for low-cost potentiometer with inherent nonlinearity
    Hua, Hongliang
    Liao, Zhenqiang
    Wu, Xiaofeng
    Chen, Yongjiang
    MEASUREMENT, 2021, 178
  • [4] The calibration and deployment of a low-cost methane sensor
    Riddick, Stuart N.
    Mauzerall, Denise L.
    Celia, Michael
    Allen, Grant
    Pitt, Joseph
    Kang, Mary
    Riddick, John C.
    ATMOSPHERIC ENVIRONMENT, 2020, 230
  • [5] LOW-COST CALIBRATION METHOD FOR THE INFRARED CAMERA
    Uemura, Derek Koji
    Shah, Sanjay B.
    Regmi, Prafulla
    Grimes, Jesse
    Wang-Li, Lingjuan
    APPLIED ENGINEERING IN AGRICULTURE, 2023, 39 (05) : 529 - 534
  • [6] GNSS Based Low-Cost Magnetometer Calibration
    Andel, Jan
    Simak, Vojtech
    Kanalikova, Alzbeta
    Pirnik, Rastislav
    SENSORS, 2022, 22 (21)
  • [7] A Low-Cost Calibration Method for Low-Cost MEMS Accelerometers Based on 3D Printing
    Garcia, Jesus A.
    Lara, Evangelina
    Aguilar, Leocundo
    SENSORS, 2020, 20 (22) : 1 - 19
  • [8] Calibration verification of a low-cost method for MEMS accelerometers
    Parsi, B.
    Bahrami, M.
    Esfahani, A. Monemian
    Sany, B. Seyedzadeh
    TRANSACTIONS OF THE INSTITUTE OF MEASUREMENT AND CONTROL, 2014, 36 (05) : 579 - 587
  • [9] Low-Cost Calibration Techniques for Smart Temperature Sensors
    Pertijs, Michiel A. P.
    Aita, Andre L.
    Makinwa, Kofi A. A.
    Huijsing, Johan H.
    IEEE SENSORS JOURNAL, 2010, 10 (06) : 1098 - 1105
  • [10] Low-Cost Formaldehyde Sensor Evaluation and Calibration in a Controlled Environment
    Chattopadhyay, Arnab
    Huertas, Andres
    Rebeiro-Hargrave, Andrew
    Fung, Pak Lun
    Varjonen, Samu
    Hieta, Tuomas
    Tarkoma, Sasu
    Petaja, Tuukka
    IEEE SENSORS JOURNAL, 2022, 22 (12) : 11791 - 11802