Multicenter laboratory evaluation of the bioMerieux Vitek antimicrobial susceptibility testing system with 11 antimicrobial agents versus members of the family Enterobacteriaceae and Pseudomonas aeruginosa

被引:36
作者
Doern, GV
Brueggemann, AB
Perla, R
Daly, J
Halkias, D
Jones, RN
Saubolle, MA
机构
[1] UNIV MASSACHUSETTS, SCH MED, WORCESTER, MA 01655 USA
[2] PRIMARY CHILDRENS MED CTR, SALT LAKE CITY, UT 84113 USA
[3] JAMES A HALEY VET ADM MED CTR, TAMPA, FL 33612 USA
[4] GOOD SAMARITAN HOSP, PHOENIX, AZ 85062 USA
关键词
D O I
10.1128/JCM.35.8.2115-2119.1997
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
A four-center study in which a total of 1,082 recent clinical isolates of members of the family Enterobacteriaceae and Pseudomonas aeruginosn were examined versus 11 antimicrobial agents with the bioMerieux Vitek susceptibility test system (Hazelwood, Mo.) and the GNS-F6 card was conducted. In addition, a challenge set consisting of the same 200 organisms was examined in each of the four participating laboratories, Results obtained with the Vitek system were compared to MICs determined by a standardized broth microdilution method, For purposes of comparison, susceptibility categories (susceptible, intermediate, or resistant) were assigned on the basis of the results of both methods, The result of the broth microdilution test was considered definitive, The total category error rate with the Vitek system and the recent clinical isolates (11,902 organism-antimicrobial comparisons) was 4.5%, i.e,, 1.7% very major errors, 0.9% major errors, and 1.9% minor errors, The total category error rate calculated from tests performed with the challenge set (i.e., 8,800 organism-antimicrobial comparisons) was 5.9%, i.e., 2.2% very major errors, 1.1% major errors, and 2.6% minor errors, Very major error rates higher than the totals were noted with Enterobacter cloacae versus ampicillin-sulbactam, aztreonam, ticarcillin, and ticarcillin-clavulanate and with P. aeruginosa versus mezlocillin, ticarcillin, and ticarcillin-clavulanate. Major error rates higher than the averages were observed with Proteus mirabilis versus imipenem and with Klebsiella pneumoniae versus ofloxacin, Excellent overall interlaboratory reproducibility was observed with the Vitek system, The importance of inoculum size as a primary determinant in the accuracy of susceptibility test results with the Vitek system was clearly demonstrated in this study, Specifically, when an inoculum density fourfold higher than that recommended by the manufacturer was used, high rates of false resistance results were obtained with cell wall-active antimicrobial agents versus both the Enterobacteriaceae and P. aeruginosa.
引用
收藏
页码:2115 / 2119
页数:5
相关论文
共 32 条
[1]  
BEIDENBACH DJ, 1995, DIAGN MCIROBIOL INFE, V21, P57
[2]   SPECIES IDENTIFICATION AND ANTIBIOTIC-RESISTANCE PATTERNS OF THE ENTEROCOCCI [J].
BRYCE, EA ;
ZEMCOV, SJV ;
CLARKE, AM .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1991, 10 (09) :745-747
[3]  
COULDRON PE, 1986, J CLIN MICROBIOL, V24, P764
[4]   MODIFICATION OF THE SCEPTOR SYSTEM FOR RAPID DETECTION OF METHICILLIN-RESISTANT STAPHYLOCOCCI [J].
DENYS, GA ;
SAHM, DF .
JOURNAL OF CLINICAL MICROBIOLOGY, 1986, 24 (03) :462-464
[5]   Accurate characterization of ofloxacin susceptibility with Enterobacteriaceae using a modified GNS F6 card and the bioMerieux Vitek System [J].
Doern, GV ;
Torres, BB ;
Jankins, M ;
Jones, RN .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1996, 25 (03) :133-135
[6]   A COMPARISON OF SERIAL PLATE AGAR DILUTION, BAUER-KIRBY DISK DIFFUSION, AND THE VITEK AUTOMICROBIC SYSTEM FOR THE DETERMINATION OF SUSCEPTIBILITIES OF KLEBSIELLA SPP, ENTEROBACTER SPP, AND PSEUDOMONAS-AERUGINOSA TO 10 ANTIMICROBIAL AGENTS [J].
FEKETE, T ;
TUMAH, H ;
WOODWELL, J ;
TRUANT, A ;
SATISHCHANDRAN, V ;
AXELROD, P ;
KRETER, B .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1994, 18 (04) :251-258
[7]   DETECTION OF INTRINSICALLY RESISTANT (HETERORESISTANT) STAPHYLOCOCCUS-AUREUS WITH THE SCEPTOR AND AUTOMICROBIC SYSTEMS [J].
HANSEN, SL ;
WALSH, TJ .
JOURNAL OF CLINICAL MICROBIOLOGY, 1987, 25 (02) :412-415
[8]   COMPARATIVE-EVALUATION OF 4 SYSTEMS FOR DETERMINING SUSCEPTIBILITY OF GRAM-POSITIVE ORGANISMS [J].
HENRY, D ;
KUNZER, L ;
NGUIYEN, J ;
SMITH, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 1986, 23 (04) :718-724
[9]   COMPARISON OF THE AMS GRAM-NEGATIVE SUSCEPTIBILITY FLEX PANEL GNS-V AND AGAR DISK DIFFUSION FOR TESTING OF PSEUDOMONAS-AERUGINOSA [J].
HENRY, DA ;
SMITH, JA ;
JONES, MA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1987, 25 (03) :586-587
[10]   EVALUATION OF THE AUTOMICROBIC SYSTEM FOR SUSCEPTIBILITY TESTING OF AMINOGLYCOSIDES AND GRAM-NEGATIVE BACILLI [J].
HINDLER, JA ;
BRUCKNER, DA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1987, 25 (03) :546-550