Validity of shoe-type inertial measurement units for Parkinson's disease patients during treadmill walking

被引:31
作者
Lee, Myeounggon [1 ]
Youm, Changhong [2 ]
Jeon, Jeanhong [2 ]
Cheon, Sang-Myung [3 ]
Park, Hwayoung [1 ]
机构
[1] Dong A Univ, Coll Hlth Sci, Biomech Lab, Hadan 2 Dong, Busan, South Korea
[2] Dong A Univ, Dept Hlth Care & Sci, Coll Hlth Sci, 37 Nakdong Daero 550 Beon Gil,Hadan 2 Dong, Busan, South Korea
[3] Dong A Univ, Sch Med, Dept Neurol, Dongdaesin Dong 3 Ga, Busan, South Korea
来源
JOURNAL OF NEUROENGINEERING AND REHABILITATION | 2018年 / 15卷
关键词
Parkinson's disease; Inertial measurement unit; Spatiotemporal parameter; Gait; Validity; GAIT VARIABILITY; SYSTEM; PARAMETERS; SENSORS; ACCELEROMETER; RELIABILITY; MOVEMENT;
D O I
10.1186/s12984-018-0384-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: When examining participants with pathologies, a shoe-type inertial measurement unit (IMU) system with sensors mounted on both the left and right outsoles may be more useful for analysis and provide better stability for the sensor positions than previous methods using a single IMU sensor or attached to the lower back and a foot However, there have been few validity analyses of shoe-type IMU systems versus reference systems for patients with Parkinson's disease (PD) walking continuously with a steady-state gait in a single direction. Therefore, the purpose of this study is to assess the validity of the shoe-type IMU system versus a 3D motion capture system for patients with PD during 1 min of continuous walking on a treadmill. Methods: Seventeen participants with PD successfully walked on a treadmill for 1 min. The shoe-type IMU system and a motion capture system comprising nine infrared cameras were used to collect the treadmill walking data with participants moving at their own preferred speeds. All participants took anti-parkinsonian medication at least 3 h before the treadmill walk An intraclass correlation coefficient analysis and the associated 95% confidence intervals were used to evaluate the validity of the resultant linear acceleration and spatiotemporal parameters for the IMU and motion capture systems. Results: The resultant linear accelerations, cadence, left step length, right step length, left step time, and right step time showed excellent agreement between the shoe-type IMU and motion capture systems. Conclusions: The shoe-type IMU system provides reliable data and can be an alternative measurement tool for objective gait analysis of patients with PD in a clinical environment.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Agreement of Gait Events Detection during Treadmill Backward Walking by Kinematic Data and Inertial Motion Units
    Gottlieb, Uri
    Balasukumaran, Tharani
    Hoffman, Jay R.
    Springer, Shmuel
    SENSORS, 2020, 20 (21) : 1 - 9
  • [42] Discriminant validity and reproducibility of spatiotemporal and kinetic parameters during treadmill walking in patients with knee osteoarthritis
    Pacifico, Daniela
    Visscher, Rosa
    List, Renate
    Item-Glatthorn, Julia F.
    Casartelli, Nicola C.
    Maffiuletti, Nicola A.
    GAIT & POSTURE, 2020, 80 : 77 - 79
  • [43] Validity and Reliability of Inertial Measurement Units on Lower Extremity Kinematics During Running: A Systematic Review and Meta-Analysis
    Ziwei Zeng
    Yue Liu
    Xiaoyue Hu
    Meihua Tang
    Lin Wang
    Sports Medicine - Open, 2022, 8
  • [44] Validity and Reliability of Inertial Measurement Units on Lower Extremity Kinematics During Running: A Systematic Review and Meta-Analysis
    Zeng, Ziwei
    Liu, Yue
    Hu, Xiaoyue
    Tang, Meihua
    Wang, Lin
    SPORTS MEDICINE-OPEN, 2022, 8 (01)
  • [45] Reliability and validity of the Beck Depression Inventory in patients with Parkinson's disease
    Visser, M
    Leentjens, AFG
    Marinus, J
    Stiggelbout, AM
    van Hilten, JJ
    MOVEMENT DISORDERS, 2006, 21 (05) : 668 - 672
  • [46] Avoiding Virtual Obstacles During Treadmill Gait in Parkinson's Disease
    Lu, Chiahao
    Twedell, Emily
    Elbasher, Reem
    McCabe, Michael
    MacKinnon, Colum D.
    Cooper, Scott E.
    FRONTIERS IN AGING NEUROSCIENCE, 2019, 11
  • [47] Wireless devices to restart walking during an episode of FOG on patients with Parkinson's disease
    Punin, C.
    Barzallo, B.
    Huerta, M.
    Bermeo, A.
    Bravo, M.
    Llumiguano, C.
    2017 IEEE SECOND ECUADOR TECHNICAL CHAPTERS MEETING (ETCM), 2017,
  • [48] Impaired Inhibitory Control During Walking in Parkinson's Disease Patients: An EEG Study
    Sosnik, Ronen
    Danziger-Schragenheim, Shani
    Possti, Daniel
    Fahoum, Firas
    Giladi, Nir
    Hausdorff, Jeffrey M.
    Mirelman, Anat
    Maidan, Inbal
    JOURNAL OF PARKINSONS DISEASE, 2022, 12 (01) : 243 - 256
  • [49] ACCELERATION PATTERN OF THE UPPER BODY DURING LEVEL WALKING IN PATIENTS WITH PARKINSON'S DISEASE
    Kim, Ji-Won
    Kwon, Yu-Ri
    Heo, Jae-Hoon
    Eom, Gwang-Moon
    Kwon, Moon-Seok
    Tack, Gye-Rae
    Koh, Seong-Beom
    Journal of Mechanics in Medicine and Biology, 2016, 16 (08)
  • [50] Walking on a treadmill improves the stride length-cadence relationship in individuals with Parkinson's disease
    Ambrus, M.
    Sanchez, Ja
    Fernandez-del-Olmo, M.
    GAIT & POSTURE, 2019, 68 : 136 - 140