Comparison of the ARK Immunoassay With High-Performance Liquid Chromatography With Ultraviolet Detection for Therapeutic Drug Monitoring of Linezolid

被引:1
|
作者
Castoldi, Simone [1 ]
Cozzi, Valeria [1 ]
Baldelli, Sara [1 ]
Fucile, Serena [1 ]
Clementi, Emilio [2 ,3 ]
Cattaneo, Dario [1 ]
机构
[1] ASST Fatebenefratelli Sacco Univ Hosp, Unit Clin Pharmacol, Via GB Grassi 74, I-20157 Milan, Italy
[2] Univ Milan, L Sacco Univ Hosp, CNR Inst Neurosci, Dept Biomed & Clin Sci,Clin Pharmacol Unit, Milan, Italy
[3] Sci Inst IRCCS Eugenio Medea, Bosisio Parini, Italy
关键词
linezolid; immunoassay; HPLC; method comparison; IN-VITRO; TUBERCULOSIS; PLASMA;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: An enzymatic immunoassay is under development by ARK Diagnostics, Inc for the quantification of plasma concentrations of linezolid (LZD). In this study, the authors aimed to assess the performance of this immunoassay using a validated high-performance liquid chromatography (HPLC) ultraviolet method as reference. Methods: Within-and between-day in vitro inaccuracy and imprecision of the ARK LZD assay were firstly tested using spiked quality controls (QC) provided by the kit manufacturer. Subsequently, the performance of the immunoassay was verified in vivo by analyzing 170 trough LZD plasma samples from patients on antibiotic therapy. Results: Imprecision of the spiked QCs resulted in every instance less than 7.0% and the inaccuracy ranged from -1.5% to 6.6%. The linear correlation between the 2 methods was documented by the Pearson analysis of plasma samples from patients on LZD therapy (coefficient = 0.9619). By Bland-Altman comparison, 8.2% of the patient samples resulted out of the limits ranging from -27.0% to + 33.5%, with most of them having LZD concentrations exceeding 10 mg/L. Conclusions: Acceptable analytical performance of the ARK LZD immunoassay has been demonstrated both with spiked QC and patients' samples, making it a viable alternative to HPLC for the therapeutic drug monitoring of LZD in clinical practice in laboratory hospitals that do not have HPLC equipment.
引用
收藏
页码:140 / 143
页数:4
相关论文
共 50 条
  • [31] TECHNICAL CONSIDERATIONS IN THE USE OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY IN THERAPEUTIC DRUG-MONITORING
    GIESE, RW
    CLINICAL CHEMISTRY, 1983, 29 (07) : 1331 - 1343
  • [32] Therapeutic drug monitoring of immunosuppressant drugs by high-performance liquid chromatography-mass spectrometry
    Taylor, PJ
    THERAPEUTIC DRUG MONITORING, 2004, 26 (02) : 215 - 219
  • [33] Development of a simple high-performance liquid chromatography- ultraviolet method for sotorasib quantification in human plasma: Implications for therapeutic drug monitoring
    Hikita, Eri
    Gando, Yoshito
    Chubachi, Hideo
    Shirota, Mikio
    Kushiyama, Akifumi
    Yasu, Takeo
    DRUG DISCOVERIES AND THERAPEUTICS, 2023, 17 (04): : 289 - 293
  • [34] High performance liquid chromatography detection of lamotrigin in therapeutic drug monitoring:: Marmara university experience
    Küçúkibrahimoglu, E
    Terzioglu, B
    Gülçebi, M
    Karaalp, A
    Gören, MZ
    Onat, F
    EPILEPSIA, 2005, 46 : 277 - 277
  • [35] DRUG-MONITORING OF ANTIEPILEPTIC DRUGS - COMPARISON OF 2 ANALYTICAL METHODS - HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND ENZYME-IMMUNOASSAY
    BLANKENHORN, V
    JURGES, U
    MAY, T
    RAMBECK, B
    EPILEPSIA, 1995, 36 : S157 - S157
  • [36] Analysis of plasma thiols by high-performance liquid chromatography with ultraviolet detection
    Bald, E
    Chwatko, G
    Glowacki, R
    Kusmierek, K
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1032 (1-2) : 109 - 115
  • [37] Determination of thiocyanate anion by high-performance liquid chromatography with ultraviolet detection
    Yang, Zi-Yuan
    Chen, Su-Hwei
    Kou, Hwang-Shang
    Wu, Shou-Mei
    Wu, Hsin-Lung
    Chinese Pharmaceutical Journal, 1994, 46 (06):
  • [38] COMPARISON OF A NONINSTRUMENTED IMMUNOASSAY FOR CARBAMAZEPINE TO HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND FLUORESCENCE POLARIZATION IMMUNOASSAY
    COCHRAN, EB
    MASSEY, KL
    PHELPS, SJ
    CRAMER, JA
    TOFTNESS, BR
    DENIO, LS
    DRAKE, ME
    EPILEPSIA, 1990, 31 (04) : 480 - 484
  • [39] COMPARISON OF THEOPHYLLINE MEASUREMENTS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND ENZYME IMMUNOASSAY
    RYDER, KW
    OEI, TO
    GLICK, MR
    JIMERSON, JC
    KELLOGG, TL
    CLINICAL CHEMISTRY, 1978, 24 (06) : 1057 - 1057
  • [40] Development and Validation of a High-Performance Liquid Chromatography With Ultraviolet Detection Method to Facilitate Therapeutic Monitoring of Teicoplanin Using Dried Blood Spots
    Ramadan, Ola
    Opitz, Patrick
    Hempel, Georg
    THERAPEUTIC DRUG MONITORING, 2024, 46 (05) : 627 - 633