Comparison of the ARK Immunoassay With High-Performance Liquid Chromatography With Ultraviolet Detection for Therapeutic Drug Monitoring of Linezolid

被引:1
|
作者
Castoldi, Simone [1 ]
Cozzi, Valeria [1 ]
Baldelli, Sara [1 ]
Fucile, Serena [1 ]
Clementi, Emilio [2 ,3 ]
Cattaneo, Dario [1 ]
机构
[1] ASST Fatebenefratelli Sacco Univ Hosp, Unit Clin Pharmacol, Via GB Grassi 74, I-20157 Milan, Italy
[2] Univ Milan, L Sacco Univ Hosp, CNR Inst Neurosci, Dept Biomed & Clin Sci,Clin Pharmacol Unit, Milan, Italy
[3] Sci Inst IRCCS Eugenio Medea, Bosisio Parini, Italy
关键词
linezolid; immunoassay; HPLC; method comparison; IN-VITRO; TUBERCULOSIS; PLASMA;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: An enzymatic immunoassay is under development by ARK Diagnostics, Inc for the quantification of plasma concentrations of linezolid (LZD). In this study, the authors aimed to assess the performance of this immunoassay using a validated high-performance liquid chromatography (HPLC) ultraviolet method as reference. Methods: Within-and between-day in vitro inaccuracy and imprecision of the ARK LZD assay were firstly tested using spiked quality controls (QC) provided by the kit manufacturer. Subsequently, the performance of the immunoassay was verified in vivo by analyzing 170 trough LZD plasma samples from patients on antibiotic therapy. Results: Imprecision of the spiked QCs resulted in every instance less than 7.0% and the inaccuracy ranged from -1.5% to 6.6%. The linear correlation between the 2 methods was documented by the Pearson analysis of plasma samples from patients on LZD therapy (coefficient = 0.9619). By Bland-Altman comparison, 8.2% of the patient samples resulted out of the limits ranging from -27.0% to + 33.5%, with most of them having LZD concentrations exceeding 10 mg/L. Conclusions: Acceptable analytical performance of the ARK LZD immunoassay has been demonstrated both with spiked QC and patients' samples, making it a viable alternative to HPLC for the therapeutic drug monitoring of LZD in clinical practice in laboratory hospitals that do not have HPLC equipment.
引用
收藏
页码:140 / 143
页数:4
相关论文
共 50 条
  • [1] Therapeutic drug monitoring of voriconazole: validation of a high performance liquid chromatography method and comparison with an ARK immunoassay
    Blanco-Dorado, Sara
    Belles Medall, Maria Dolores
    Pascual-Marmaneu, Oscar
    Campos-Toimil, Manuel
    Javier Otero-Espinar, Francisco
    Rodriguez-Riego, Rafael
    Rodriguez-Jato, Teresa
    Zarra-Ferro, Irene
    Jesus Lamas, Maria
    Fernandez-Ferreiro, Anxo
    EUROPEAN JOURNAL OF HOSPITAL PHARMACY, 2021, 28 (03) : 154 - 159
  • [2] High-performance liquid chromatography with ultraviolet detection for therapeutic drug monitoring of everolimus
    Baldelli, S
    Murgia, S
    Merlini, S
    Zenoni, S
    Perico, N
    Remuzzi, G
    Cattaneo, D
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 816 (1-2): : 99 - 105
  • [3] Comparison of ultra-performance liquid chromatography and ARK immunoassay for therapeutic drug monitoring of voriconazole
    Pena-Lorenzo, Diego
    Rebollo, Noemi
    German Sanchez-Hernandez, Jose
    Zarzuelo-Castaneda, Aranzazu
    ANNALS OF CLINICAL BIOCHEMISTRY, 2021, 58 (06) : 657 - 660
  • [4] Therapeutic drug monitoring of voriconazole: validation of a novel ARK™ immunoassay and comparison with ultra-high performance liquid chromatography
    Cattoir, Lien
    Fauvarque, Gregoire
    Degandt, Simon
    Ghys, Timothy
    Verstraete, Alain G.
    Stove, Veronique
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (05) : 6135 - 6139
  • [5] A comparison of high-performance liquid chromatography and fluorescence polarization immunoassay for therapeutic drug monitoring of tricyclic antidepressants
    Hackett, LP
    Dusci, LJ
    Ilett, KF
    THERAPEUTIC DRUG MONITORING, 1998, 20 (01) : 30 - 34
  • [6] Therapeutic Drug Monitoring of Voriconazole: Comparison of Bioassay with High-Performance Liquid Chromatography
    Badiee, Parisa
    Hashemizadeh, Zahra
    Montaseri, Hashem
    JUNDISHAPUR JOURNAL OF MICROBIOLOGY, 2017, 10 (04)
  • [7] Determination of Linezolid in Human Plasma by High-Performance Liquid Chromatography With Ultraviolet Detection
    Cattaneo, Dario
    Baldelli, Sara
    Conti, Francesca
    Cozzi, Valeria
    Clementi, Emilio
    THERAPEUTIC DRUG MONITORING, 2010, 32 (04) : 520 - 524
  • [8] Therapeutic drug monitoring for sirolimus in whole blood of organ transplants by high-performance liquid chromatography with ultraviolet detection
    Campanero, MA
    Cardenas, E
    Sádaba, B
    García-Quetglas, E
    Muñoz-Juarez, MJ
    Gil-Aldea, I
    Pazo, D
    Azanza, JR
    Honorato, J
    JOURNAL OF CHROMATOGRAPHY A, 2004, 1031 (1-2) : 265 - 273
  • [9] High-performance liquid chromatography with ultraviolet detection for real-time therapeutic drug monitoring of meropenem in plasma
    Ikeda, Kayo
    Ikawa, Kazuro
    Morikawa, Norifumi
    Miki, Mizuka
    Nishimura, Shin-Ichiro
    Kobayashi, Masao
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 856 (1-2): : 371 - 375
  • [10] Therapeutic drug monitoring of antipsychotic drugs with high-performance liquid chromatography and simultaneous electrochemical-ultraviolet detection
    Schmid, I
    Müller-Dithenhof, A
    Patsch, W