3D versus 2D ultrasound - Accuracy of volume measurement in human cadaver kidneys

被引:69
作者
Partik, BL
Stadler, A
Schamp, S
Koller, A
Voracek, M
Heinz, G
Helbich, TH
机构
[1] Stanford Univ, Ctr Med, Dept Radiol, Stanford, CA 94305 USA
[2] Univ Vienna, Dept Radiol, A-1010 Vienna, Austria
[3] Univ Vienna, Inst Clin Pathol, A-1010 Vienna, Austria
[4] Univ Vienna, Dept Psychoanal & Psychotherapy, Documentat & Stat Branch, A-1010 Vienna, Austria
关键词
kidney; ultrasound; measurement; concordance correlation; comparative studies;
D O I
10.1097/00004424-200209000-00003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. Comparison of the accuracy of 3D and 2D ultrasound in assessing the volume of human cadaver kidneys. MATERIALS AND METHODS. Before autopsy the volume of 22 kidneys was assessed from a 3D data set after manually tracing organ contours (3D volumetry) and by applying a 3D ellipsoid formula both on a 3D data set and 2D images. Measurements by water-displacement served as the gold standard. RESULTS. 3D volumetry showed a mean absolute deviation of 31 mL (18.5%) compared with the mean gold standard measurement (168 mL), yielding a concordance correlation (Lin's rho(c)) of 0.71. Calculation based on the ellipsoid formula revealed a mean absolute deviation of 37 mL (22.0%) when applied on the 3D data set (rho(c) = 0.65) and of 42 mL (25.0%) when applied on 2D images (rho(c) = 0.61), respectively. CONCLUSIONS. 3D volumetry showed a satisfactory concordance correlation and is superior to volume calculation based on the ellipsoid formula either applied to a 3D data set or to conventional 2D images in assessing the volume of human cadaver kidneys.
引用
收藏
页码:489 / 495
页数:7
相关论文
共 34 条
  • [21] 3-DIMENSIONAL SPATIAL REGISTRATION AND INTERACTIVE DISPLAY OF POSITION AND ORIENTATION OF REAL-TIME ULTRASOUND IMAGES
    KING, DL
    KING, DL
    SHAO, MYC
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 1990, 9 (09) : 525 - 532
  • [22] Lee A, 1996, AM J OBSTET GYNECOL, V175, P588
  • [23] Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging
    Leotta, DF
    Detmer, PR
    Martin, RW
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (04) : 597 - 609
  • [24] A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY
    LIN, LI
    [J]. BIOMETRICS, 1989, 45 (01) : 255 - 268
  • [25] ASSAY VALIDATION USING THE CONCORDANCE CORRELATION-COEFFICIENT
    LIN, LIK
    [J]. BIOMETRICS, 1992, 48 (02) : 599 - 604
  • [26] Matre K, 1999, Eur J Ultrasound, V10, P65, DOI 10.1016/S0929-8266(99)00044-0
  • [27] VOLUME SCANNING IN THE EVALUATION OF FETAL MALFORMATIONS - A NEW DIMENSION IN PRENATAL-DIAGNOSIS
    MERZ, E
    BAHLMANN, F
    WEBER, G
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 1995, 5 (04) : 222 - 227
  • [28] Relationships among liver and kidney volumes, lean body mass and drug clearance
    Nawaratne, S
    Brien, JE
    Seeman, E
    Fabiny, R
    Zalcberg, J
    Cosolo, W
    Angus, P
    Morgan, DJ
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1998, 46 (05) : 447 - 452
  • [29] Is blood pressure related to kidney size and shape?
    Raman, GV
    Clark, A
    Campbell, S
    Watkins, L
    Osmond, C
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 1998, 13 (03) : 728 - 730
  • [30] Three-dimensional ultrasound: Accuracy of distance and volume measurements
    Riccabona, M
    Nelson, TR
    Pretorius, DH
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 1996, 7 (06) : 429 - 434