Comparison of the pharyngeal pressure provided by two heated, humidified high-flow nasal cannulae devices in premature infants

被引:29
|
作者
Collins, Clare L. [1 ]
Holberton, James R. [1 ]
Koenig, Kai [1 ]
机构
[1] Mercy Hosp Women, Dept Paediat, Melbourne, Vic 3084, Australia
关键词
continuous positive airway pressure; heated; humidified high-flow nasal cannula; infant; non-invasive ventilation; premature; POSITIVE AIRWAY PRESSURE; NARES;
D O I
10.1111/jpc.12277
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Aims: This study aims to determine if there is a difference in the pharyngeal pressure, measured as a surrogate for continuous positive distending airway pressure, delivered to premature infants between two commonly used heated, humidified high-flow nasal cannulae (HHHFNC) devices: Fisher & Paykel Healthcare HHHFNC and Vapotherm 2000i. Methods: Pharyngeal pressure measurements were taken from stable premature infants receiving HHHFNC for respiratory support. Flow rates of 2-8 L/min were studied. Results: Nine infants had pharyngeal pressure measurements recorded with both HHHFNC devices at flow rates of 2-8 L/min. There was no difference in pharyngeal pressures recorded between devices at flow rates of 2-6 L/min; measured pressure was linearly associated with flow (R-2 = 0.9). At flow rates of 7 L/min, Vapotherm delivered a mean (standard deviation) pharyngeal pressure of 4.7 (2.2) cmH(2)O compared with 4.23 (2.2) cmH(2)O by the Fisher & Paykel device (P = 0.04). At a flow of 8 L/min, the mean pharyngeal pressure via Vapotherm was 4.9 (2.2) cmH(2)O compared with 4.1 (2.3) cmH(2)O with the Fisher & Paykel device (P = 0.05). Conclusions: Both HHHFNC delivered similar pharyngeal pressures at flow rates of 2-6 L/min. The pressure limiter valve of the Fisher & Paykel device attenuated the pharyngeal pressures at flows of 7 and 8 L/min. Vapotherm trended towards higher delivered pharyngeal pressure at flow rates 7 and 8 L/min, but the clinical significance of the difference remains unclear.
引用
收藏
页码:554 / 556
页数:3
相关论文
共 50 条
  • [1] Pharyngeal pressure with high-flow nasal cannulae in premature infants
    Wilkinson, D. J.
    Andersen, C. C.
    Smith, K.
    Holberton, J.
    JOURNAL OF PERINATOLOGY, 2008, 28 (01) : 42 - 47
  • [2] Pharyngeal pressure with high-flow nasal cannulae in premature infants
    D J Wilkinson
    C C Andersen
    K Smith
    J Holberton
    Journal of Perinatology, 2008, 28 : 42 - 47
  • [3] A randomized pilot study comparing heated humidified high-flow nasal cannulae with NIPPV for RDS
    Kugelman, Amir
    Riskin, Arieh
    Said, Waseem
    Shoris, Irit
    Mor, Frida
    Bader, David
    PEDIATRIC PULMONOLOGY, 2015, 50 (06) : 576 - 583
  • [4] Pharyngeal pressure due to high-flow nasal cannula devices in preterm infants
    Kodera, Takayuki
    Takatera, Akihiro
    Morisawa, Takeshi
    Yonetani, Masahiko
    PEDIATRICS INTERNATIONAL, 2021, 63 (10) : 1212 - 1217
  • [5] Heated Humidified High-Flow Nasal Cannula for Prevention of Extubation Failure in Preterm Infants
    Soonsawad, Sasivimon
    Swatesutipun, Buranee
    Limrungsikul, Anchalee
    Nuntnarumit, Pracha
    INDIAN JOURNAL OF PEDIATRICS, 2017, 84 (04) : 262 - 266
  • [6] High-Flow Nasal Cannulae for Respiratory Support of Preterm Infants: A Review of the Evidence
    Manley, Brett J.
    Dold, Simone K.
    Davis, Peter G.
    Roehr, Charles C.
    NEONATOLOGY, 2012, 102 (04) : 300 - 308
  • [7] Evidence Support and Guidelines for Using Heated, Humidified, High-Flow Nasal Cannulae in Neonatology Oxford Nasal High-Flow Therapy Meeting, 2015
    Roehr, Charles C.
    Yoder, Bradley A.
    Davis, Peter G.
    Ives, Kevin
    CLINICS IN PERINATOLOGY, 2016, 43 (04) : 693 - +
  • [8] Heated, Humidified High-Flow Nasal Cannulae as a Form of Noninvasive Respiratory Support for Preterm Infants and Children with Acute Respiratory Failure
    Mardegan, Veronica
    Priante, Elena
    Lolli, Elisabetta
    Lago, Paola
    Baraldi, Eugenio
    AMERICAN JOURNAL OF PERINATOLOGY, 2016, 33 (11) : 1058 - 1061
  • [9] Heated Humidified High-Flow Nasal Cannula for Prevention of Extubation Failure in Preterm Infants
    Sasivimon Soonsawad
    Buranee Swatesutipun
    Anchalee Limrungsikul
    Pracha Nuntnarumit
    The Indian Journal of Pediatrics, 2017, 84 : 262 - 266
  • [10] Nursing perceptions of high-flow nasal cannulae treatment for very preterm infants
    Roberts, Calum T.
    Manley, Brett J.
    Dawson, Jennifer A.
    Davis, Peter G.
    JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2014, 50 (10) : 806 - 810