Evaluation of the articular cartilage of the knee joint with vastly undersampled isotropic projection reconstruction steady-state free precession Imaging
被引:33
作者:
Kijowski, Richard
论文数: 0引用数: 0
h-index: 0
机构:Univ Wisconsin Hosp, Dept Radiol, Ctr Clin Sci, Madison, WI 53792 USA
Kijowski, Richard
Lu, Aiming
论文数: 0引用数: 0
h-index: 0
机构:Univ Wisconsin Hosp, Dept Radiol, Ctr Clin Sci, Madison, WI 53792 USA
Lu, Aiming
Block, Walter
论文数: 0引用数: 0
h-index: 0
机构:Univ Wisconsin Hosp, Dept Radiol, Ctr Clin Sci, Madison, WI 53792 USA
Block, Walter
Grist, Thomas
论文数: 0引用数: 0
h-index: 0
机构:Univ Wisconsin Hosp, Dept Radiol, Ctr Clin Sci, Madison, WI 53792 USA
Grist, Thomas
机构:
[1] Univ Wisconsin Hosp, Dept Radiol, Ctr Clin Sci, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Biomed Engn, Madison, WI USA
[3] Univ Wisconsin, Dept Med Phys, Madison, WI USA
magnetic resonance imaging;
cartilage;
knee;
steady-state free precession;
isotropic resolution;
D O I:
10.1002/jmri.20596
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Purpose: To determine the feasibility of the vastly under-sampled isotropic projection reconstruction steady-state free precession (VIPR-SSFP) sequence for evaluating the articular cartilage of the knee joint. Materials and Methods: A magnetic resonance (MR) examination of the knee was performed on 33 subjects using a GE 1.5T scanner and a phased-array extremity coil. VIPR-SSFP, proton density-weighted fast spin-echo (PP-FSE), fat-suppressed T2-weighted fast spin-echo (T2-FSE), and three-dimensional fat-suppressed spoiled gradient recall-echo (SPGR) sequences were performed on three asymptomatic volunteers and 10 patients with osteoarthritis of the knee joint. Signal-to-noise efficiency, and contrast-to-noise ratio (CNR) measurements were calculated for all sequences and compared with the use of paired t-tests. The VIPR-SSFP sequence was then performed on 20 consecutive patients who were undergoing a routine MR examination of the knee. Results: The cartilage signal-to-noise efficiency of the VIPR-SSFP sequence was not significantly different from that of the PD-FSE and SPGR sequences. The cartilage signal-to-noise efficiency of the VIPR-SSFP sequence was significantly higher (P < 0.05) than that of the T2-FSE sequence. The VIPR-SSFP, sequence produced images with significantly higher (P < 0.05) CNR between cartilage and synovial fluid than the PD-FSE and SPGR sequences, and significantly higher (P < 0.05) CNR between cartilage and subchondral bone than the T2-FSE sequence. The VIPR-SSFP sequence allowed excellent visualization of the articular cartilage of the knee joint in all subjects. All articular cartilage defects identified on the PD-FSE, T2-FSE, and SPGR images were well visualized on the VIPR-SSFP images. Conclusion: VIPR-SSFP images had high cartilage signal-to-noise efficiency and high CNR between cartilage and adjacent synovial fluid and subehondral bone; therefore, the sequence is well suited for evaluating the articular cartilage of the knee joint.