Characterization measurement of a thick CdTe detector for BNCT-SPECT - Detection efficiency and energy resolution

被引:21
作者
Murata, Isao [1 ]
Nakamura, Soichiro [1 ]
Manabe, Masanobu [1 ]
Miyamaru, Hiroyuki [2 ]
Kato, Itsuro [3 ]
机构
[1] Osaka Univ, Div Elect Elect & Informat Engn, Grad Sch Engn, Suita, Osaka 5650871, Japan
[2] Osaka Prefecture Univ, Ctr Radiat Res, Sakai, Osaka 5998531, Japan
[3] Osaka Univ, Grad Sch Dent, Dept Oral & Maxillofacial Surg 2, Suita, Osaka 5650871, Japan
关键词
BNCT; SPECT; CdTe; Gamma-ray; B-10(n; alpha)Li-7; 478; keV;
D O I
10.1016/j.apradiso.2014.01.023
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Author's group is carrying out development of BNCT-SPECT with CdTe device, which monitors the therapy effect of BNCT in real-time. From the design calculations, the dimensions were fixed to 1.5 x 2 x 30 mm(3). For the collimator it was confirmed that it would have a good spatial resolution and simultaneously the number of counts would be acceptably large. After producing the CdTe crystal, the characterization measurement was carried out. For the detection efficiency an excellent agreement between calculation and measurement was obtained. Also, the detector has a very good energy resolution so that gamma-rays of 478 key and 511 keV could be distinguished in the spectrum. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:129 / 133
页数:5
相关论文
共 12 条
  • [1] Imahori Y, 1998, J NUCL MED, V39, P325
  • [2] Kato I, 2008, COMMUNICATION
  • [3] A noninvasive dose estimation system for clinical BNCT based on PG-SPECT - Conceptual study and fundamental experiments using HPGe and CdTe semiconductor detectors
    Kobayashi, T
    Sakurai, Y
    Ishikawa, M
    [J]. MEDICAL PHYSICS, 2000, 27 (09) : 2124 - 2132
  • [4] Development of JCDS, a computational dosimetry system at JAEA for boron neutron capture therapy
    Kumada, H.
    Yamamoto, K.
    Matsumura, A.
    Yamamoto, T.
    Nakagawa, Y.
    [J]. FIRST EUROPEAN WORKSHOP ON MONTE CARLO TREATMENT PLANNING, 2007, 74
  • [5] Manabe M., 2012, P 15 INT C NEUTR CAP, p6B
  • [6] Experimental feasibility studies on a SPECT tomograph for BNCT dosimetry
    Minsky, D. M.
    Valda, A. A.
    Kreiner, A. J.
    Green, S.
    Wojnecki, C.
    Ghani, Z.
    [J]. APPLIED RADIATION AND ISOTOPES, 2009, 67 (7-8) : S179 - S182
  • [7] Mukai T, 2010, JAEA C, P75
  • [8] Development of a thick CdTe detector for BNCT-SPECT
    Murata, I.
    Mukai, T.
    Nakamura, S.
    Miyamarua, H.
    Kato, I.
    [J]. APPLIED RADIATION AND ISOTOPES, 2011, 69 (12) : 1706 - 1709
  • [9] MURATA I, 2011, PROG NUCL SCI TECHNO, V0001, P00267
  • [10] Nakamura S., 2012, P 2011 ANN S NUCL DA, P165