Validation of Oscillometric Pulse Wave Analysis Measurements in Children

被引:27
作者
Stoner, Lee [1 ]
Lambrick, Danielle M. [2 ]
Westrupp, Nicole [1 ]
Young, Joanna [3 ]
Faulkner, James [1 ]
机构
[1] Massey Univ, Sch Sport & Exercise, Wellington, New Zealand
[2] Massey Univ, Inst Food Nutr & Human Hlth, Wellington, New Zealand
[3] Diabet Res Inst, Lipid & Diabet Res Grp, Christchurch, New Zealand
关键词
arterial stiffness; blood pressure; cardiovascular disease; central blood pressures; hypertension; pediatric; pulse wave analysis; validation; CENTRAL BLOOD-PRESSURE; CORONARY-ARTERY-DISEASE; NONINVASIVE ASSESSMENT; CARDIOVASCULAR EVENTS; AUGMENTATION INDEX; AORTIC PRESSURE; HEART-DISEASE; RADIAL ARTERY; ALL-CAUSE; STIFFNESS;
D O I
10.1093/ajh/hpt243
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
BACKGROUND Pulse wave analysis (PWA) has emerged as a noninvasive, valid, reliable, and widely used technique to investigate central blood pressures and systemic arterial wave reflection (augmentation index). The goldstandard technique is tonometry, but this technique can be challenging, especially when used on children. The purpose of this study was to validate oscillometric PWA for use in children. METHODS Fifty-seven healthy children were recruited for participation. Central blood pressures and peripheral augmentation index (pAIx) were measured objectively using oscillometric (Pulsecor R7) and tonometric (SphygmaCor) devices. All measurements were made during the same visit under standardized conditions between the hours of 8 am and 10 am in the fasted state. RESULTS Tonometric measurements were unsuccessful on 1 child. Comparisons were made on 56 children ( mean age = 9.8 +/- 1.0 y; 57% male). A very strong relationship was found between devices for central systolic ( r = 0.94; P < 0.001), diastolic ( r = 0.99; P < 0.001) and mean ( r = 0.96; P < 0.001) blood pressures. However, Bland-Altman analysis indicated a bias toward greater systolic blood pressures with the oscillometric monitor ( mean difference = 4.5 mm Hg; 95% confidence interval (CI) = -5.16 to -3.89). A good relationship was found for pAIx ( r=0.71; p < 0.001); the mean difference between devices was -1.70% ( 95% CI = 4.47% to 1.08%), which is not significantly different from zero. CONCLUSIONS Findings from this study suggest that oscillometric PWA provides valid measures of central blood pressure and arterial wave reflection in children aged 8-10 years.
引用
收藏
页码:865 / 872
页数:8
相关论文
共 39 条
  • [1] Role of Pulse Pressure Amplification in Arterial Hypertension Experts' Opinion and Review of the Data
    Avolio, Alberto P.
    Van Bortel, Luc M.
    Boutouyrie, Pierre
    Cockcroft, John R.
    McEniery, Carmel M.
    Protogerou, Athanase D.
    Roman, Mary J.
    Safar, Michel E.
    Segers, Patrick
    Smulyan, Harold
    [J]. HYPERTENSION, 2009, 54 (02) : 375 - 383
  • [2] Clustering of multiple lifestyle behaviours and its association to cardiovascular risk factors in children: the IDEFICS study
    Bel-Serrat, S.
    Mouratidou, T.
    Santaliestra-Pasias, A. M.
    Iacoviello, L.
    Kourides, Y. A.
    Marild, S.
    Molnar, D.
    Reisch, L.
    Siani, A.
    Stomfai, S.
    Vanaelst, B.
    Veidebaum, T.
    Pigeot, I.
    Ahrens, W.
    Krogh, V.
    Moreno, L. A.
    [J]. EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2013, 67 (08) : 848 - 854
  • [3] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [4] Non-invasively measured aortic wave reflection and pulse pressure amplification are related to the severity of coronary artery disease
    Cho, Sung Woo
    Kim, Byung Kyu
    Kim, Jeong Hoon
    Byun, Young Sup
    Goh, Choong Won
    Rhee, Kun Joo
    Ahn, Hyo Seung
    Lee, Byoung Kwon
    Kim, Byung Ok
    [J]. JOURNAL OF CARDIOLOGY, 2013, 62 (1-2) : 131 - 137
  • [5] Validity and Reliability of Central hood Pressure Estimated by Upper Arm Oscillometric Cuff Pressure
    Climie, Rachel E. D.
    Schultz, Martin G.
    Nikolic, Sonja B.
    Ahuja, Kiran D. K.
    Fell, James W.
    Sharman, James E.
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2012, 25 (04) : 414 - 420
  • [6] Increased arterial stiffness in children on haemodialysis
    Covic, A
    Mardare, N
    Gusbeth-Tatomir, P
    Brumaru, O
    Gavrilovici, C
    Munteanu, M
    Prisada, O
    Goldsmith, DJA
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2006, 21 (03) : 729 - 735
  • [7] Craig E, 2007, MONITORING HLTH NZ
  • [8] Validation of the Noninvasive Assessment of Central Blood Pressure by the SphygmoCor and Omron Devices Against the Invasive Catheter Measurement
    Ding, Feng-Hua
    Fan, Wang-Xiang
    Zhang, Rui-Yan
    Zhang, Qi
    Li, Yan
    Wang, Ji-Guang
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2011, 24 (12) : 1306 - 1311
  • [9] Fleiss J., 1986, Reliability of measurement: the design and analysis of clinical experiments
  • [10] The pulse wave velocity and extensibility of the brachial and radial artery in man.
    Fulton, JS
    McSwiney, BA