Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:16
|
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2018年 / 19卷 / 04期
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 50 条
  • [31] How Inaccurate is Weight as a Metric for Patient Size? Comparing Patient Weight to Effective Diameter for Size-Specific Dose Estimation
    Cook, Tessa S.
    Chadalavada, Seetharam C.
    Boonn, William W.
    MEDICAL IMAGING 2013: ADVANCED PACS-BASED IMAGING INFORMATICS AND THERAPEUTIC APPLICATIONS, 2013, 8674
  • [32] Development of a novel automated algorithm for patient dosimetry in computed tomography: a step towards the facilitation of size-specific dose estimates and organs dosimetry estimations in a busy clinical workflow
    Sekkat, Hamza
    Khallouqi, Abdellah
    Madkouri, Youssef
    El Atifi, Wiam
    Rhouch, Imane
    El Rhazouani, Omar
    Tahiri, Zakaria
    Talbi, Mohammed
    Halimi, Abdellah
    PHYSICA SCRIPTA, 2024, 99 (12)
  • [33] Relationships of Radiation Dose Indices with Body Size Indices in Adult Body Computed Tomography
    Inoue, Yusuke
    Itoh, Hiroyasu
    Nagahara, Kazunori
    Hata, Hirofumi
    Mitsui, Kohei
    TOMOGRAPHY, 2023, 9 (04) : 1381 - 1392
  • [34] Dependence of Water-equivalent Diameter and Size-specific Dose Estimates on CT Tube Potential
    McCollough, Cynthia H.
    McCollough, Shannon L.
    Schneider, Justine J.
    Moen, Taylor R.
    Weaver, Jayse M.
    Vrieze, Thomas J.
    Yu, Lifeng
    Leng, Shuai
    RADIOLOGY, 2022, 303 (02)
  • [35] RADIATION DOSE DETERMINATION IN ABDOMINAL CT EXAMINATIONS OF CHILDREN AT SUDANESE HOSPITALS USING SIZE-SPECIFIC DOSE ESTIMATES
    Bashier, Einas H.
    Suliman, I. I.
    RADIATION PROTECTION DOSIMETRY, 2019, 183 (04) : 444 - 449
  • [36] Image quality and dose in computed tomography
    Jurik, AG
    Jessen, KA
    Hansen, J
    EUROPEAN RADIOLOGY, 1997, 7 (01) : 77 - 81
  • [37] PATIENT DOSE SURVEY BASED ON SIZE-SPECIFIC DOSE ESTIMATE AND ACCEPTABLE QUALITY DOSE IN CHEST AND ABDOMEN/PELVIS CT EXAMINATIONS
    Mehdipour, Ali
    Parsi, Masoumeh
    Khorram, Faezeh-Sadat
    RADIATION PROTECTION DOSIMETRY, 2019, 185 (02) : 176 - 182
  • [38] A comparison study of size-specific dose estimate calculation methods
    Parikh, Roshni A.
    Wien, Michael A.
    Novak, Ronald D.
    Jordan, David W.
    Klahr, Paul
    Soriano, Stephanie
    Ciancibello, Leslie
    Berlin, Sheila C.
    PEDIATRIC RADIOLOGY, 2018, 48 (01) : 56 - 65
  • [39] Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults
    Desai, Gaurav S.
    Uppot, Raul N.
    Yu, Elaine W.
    Kambadakone, Avinash R.
    Sahani, Dushyant V.
    EUROPEAN RADIOLOGY, 2012, 22 (08) : 1631 - 1640
  • [40] CNN application for automated determination of the patient's size to obtain the size-specific dose estimated in CT
    Hernandez-Davila, Erik R.
    Torres-Garcia, Eugenio
    Aranda-Lara, Liliana
    Roldan-Valadez, Ernesto
    Isaac-Olive, Keila
    Flores-Reyes, Mario
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (03)