Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:18
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2018年 / 19卷 / 04期
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 24 条
[1]  
AAPM TG 220, 2014, AAPM REP 220, V220, P1
[2]  
Boone J.M., 2011, SIZE SPECIFIC DOSE E
[3]   Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[4]   Investigation of American Association of Physicists in Medicine Report 204 Size-specific Dose Estimates for Pediatric CT Implementation [J].
Brady, Samuel L. ;
Kaufman, Robert A. .
RADIOLOGY, 2012, 265 (03) :832-840
[5]   Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility [J].
Braun, Franziska M. ;
Johnson, Thorsten R. C. ;
Sommer, Wieland H. ;
Thierfelder, Kolja M. ;
Meinel, Felix G. .
EUROPEAN RADIOLOGY, 2015, 25 (06) :1598-1606
[6]   Size-specific Dose Estimation for CT: How Should It Be Used and What Does It Mean? [J].
Brink, James A. ;
Morin, Richard L. .
RADIOLOGY, 2012, 265 (03) :666-668
[7]   Size-specific Dose Estimates for Adult Patients at CT of the Torso [J].
Christner, Jodie A. ;
Braun, Natalie N. ;
Jacobsen, Megan C. ;
Carter, Rickey E. ;
Kofler, James M. ;
McCollough, Cynthia H. .
RADIOLOGY, 2012, 265 (03) :841-847
[8]   Automatic exposure control at MDCT based on the contrast-to-noise ratio: Theoretical background and phantom study [J].
Funama, Yoshinori ;
Sugaya, Yoshiaki ;
Miyazaki, Osamu ;
Utsunomiya, Daisuke ;
Yamashita, Yasuyuki ;
Awai, Kazuo .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (01) :39-47
[9]   Patient Size and Radiation Exposure in Thoracic, Pelvic, and Abdominal CT Examinations Performed With Automatic Exposure Control [J].
Israel, Gary M. ;
Cicchiello, Lawrence ;
Brink, James ;
Huda, Walter .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (06) :1342-1346
[10]  
J-RIME 2015, 2015, DIAGN REF LEV BAS LA