Evaluation of the Abbott ARCHITECT Toxo IgM assay

被引:9
作者
Sickinger, Eva [2 ]
Braun, Hans-Bertram [2 ]
Praast, Gerald [2 ]
Stieler, Myriam [2 ]
Gundlach, Cordelia [2 ]
Birkenbach, Claudia [2 ]
Prostko, John [1 ]
Palafox, Mary Ann [1 ]
Frias, Edwin [1 ]
Hsu, Stephen [1 ]
Matias, Matthew [1 ]
Pucci, Dominick [1 ]
Hausmann, Michael [2 ]
Sagel, Ulrich [3 ]
Smith, Darwin [1 ]
机构
[1] Abbott Diagnost, Abbott Pk, IL 60064 USA
[2] Abbott GmbH & Co KG, D-65205 Wiesbaden, Germany
[3] Austrian Agcy Hlth & Food Safety AGES, Inst Med Microbiol & Hyg, A-1096 Vienna, Austria
关键词
Toxoplasmosis; Toxo IgM; Acute infection; TOXOPLASMA-GONDII INFECTION; IMMUNOGLOBULIN-M ANTIBODIES; AGGLUTINATION-TEST; PREGNANT-WOMEN; RISK-FACTORS; AVIDITY; DIAGNOSIS; SYSTEM;
D O I
10.1016/j.diagmicrobio.2009.03.016
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Development of the ARCHITECT Toxo IgM assay has been done to assist the clinician in acute Toxoplasma gondii infection detection, especially in pregnant women. Its use, in conjunction with ARCHITECT Toxo IgG and Toxo Avidity assays, will provide an array of assays particularly useful in the monitoring of pregnant females to determine the risk of maternal transmission of the parasite. Specificity results from 2 testing sites, using populations of pregnant females, hospital patients, and blood donors, demonstrated that the assay has an overall resolved relative specificity of 99.89% (confidence interval, 99.68-99.98%). Relative specificity for pregnant female specimens was 99.95% (n = 2031). Excellent seroconversion sensitivity was observed for the ARCHITECT Toxo IgM assay, which was similar to the Abbott AxSYM Toxo IgM assay (Abbott Laboratories, Abbott Park, IL). In more than 90% of the panels tested, the 1st bleed detected in the serial bleeds was the same for both assays. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:275 / 282
页数:8
相关论文
共 50 条
  • [21] 宁波地区妇女不良妊娠结局与TOXO-IgM关系的分析
    翁华
    陈健
    中国优生与遗传杂志, 2009, 17 (10) : 78 - 78
  • [22] TOXO-IgM实验室检测室内质控方法的建立
    高志伟
    江晓华
    中国计划生育学杂志, 2011, 19 (01) : 43 - 45
  • [23] Comparative analysis of the VIDAS Toxo IgG IV assay in the detection of antibodies to Toxoplasma gondii
    Roux-Buisson, N
    Fricker-Hidalgo, H
    Foussadier, A
    Rolland, D
    Suchel-Jambon, AS
    Brenier-Pinchart, MP
    Pelloux, H
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2005, 53 (01) : 79 - 81
  • [24] Evaluation of a SARS-CoV-2 Capture IgM Antibody Assay in Convalescent Sera
    Ha, Binh
    Jadhao, Samadhan
    Hussaini, Laila
    Gibson, Theda
    Stephens, Kathy
    Ciric, Caroline
    Taylor, Meg
    Rouphael, Nadine
    Edupuganti, Srilatha
    Rostad, Christina A.
    Tompkins, S. Mark
    Anderson, Evan J.
    Anderson, Larry J.
    MICROBIOLOGY SPECTRUM, 2021, 9 (02):
  • [25] Performance Evaluation of Open Channel Buhlmann Fecal Calprotectin Turbo Assay on Abbott Alinity C Analyzer
    Sataranatarajan, Kavithalakshmi
    Adhikari, Shishir
    Nguyen, Ngoc
    Narasimhan, Madhusudhanan
    Balani, Jyoti
    Muthukumar, Alagarraju
    DIAGNOSTICS, 2024, 14 (16)
  • [26] Evaluation of an in-house IgM/IgG lateral flow assay for serodiagnosis of human brucellosis
    Shome, Rajeswari
    Patra, Sudipta
    Sahib, Muneera Mohamed
    Shanmugam, G.
    Dubey, Shiva
    Skariah, Somy
    Shamshad, Samer
    Barman, Nagendra Nath
    Bora, Durlav Prasad
    Shome, Arijit
    Mohandoss, Nagalingam
    Shome, Bibek Ranjan
    INDIAN JOURNAL OF MEDICAL MICROBIOLOGY, 2023, 42 : 55 - 58
  • [27] Evaluation of an Aspergillus IgG/IgM lateral flow assay for serodiagnosis of fungal asthma in Uganda
    Kwizera, Richard
    Bongomin, Felix
    Olum, Ronald
    Worodria, William
    Bwanga, Freddie
    Meya, David B.
    Kirenga, Bruce J.
    Gore, Robin
    Denning, David W.
    Fowler, Stephen J.
    PLOS ONE, 2021, 16 (05):
  • [28] Evaluation of the abbott LCx Mycobacterium tuberculosis assay for direct detection of Mycobacterium bovis in bovine tissue samples
    Valente, C
    Cuteri, V
    Ausili, E
    Piersimoni, C
    VETERINARY RESEARCH COMMUNICATIONS, 2002, 26 (01) : 21 - 27
  • [29] Evaluation of the Abbott LCx Mycobacterium Tuberculosis Assay for Direct Detection of Mycobacterium Bovis in Bovine Tissue Samples
    C. Valente
    V. Cuteri
    E. Ausili
    C. Piersimoni
    Veterinary Research Communications, 2002, 26 : 21 - 27
  • [30] Evaluation of a New Immunochromatography Technology Test (LDBio Diagnostics) To Detect Toxoplasma IgG and IgM: Comparison with the Routine Architect Technique
    Mahinc, Caroline
    Flori, Pierre
    Delaunay, Edouard
    Guillerme, Cecile
    Charaoui, Sana
    Raberin, Helene
    Hafid, Jamal
    L'Ollivier, Coralie
    JOURNAL OF CLINICAL MICROBIOLOGY, 2017, 55 (12) : 3395 - 3404