Correction of megavoltage cone-beam CT images of the pelvic region based on phantom measurements for dose calculation purposes

被引:11
作者
Aubry, Jeanm Francois [1 ]
Cheung, Joey [1 ]
Morin, Olivier [1 ]
Gottschalk, Alexander [1 ]
Beaulieu, Luc [2 ]
Pouliot, Jean [1 ]
机构
[1] UCSF Helen Diller Family Comprehens Canc Ctr, Dept Radiat Oncol, San Francisco, CA USA
[2] Univ Laval, Dept Phys, Quebec City, PQ G1K 7P4, Canada
关键词
megavoltage cone-beam CT; cupping artifact; missing data artifact; dose calculation; pelvis imaging; prostate cancer; GUIDED RADIATION-THERAPY; MVCT;
D O I
10.1120/jacmp.v10i1.2852
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Megavoltage cone-beam CT (MVCBCT) is an imaging technology that provides a 3D representation of the patient in treatment position. Because it is a form of x-ray tomography, MVCBCT images give information about the attenuation coefficients of the imaged tissues, and thus could be used for dose calculation. However, the cupping and missing data artifacts seen on MVCBCT images can cause inaccuracies in dose calculations. To eliminate these inaccuracies, a correction method specific to pelvis imaging and based on phantom measurements has been devised. Pelvis-shaped water phantoms of three different sizes were designed and imaged with MVCBCT Three sets of correction factors were created from the artifacts observed in these MVCBCT images by dividing the measured CT number by the predefined CT number for water. Linear interpolation is performed between the sets of correction factors to take into account the varying size of different patients. To compensate for the missing anatomy due to the limited field of view of the MVCBCT system, the MVCBCT image is complemented with the kilovoltage CT (kVCT) image acquired for treatment planning. When the correction method is applied to an anthropomorphic pelvis phantom, the standard deviation between dose calculations performed with kVCT and MVCBCT images is 0.6% with 98% of the dose points agreeing within +/- 3% With uncorrected MVCBCT images this percentage falls to 75% An example of dose calculation performed with a corrected clinical MVCBCT image of a prostate cancer patient shows that changes in anatomy of normal tissues result in variation of the dose distribution received by these tissues. This correction method enables MVCBCT images to be used for the verification of the daily dose distribution for patients treated in the pelvis region.
引用
收藏
页码:33 / 42
页数:10
相关论文
共 27 条
[1]   Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region [J].
Aubrya, Jean-Francois ;
Pouliot, Jean ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2008, 35 (03) :900-907
[2]   In-room CT techniques for image-guided radiation therapy [J].
Charlie, CM ;
Paskalev, K .
MEDICAL DOSIMETRY, 2006, 31 (01) :30-39
[3]   Dose-guided radiation therapy with megavoltage cone-beam CT [J].
Chen, J. ;
Morin, O. ;
Aubin, M. ;
Bucci, M. K. ;
Chuang, C. F. ;
Pouliot, J. .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 :S87-S98
[4]   CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS [J].
CHO, PS ;
JOHNSON, RH ;
GRIFFIN, TW .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11) :1863-1883
[5]   THE DEVELOPMENT OF CONFORMAL RADIATION-THERAPY [J].
FRAASS, BA .
MEDICAL PHYSICS, 1995, 22 (11) :1911-1921
[6]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[7]   The use of megavoltage CT (MVCT) images for dose recomputations [J].
Langen, KM ;
Meeks, SL ;
Poole, DO ;
Wagner, TH ;
Willoughby, TR ;
Kupelian, PA ;
Ruchala, KJ ;
Haimerl, J ;
Olivera, GH .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) :4259-4276
[8]  
LING CC, 1983, COMPUT TOMOGR, P284
[9]   Deformable registration of the planning image (kVCT) and the daily images (MVCT) for adaptive radiation therapy [J].
Lu, Weiguo ;
Olivera, Gustavo H. ;
Chen, Quan ;
Ruchala, Kenneth J. ;
Haimerl, Jason ;
Meeks, Sanford L. ;
Langen, Katja M. ;
Kupelian, Patrick A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (17) :4357-4374
[10]   Image guidance for precise conformal radiotherapy [J].
Mackie, TR ;
Kapatoes, J ;
Ruchala, K ;
Lu, WG ;
Wu, C ;
Olivera, G ;
Forrest, L ;
Tome, W ;
Welsh, J ;
Jeraj, R ;
Harari, P ;
Reckwerdt, P ;
Paliwal, B ;
Ritter, M ;
Keller, H ;
Fowler, J ;
Mehta, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :89-105