Safety and efficacy of argon plasma coagulation trimming of malpositioned and migrated biliary metal stents: A controlled study in the porcine model

被引:19
作者
Chen, Yang K.
Jakribettuu, Vaman
Springer, Erik W.
Shah, Raj J.
Penberthy, Jeanine
Nash, S. Russell
机构
[1] Univ Colorado, Hlth Sci Ctr, Dept Pathol, Denver, CO 80262 USA
[2] Univ Colorado, Hlth Sci Ctr, Div Gastroenterol & Hepatol, Denver, CO USA
关键词
D O I
10.1111/j.1572-0241.2006.00744.x
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
BACKGROUND: Argon plasma coagulation (APC) has been used to trim uncovered Elgiloy((R)) stents, but the extent of thermal damage and bile duct injury is not known. The goal of this study was to evaluate the safety and efficacy of APC for this application. METHODS: Eight Elgiloy((R)) (covered and uncovered) and four nitinol stents were deployed in the bile duct at ERCP in 12 anesthetized pigs. In nine pigs, the excess distal ends were trimmed in vivo using short bursts of APC. Three pigs served as controls. Bile ducts and stent specimens were then harvested for gross and histological examinations by a single-blinded pathologist. RESULTS: APC effectively trimmed all the stents. Seven APC-treated bile ducts and three controls showed epithelial distortion consistent with pressure injury from stent expansion. Two APC-treated bile ducts showed mild thermal injury. The damage was superficial, extending to a maximum depth of 0.1 mm with rare foci involving subepithelial connective tissue. CONCLUSION: APC at indicated settings effectively cuts covered and uncovered Elgiloy((R)) as well as nitinol stents, but can cause biliary epithelial injury secondary to conduction of heat and electrical energy. Proper technique and settings should be followed and short bursts of energy judiciously applied in order to minimize this danger.
引用
收藏
页码:2025 / 2030
页数:6
相关论文
共 18 条
  • [1] AHMED A, 1997, GASTROINTEST ENDOSC, V50, P279
  • [2] Endoscopic removal of a ''floating'' biliary Gianturco Z stent five years after placement for a benign anastomotic stricture in a liver transplant patient
    Baron, TH
    Blackard, WG
    Morgan, DE
    [J]. GASTROINTESTINAL ENDOSCOPY, 1997, 46 (01) : 80 - 82
  • [3] A comparison study on physical properties of self-expandable esophageal metal stents
    Chan, ACW
    Shin, FG
    Lam, YH
    Ng, EKW
    Sung, JJY
    Lau, JYW
    Chung, SCS
    [J]. GASTROINTESTINAL ENDOSCOPY, 1999, 49 (04) : 462 - 465
  • [4] Randomized controlled trial comparing epinephrine injection plus heat probe coagulation versus epinephrine injection plus argon plasma coagulation for bleeding peptic ulcers
    Chau, CH
    Siu, WT
    Law, BKB
    Tang, CN
    Kwok, SY
    Luk, YW
    Lao, WC
    Li, MKW
    [J]. GASTROINTESTINAL ENDOSCOPY, 2003, 57 (04) : 455 - 461
  • [5] Endoscopic palliation of patients with biliary obstruction caused by nonresectable hilar cholangiocarcinoma: efficacy of self-expandable metallic Wallstents
    Cheng, JLS
    Bruno, MJ
    Bergman, JJ
    Rauws, EA
    Tytgat, GN
    Huibregtse, K
    [J]. GASTROINTESTINAL ENDOSCOPY, 2002, 56 (01) : 33 - 39
  • [6] Biliary strictures in liver transplant recipients: Treatment with metal stents
    Culp, WC
    McCowan, TC
    Lieberman, RP
    Goertzen, TC
    LeVeen, RF
    Heffron, TG
    [J]. RADIOLOGY, 1996, 199 (02) : 339 - 346
  • [7] Argon plasma endoscopic section of biliary metallic prostheses
    Demarquay, JF
    Dumas, R
    Peten, EP
    Rampal, P
    [J]. ENDOSCOPY, 2001, 33 (03) : 289 - 290
  • [8] DIAMOND G, 1995, J VASC INTERV RADIOL, V6, P755
  • [9] Self-expandable metal stents
    Dumonceau, JM
    Devière, J
    [J]. BEST PRACTICE & RESEARCH CLINICAL GASTROENTEROLOGY, 1999, 13 (01) : 109 - 130
  • [10] Endoscopic removal of an embedded biliary Wallstent by piecemeal extraction
    Egan, LJ
    Baron, TH
    [J]. ENDOSCOPY, 2000, 32 (06) : 492 - 494