共 31 条
Wide-range mass calibration for negative-ion DART mass spectrometry
被引:13
作者:
Gross, Juergen H.
[1
]
机构:
[1] Heidelberg Univ, Inst Organ Chem, D-69120 Heidelberg, Germany
关键词:
Direct analysis in real time;
DART;
mass spectrometry;
Agilent ES tune mix;
phosphazene;
mass calibration;
accurate mass;
Fourier transform ion cyclotron resonance;
FT-MS;
Orbitrap;
REAL-TIME;
ELECTROSPRAY-IONIZATION;
NONPOLAR COMPOUNDS;
REFERENCE COMPOUND;
OPEN-AIR;
IDENTIFICATION;
ULTRAMARK-1621;
MS;
D O I:
10.1255/ejms.1266
中图分类号:
O64 [物理化学(理论化学)、化学物理学];
O56 [分子物理学、原子物理学];
学科分类号:
070203 ;
070304 ;
081704 ;
1406 ;
摘要:
A method for accurate mass calibration covering a wide m/z range in negative-ion direct analysis in real time mass spectrometry (DART-MS) is described. It is based on a commercially available tuning solution (Agilent ES tune mix) normally applied in electrospray ionization. This acetonitrile solution is composed of ammonium trifluoroacetate, betaine, 2,4,6-tris(heptafluoropropyl)-1,3,5-triazine, and various symmetrical hexakis-(fluoroalkoxy)-phosphazenes. Upon addition of 2% (v/v) trifluoroacetic acid in negative-ion DART-MS, this mixture yields negative ions covering a range of m/z 69 to m/z 2834. Validity of the calibration procedure is demonstrated on a Fourier transform ion cyclotron resonance mass spectrometer by application to polyethylene glycol diacid and [60]fullerene. Experimental conditions and a mass reference list are provided.
引用
收藏
页码:155 / 161
页数:7
相关论文