Monte Carlo study shows no significant difference in second cancer risk between 6-and 18-MV intensity-modulated radiation therapy

被引:41
作者
Kry, Stephen F. [1 ]
Salehpour, Mohammad [1 ]
Titt, Uwe [1 ]
White, R. Allen [2 ]
Stovall, Marilyn [1 ]
Followill, David [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Biostat & Appl Math, Houston, TX 77030 USA
关键词
IMRT; Out-of-field dose; Second cancer; Monte Carlo; Neutrons; DYNAMIC MLC IMRT; SECONDARY MALIGNANCIES; CALCULATED RISK; DOSE-EQUIVALENT; MV BEAM; RADIOTHERAPY; NEUTRON; PHOTON; ENERGIES; SIMULATION;
D O I
10.1016/j.radonc.2008.11.020
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the photon and neutron out-of-field dose equivalents from 6- and 18-MV intensity-modulated radiation therapy (IMRT) and to investigate the impact of the differences on the associated risk of induced second malignancy using a Monte Carlo model. Methods and materials: A Monte Carlo model created with MCNPX was used to calculate the out-of-field photon dose and neutron dose equivalent from simulated IMRT of the prostate conducted at beam energies of 6 and 18 MV. The out-of-field dose equivalent was calculated at the locations of sensitive organs in an anthropomorphic phantom. Based on these doses, the risk of secondary malignancy was calculated based on organ-, gender-, and age-specific risk coefficients for a 50-year-old man. Results: The Monte Carlo model predicted much lower neutron dose equivalents than had been determined previously. Further analysis illuminated the large uncertainties in the neutron dose equivalent and demonstrated the need for better determination of this value, which plays a large role in estimating the risk of secondary malignancies. The Monte Carlo calculations found that the differences in the risk of secondary malignancies conferred by high-energy IMRT versus low-energy IMRT are minimal and insignificant, contrary to prior findings. Conclusions: The risk of secondary malignancy associated with high-energy radiation therapy may not be as large as previously reported, and likely should not deter the use of high-energy beams. However, the large uncertainties in neutron dose equivalents at specific locations within the patient warrant further study so that the risk of secondary cancers can be estimated with greater accuracy. (C) 2008 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 91 (2009) 132-137
引用
收藏
页码:132 / 137
页数:6
相关论文
共 30 条
[1]   Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18 MV x-ray LINAC-based radiotherapy [J].
Barquero, R ;
Edwards, TM ;
Iñiguez, MP ;
Vega-Carrillo, HR .
MEDICAL PHYSICS, 2005, 32 (12) :3579-3588
[2]  
*BEIR, 2006, 7 BEIR
[3]  
BOICE JD, 1985, JNCI-J NATL CANCER I, V74, P955
[4]  
Brenner DJ, 2000, CANCER, V88, P398, DOI 10.1002/(SICI)1097-0142(20000115)88:2<398::AID-CNCR22>3.0.CO
[5]  
2-V
[6]   Depth dose-equivalent and effective energies of photoneutrons generated by 6-18 MV X-ray beams for radiotherapy [J].
d'Errico, F ;
Luszik-Bhadra, M ;
Nath, R ;
Siebert, BRL ;
Wolf, U .
HEALTH PHYSICS, 2001, 80 (01) :4-11
[7]   The use of risk estimation models for the induction of secondary cancers following radiotherapy [J].
Dasu, A ;
Toma-Dasu, I ;
Olofsson, J ;
Karlsson, M .
ACTA ONCOLOGICA, 2005, 44 (04) :339-347
[8]   Second primary tumors after radiotherapy for malignancies -: Treatment-related parameters [J].
Dörr, W ;
Herrmann, T .
STRAHLENTHERAPIE UND ONKOLOGIE, 2002, 178 (07) :357-362
[9]   Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy [J].
Followill, D ;
Geis, P ;
Boyer, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :667-672
[10]   IMRT should not be administered at photon energies greater than 10 MV [J].