IMRT delivery to a moving target by dynamic MLC tracking: delivery for targets moving in two dimensions in the beam's eye view

被引:61
作者
McQuaid, D.
Webb, S.
机构
[1] Inst Canc Res, Joint Dept Phys, Sutton SM2 5PT, Surrey, England
[2] Royal Marsden NHSF Trust, Sutton SM2 5PT, Surrey, England
关键词
D O I
10.1088/0031-9155/51/19/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new modification of the dynamic multileaf collimator (dMLC) delivery technique for intensity-modulated therapy (IMRT) is outlined. This technique enables the tracking of a target moving through rigid-body translations in a 2D trajectory in the beam's eye view. The accuracy of the delivery versus that of deliveries with no tracking and of 1D tracking techniques is quantified with clinically derived intensity-modulated beams (IMBs). Leaf trajectories calculated in the target-reference frame were iteratively synchronized assuming regular target motion. This allowed the leaves defined in the lab-reference frame to simultaneously follow the target motion and to deliver the required IMB without violation of the leaf maximum-velocity constraint. The leaves are synchronized until the gradient of the leaf position at every instant is less than a calculated maximum. The delivered fluence in the target-reference frame was calculated with a simple primary-fluence model. The new 2D tracking technique was compared with the delivered fluence produced by no-tracking deliveries and by 1D tracking deliveries for 33 clinical IMBs. For the clinical IMBs normalized to a maximum fluence of 200 MUs, the rms difference between the desired and the delivered IMB was 15.6 +/- 3.3 MU for the case of a no-tracking delivery, 7.9 +/- 1.6 MU for the case where only the primary component of motion was corrected and 5.1 +/- 1.1 MU for the 2D tracking delivery. The residual error is due to interpolation and sampling effects. The 2D tracking delivery technique requires an increase in the delivery time evaluated as between 0 and 50% of the unsynchronized delivery time for each beam with a mean increase of 13% for the IMBs tested. The 2D tracking dMLC delivery technique allows an optimized IMB to be delivered to moving targets with increased accuracy and with acceptable increases in delivery time. When combined with real-time knowledge of the target motion at delivery time, this technique facilitates improved target conformality relative to no-tracking deliveries and allows PTV margin reduction.
引用
收藏
页码:4819 / 4839
页数:21
相关论文
共 19 条
[1]   Effects of intra-fraction motion on IMRT dose delivery: statistical analysis and simulation [J].
Bortfeld, T ;
Jokivarsi, K ;
Goitein, M ;
Kung, J ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (13) :2203-2220
[2]   Effects of averaging over motion and the resulting systematic errors in radiation therapy [J].
Evans, PM ;
Coolens, C ;
Nioutsikou, E .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (01) :N1-N7
[3]   Organ and tumor motion: An overview [J].
Goitein, M .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :2-9
[4]   Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking [J].
Keall, PJ ;
Joshi, S ;
Vedam, SS ;
Siebers, JV ;
Kini, VR ;
Mohan, R .
MEDICAL PHYSICS, 2005, 32 (04) :942-951
[5]   High-tech will improve radiotherapy of NSCLC: A hypothesis waiting to be validated [J].
Ling, CC ;
Yorke, E ;
Amols, H ;
Mechalakos, J ;
Erdi, Y ;
Leibel, S ;
Rosenzweig, K ;
Jackson, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01) :3-7
[6]   A method for incorporating organ motion due to breathing into 3D dose calculations in the liver: Sensitivity to variations in motion [J].
Lujan, AE ;
Balter, JM ;
Ten Haken, RK .
MEDICAL PHYSICS, 2003, 30 (10) :2643-2649
[7]   Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment [J].
Mageras, GS ;
Yorke, E .
SEMINARS IN RADIATION ONCOLOGY, 2004, 14 (01) :65-75
[8]   Real-time DMLC IMRT delivery for mobile and deforming targets [J].
Papiez, L ;
Rangaraj, D ;
Keall, P .
MEDICAL PHYSICS, 2005, 32 (09) :3037-3048
[9]   DMLC leaf-pair optimal control for mobile, deforming target [J].
Papiez, L ;
Rangaraj, D .
MEDICAL PHYSICS, 2005, 32 (01) :275-285
[10]   DMLC leaf-pair optimal control of IMRT delivery for a moving rigid target [J].
Papiez, L .
MEDICAL PHYSICS, 2004, 31 (10) :2742-2754