DigiPET: sub-millimeter spatial resolution small-animal PET imaging using thin monolithic scintillators

被引:95
作者
Espana, Samuel [1 ]
Marcinkowski, Radoslaw [1 ]
Keereman, Vincent [1 ]
Vandenberghe, Stefaan [1 ]
Van Holen, Roel [1 ]
机构
[1] Ghent Univ IMinds IBiTech, MEDISIP, Dept Elect & Informat Syst, B-9000 Ghent, Belgium
关键词
PET; monolithic crystals; dSiPM; PERFORMANCE EVALUATION; CONTINUOUS LSO; BAF2; CRYSTALS; SCANNER; DETECTORS; MICROPET; POSITION; DESIGN; PHOTOMULTIPLIER; CALIBRATION;
D O I
10.1088/0031-9155/59/13/3405
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new preclinical PET system based on dSiPMs, called DigiPET, is presented. The system is based on thin monolithic scintillation crystals and exhibits superior spatial resolution at low-cost compared to systems based on pixelated crystals. Current dedicated small-rodent PET scanners have a spatial resolution in the order of 1 mm. Most of them have a large footprint, requiring considerable laboratory space. For rodent brain imaging, a PET scanner with sub-millimeter resolution is desired. To achieve this, crystals with a pixel pitch down to 0.5 mm have been used. However, fine pixels are difficult to produce and will render systems expensive. In this work, we present the first results with a high-resolution preclinical PET scanner based on thin monolithic scintillators and a large solid angle. The design is dedicated to rat-brain imaging and therefore has a very compact geometry. Four detectors were placed in a square arrangement with a distance of 34.5 mm between two opposing detector modules, defining a field of view (FOV) of 32 x 32 x 32 mm(3). Each detector consists of a thin monolithic LYSO crystal of 32 x 32 x 2 mm3 optically coupled to a digital silicon photomultiplier (dSiPM). Event positioning within each detector was obtained using the maximum likelihood estimation (MLE) method. To evaluate the system performance, we measured the energy resolution, coincidence resolving time (CRT), sensitivity and spatial resolution. The image quality was evaluated by acquiring a hot-rod phantom filled with F-18-FDG and a rat head one hour after an 18F-FDG injection. The MLE yielded an average intrinsic spatial resolution on the detector of 0.54 mm FWHM. We obtained a CRT of 680 ps and an energy resolution of 18% FWHM at 511 keV. The sensitivity and spatial resolution obtained at the center of the FOV were 6.0 cps kBq(-1) and 0.7 mm, respectively. In the reconstructed images of the hot-rod phantom, hot rods down to 0.7 mm can be discriminated. In conclusion, a compact PET scanner was built using dSiPM technology and thin monolithic LYSO crystals. Excellent spatial resolution and acceptable sensitivity were demonstrated. Promising results were also obtained in a hot-rod phantom and in rat-brain imaging.
引用
收藏
页码:3405 / 3420
页数:16
相关论文
共 39 条
[1]   Scanner calibration of a small animal PET camera based on continuous LSO crystals and flat panel PSPMTs [J].
Benlloch, J. M. ;
Carrilero, V. ;
Gonzalez, A. J. ;
Catret, J. ;
Lerche, Ch. W. ;
Abellan, D. ;
de Quiros, F. Garcia ;
Gimenez, M. ;
Modia, J. ;
Sanchez, F. ;
Pavon, N. ;
Ros, A. ;
Martinez, J. ;
Sebastia, A. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :26-29
[2]   Neural network-based position estimators for PET detectors using monolithic LSO blocks [J].
Bruyndonckx, P ;
Léonard, S ;
Tavernier, S ;
Lemaître, C ;
Devroede, O ;
Wu, YB ;
Krieguer, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2520-2525
[3]   Design and physical characteristics of a small animal PET using BaF2 crystals and a photosensitive wire chamber [J].
Bruyndonckx, P ;
Liu, X ;
Rajeswaran, S ;
Smolik, W ;
Tavernier, S ;
Zhang, SP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 382 (03) :589-600
[4]   Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber [J].
Bruyndonckx, P ;
Liu, XA ;
Tavernier, S ;
Zhang, SP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3) :407-413
[5]  
Chatziioannou AF, 1999, J NUCL MED, V40, P1164
[6]   Multimodality in vivo imaging systems: Twice the power or double the trouble? [J].
Cherry, Simon R. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2006, 8 :35-62
[7]  
Cherry SR, 2004, PHYS MED BIOL, V49, pR13, DOI 10.1088/0031-9155/49/3/R01
[8]   MicroPET: A high resolution PET scanner for imaging small animals [J].
Cherry, SR ;
Shao, Y ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Chatziioannou, A ;
Young, JW ;
Jones, WF ;
Moyers, JC ;
Newport, D ;
Boutefnouchet, A ;
Farquhar, TH ;
Andreaco, M ;
Paulus, MJ ;
Binkley, DM ;
Nutt, R ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1161-1166
[9]  
Degenhardt C, 2010, IEEE NUCL SCI CONF R, P1954, DOI 10.1109/NSSMIC.2010.5874115
[10]   Use of a YRP:Ce matrix coupled to a position-sensitive photomultiplier for high resolution position emission tomography [J].
DelGuerra, A ;
deNotaristefani, F ;
DiDomenico, G ;
Giganti, M ;
Pani, R ;
Piffanelli, A ;
Zavattini, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1958-1962